Differences between revisions 27 and 28
Revision 27 as of 2019-11-12 18:40:42
Size: 1988
Comment:
Revision 28 as of 2019-11-13 12:28:59
Size: 2235
Comment:
Deletions are marked like this. Additions are marked like this.
Line 34: Line 34:
Top filter

([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7758|7758]])

Standard Flter

([[http://p://klog.icrr.u-tokyo.ac.jp/osl/?r=4720|4720]])

Bottom Filter

Other related entries

 [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8658|8658]],

Describe KAGRA/Subgroups/VIS/OpsManual/FishingRods here.

This page describes the status of the fishing rods.

Beam splitter

Top filter

  • At 25 ºC it's possible to use the fishing rod to set the keystone at nominal position (8125).

  • at 24.5 ºC or lower it's not (8645, 11465).

  • Range: from -2500 µm to +4900 µm in the LVDT and from -1,400,000 to +1,400,000 steps in the stepper motor (7694). What was the temperature that day?

Standard Filter

  • It's been at its lowest position since the temperature was 25 ºC. We set the keystone to 488 µm and couldn't move it down more (8125, 8645).

  • Range: from -700 µm to +700 µm in the LVDT and from -870,000 to +835,000 steps in the stepper motor (3558). What was the temperature that day?

Bottom filter

  • This fishing rod didn't seem to have any problems before but we haven't tried to use it at every temperature condition we've had.
  • Range: from -1000 µm to +1800 µm in the LVDT and from -746,000 to +746,000 steps in the stepper motor (3280).What was the temperature that day?

SR2

Top filter

Range: [ -1,700,000 , 1,500,000 ]

Stardard filter

Range: from -750 µm to 1000 µm in the LVDT and from 300,000 to 1,000,00 steps in the stepper motor (4431).

Bottom filter

Range: from -500 µm to 600 µm in the LVDT and from -550,000 to -250,000 (4430).

8082

SR3

Top filter

(7758)

Standard Flter

(4720)

Bottom Filter

Other related entries

KAGRA/Subgroups/VIS/OpsManual/FishingRods (last edited 2020-09-18 10:45:40 by fabian.arellano)