=== Check list === === Mode Cleaner Suspensions ===
Each MC suspension has 4 coils and 2 pico-motors (pitch, yaw).
Line 3: Line 4:
 * === PR3 ===
Line 5: Line 6:
=== In the case of VIS/subsystem failure or emergency ===
Line 7: Line 7:
 * === ETM ===

Each iKAGRA ETM suspension has 4 coils and 4 pico-motors connected.
If the coil-magnet actuator seems to have a problem, please check 2 things.
1. Coil drivers in particular for Y-end had a loose connection with power line (-18V, -15V).
Please check the LED on coil driver. If the LED is out, try to fasten the power line.
2. Sometimes watchdog? system stop the output current from the coil driver in order to
protect mechanical system.

Mode Cleaner Suspensions

Each MC suspension has 4 coils and 2 pico-motors (pitch, yaw).

PR3

ETM

Each iKAGRA ETM suspension has 4 coils and 4 pico-motors connected. If the coil-magnet actuator seems to have a problem, please check 2 things. 1. Coil drivers in particular for Y-end had a loose connection with power line (-18V, -15V). Please check the LED on coil driver. If the LED is out, try to fasten the power line. 2. Sometimes watchdog? system stop the output current from the coil driver in order to protect mechanical system.