= EX0 rack =
measured behind EX0 rack
[[attachment:ex0_2.jpg|jpg]]
= Cryocooler P53 =
measured near Cryocooler P53
[[attachment:cryocooler.jpg|jpg]]
Line 5: Line 5:
There is noise in 60-150 MHz. measured in 6 situations
|| situation || small cryocooler || big cryocooler || FFU filters & fan for EXC chamber ||
|| 1 || off || off || on ||
|| 2 || on || off || on ||
|| 3 || on || on || on ||
|| 4 || on || on || off ||
|| 5 || off || off || off ||
Line 7: Line 13:
The peak of the noise is around 112.5 MHz. Situation 6 is situation 5 with an ele-mag protection sheet
Line 9: Line 15:
= 0-250 MHz =
{{attachment:ex0_0_250_2.jpg|jpg}}
== situation 1 ==
Line 12: Line 17:
= 0-5 MHz =
{{attachment:ex0_0_5_2.jpg|jpg}}
||<:> ~+0-250 MHz+~ ||<:> ~+0-20 MHz+~ ||<:> ~+0-5 MHz+~ ||
||{{attachment:cryocooler1_0_250.png|png}}||{{attachment:cryocooler1_0_20.png|png}}||{{attachment:cryocooler1_0_5.png|png}}||
Line 15: Line 20:
= 50-150 MHz =
{{attachment:ex0_50_150_2.jpg|jpg}}
== situation 2 ==
Line 18: Line 22:
= 110-115 MHz =
{{attachment:ex0_110_115_2.jpg|jpg}}
||<:> ~+0-250 MHz+~ ||<:> ~+0-20 MHz+~ ||<:> ~+0-5 MHz+~ ||
||{{attachment:cryocooler2_0_250.png|png}}||{{attachment:cryocooler2_0_20.png|png}}||{{attachment:cryocooler2_0_5_2.png|png}}||

== situation 3 ==

||<:> ~+0-250 MHz+~ ||<:> ~+0-20 MHz+~ ||<:> ~+0-5 MHz+~ ||
||{{attachment:cryocooler3_0_250.png|png}}||{{attachment:cryocooler3_0_20.png|png}}||{{attachment:cryocooler3_0_5.png|png}}||

== situation 4 ==

||<:> ~+0-250 MHz+~ ||<:> ~+0-20 MHz+~ ||<:> ~+0-5 MHz+~ ||
||{{attachment:cryocooler4_0_250.png|png}}||{{attachment:cryocooler4_0_20.png|png}}||{{attachment:cryocooler4_0_5.png|png}}||

== situation 5 ==

||<:> ~+0-250 MHz+~ ||<:> ~+0-20 MHz+~ ||<:> ~+0-5 MHz+~ ||
||{{attachment:cryocooler5_0_250.png|png}}||{{attachment:cryocooler5_0_20.png|png}}||{{attachment:cryocooler5_0_5.png|png}}||

== situation 6 ==

||<:> ~+0-250 MHz+~ ||<:> ~+0-20 MHz+~ ||<:> ~+0-5 MHz+~ ||
||{{attachment:cryocooler6_0_250.png|png}}||{{attachment:cryocooler6_0_20.png|png}}||{{attachment:cryocooler6_0_5.png|png}}||

Cryocooler P53

measured near Cryocooler P53 jpg

measured in 6 situations

situation

small cryocooler

big cryocooler

FFU filters & fan for EXC chamber

1

off

off

on

2

on

off

on

3

on

on

on

4

on

on

off

5

off

off

off

Situation 6 is situation 5 with an ele-mag protection sheet

situation 1

0-250 MHz

0-20 MHz

0-5 MHz

png

png

png

situation 2

0-250 MHz

0-20 MHz

0-5 MHz

png

png

png

situation 3

0-250 MHz

0-20 MHz

0-5 MHz

png

png

png

situation 4

0-250 MHz

0-20 MHz

0-5 MHz

png

png

png

situation 5

0-250 MHz

0-20 MHz

0-5 MHz

png

png

png

situation 6

0-250 MHz

0-20 MHz

0-5 MHz

png

png

png