= KAGRA VIS Operations Manual - Troubleshooting = == Check List == * デジタルシステムトラブル * RTモデルが動かない等の問題はひとまず以下を参照:[[KAGRA/Subgroups/DGS/Manual/CheckList|DGS Checklist]] * 伝達関数等を測定中にdiagguiが落ちて、excitationが止まらない:[[#tpclear| [解決策] ]] * * ---- <> === Clearing Excitation Testpoints === If excitation from `diaggui` won't stop, the fix is to clear all the test points for the real time model. You need to know the number of the user process running on the front end, which can be read from the `GDS_TP` screen, accessible from a button on the `OVERVIEW` screen and other places:<
> {{attachment:K1VISSR2GDSTP Screen.png||width=600}}<