Deletions are marked like this. Additions are marked like this.
Line 10: Line 10:
|| || [[https://www.vaisala.com/sites/default/files/documents/PTB110%20User%20Guide%20in%20English.pdf|(Manual)]] PTB110(Vaisala) || Pressure || || cRIO, KAGRA CDS ||
|| || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7554|(Manual)]] Trillium 120QA|| seismic || || KAGRA CDS ||
||<|2>for KAGRA || [[https://www.vaisala.com/sites/default/files/documents/PTB110%20User%20Guide%20in%20English.pdf|(Manual)]] PTB110(Vaisala) || Pressure || || cRIO, KAGRA CDS ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7554|(Manual)]] Trillium 120QA|| seismic || || KAGRA CDS ||

Environmental Sensors

Overview

  • {*} ひずみ計に対する各センサーをプロジェクションしたASDを載せる。

Sensors List

a

Sensor

Use

Spec

DAQ1

Air Environment

(Manual) HMP155(Vaisala)

Temperature

cRIO, KAGRA CDS

Humidity

cRIO, KAGRA CDS

for KAGRA

(Manual) PTB110(Vaisala)

Pressure

cRIO, KAGRA CDS

(Manual) Trillium 120QA

seismic

KAGRA CDS

(Manual) Trillium Compact

seismic

KAGRA CDS

(Manual) Trillium 240

seismic

cRIO

(Manual) CMG3T

seismic

cRIO

Deployment

  • {*} 配置図を載せる。

Circuit

  • {*} アンプの回路図、Dsubの配線図等のJGWのリンクを載せる。

Calibration

cRIO01 500m, y[V]がADの電圧値。上からAD00,01,02,03

t[℃]=(y[V]/10/(1+3300/430)/1.e-3-107.79)/.388+20.0
h[%]=-y[V]/5*100/2
p[hPa]=-y[V]/2/5*(1100-800)+800
vac[Pa]=10^[2*y[V]-13.0]

cRIO02 2000m

t[℃]=(y[V]/10/(1+3300/430)/1.e-3-107.79)/.388+20.0
h[%]=y[V]/5*100/2
p[hPa]=y[V]/2/5*(1100-800)+800
vac[Pa]=10^[2*y[V]-13.0]

Reference

Envrionmental Air Sensors


  1. More detail information about cRIO is here. (1)