Size: 1125
Comment:
|
Size: 1380
Comment:
|
Deletions are marked like this. | Additions are marked like this. |
Line 3: | Line 3: |
* Measured IMC length: 53.3 m http://klog.icrr.u-tokyo.ac.jp/osl/?r=616 |
* Measured IMC length: 53.3 m * http://klog.icrr.u-tokyo.ac.jp/osl/?r=616 |
Line 7: | Line 7: |
* Measured IMC mode matching ratio: 86.2 % https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=5019 |
* Measured IMC mode matching ratio: 86.2 % * https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=5019 |
Line 11: | Line 11: |
* Measured MCe RoC: 37.33(9) m http://klog.icrr.u-tokyo.ac.jp/osl/?r=711 |
* Measured MCe RoC: 37.33(9) m * http://klog.icrr.u-tokyo.ac.jp/osl/?r=711 |
Line 14: | Line 14: |
* Measured AOM actuation efficiency: 5.064(7) MHz/V http://klog.icrr.u-tokyo.ac.jp/osl/?r=3172 |
* Measured AOM actuation efficiency: 5.064(7) MHz/V * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3172 |
Line 17: | Line 17: |
* Measured MCe actuation efficiency: 6.9 um/V (any data for MCi and MCo?) http://gwclio.icrr.u-tokyo.ac.jp/lcgtsubgroup/inoutoptics/2015/06/mce625-26.html#comments |
* Measured MCe actuation efficiency: 6.9 um/V (any data for MCi and MCo?) * http://gwclio.icrr.u-tokyo.ac.jp/lcgtsubgroup/inoutoptics/2015/06/mce625-26.html#comments |
Line 21: | Line 20: |
* | * Calibrated IMC WFS noise level, and some noise hunting: * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3366 * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3335 |
Line 23: | Line 24: |
* Calibrated IMC WFS noise level, and some noise hunting: http://klog.icrr.u-tokyo.ac.jp/osl/?r=3366 http://klog.icrr.u-tokyo.ac.jp/osl/?r=3335 |
* Accelerometer on the PSL table: * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3373 * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3435 * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3689 |
Line 27: | Line 29: |
* Accelerometer on the PSL table: http://klog.icrr.u-tokyo.ac.jp/osl/?r=3373 http://klog.icrr.u-tokyo.ac.jp/osl/?r=3435 http://klog.icrr.u-tokyo.ac.jp/osl/?r=3689 |
* IMC length Noise``Budget code: * (UPDATE SVN) |
Line 32: | Line 32: |
* IMC length NoiseBudget code: (UPDATE SVN) |
* Frequency stabilization servo modeling (design): * https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=3330 * https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=2908 |
IMC Information
- Measured IMC length: 53.3 m
(Could be slightly different right now since we had re-installation work after this measurement)
- Measured IMC mode matching ratio: 86.2 %
(Could be fairly different right now since we had re-installation work after this measurement)
- Measured MCe RoC: 37.33(9) m
- Measured AOM actuation efficiency: 5.064(7) MHz/V
- Measured MCe actuation efficiency: 6.9 um/V (any data for MCi and MCo?)
- Calibrated IMC WFS noise level, and some noise hunting:
- Accelerometer on the PSL table:
IMC length NoiseBudget code:
- (UPDATE SVN)
- Frequency stabilization servo modeling (design):