Size: 2039
Comment:
|
← Revision 22 as of 2017-11-28 19:23:27 ⇥
Size: 3368
Comment:
|
Deletions are marked like this. | Additions are marked like this. |
Line 1: | Line 1: |
[[KAGRA/Commissioning/CommissioningWorkshop2017|Back to CommissioningWorkshop2017]] | |
Line 6: | Line 7: |
* slow path to AOM??? | * SLOW ('''K1:IMC-SERVO_SLOW_CON_IN1''') goes to AOM (and IMC length) * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3190 * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3544 |
Line 11: | Line 14: |
* Measured IMC modulation depth: ~0.1 rad (measured??) | * Measured IMC modulation depth: ~0.1 rad |
Line 14: | Line 17: |
* Measured IMC finesse: 540 +/- 70 | * Measured IMC finesse: 540 +/- 70 (NOTE: 458 from cavity pole, not klogged) |
Line 17: | Line 20: |
* Measured IMC length: 53.3 m * http://klog.icrr.u-tokyo.ac.jp/osl/?r=616 (Could be slightly different right now since we had re-installation work after this measurement) |
* Measured IMC round-trip length: 53.3 m (Could be slightly different right now since we had re-installation work after this measurement) * http://klog.icrr.u-tokyo.ac.jp/osl/?r=616 |
Line 20: | Line 23: |
* Measured IMC mode matching ratio: 86.2 % * https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=5019 (Could be fairly different right now since we had re-installation work after this measurement) |
* Measured IMC mode matching ratio: 86.2 % (Could be fairly worse right now since we had re-installation work after this measurement) * https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=5019 |
Line 26: | Line 29: |
* Measured AOM actuation efficiency: 5.064(7) MHz/V | * Measured AOM actuation efficiency: 5.064(7) MHz/V (x2 for double-pass) |
Line 29: | Line 32: |
* Measured actuation efficiency: 17.2 MHz/V(MCe), 15.9 MHz/V(MCi), 14.9 MHz/V(MCo) @DC | * Measured laser PZT actuation efficiency: 1.48 MHz/V (NOTE: 1.26 MHz/V using AOM, not klogged) * http://klog.icrr.u-tokyo.ac.jp/osl/?r=2853 * Laser temperature actuation efficiency: not measured yet * Measured mirror actuation efficiency: 17.2 MHz/V(MCe), 15.9 MHz/V(MCi), 14.9 MHz/V(MCo) @DC (could be wrong; inconsistent between measurements) |
Line 32: | Line 40: |
* MCe pendulum frequency: 0.955 Hz | |
Line 33: | Line 42: |
* Measured IMC optical gain: 18.6 V/MHz (cavity pole=5.6kHz) | * Measured IMC optical gain: 18.6 V/MHz (cavity pole=5.6kHz) (NOT UP TO DATE; we changed laser power, detection optics since then) |
Line 35: | Line 44: |
* Reference cavity parameters: finesse 1.1e4, length 10 cm (spec), cavity pole 68 kHz * Counts to volts, volts to counts: * ADC: 2^16/40 counts/V * DAC: 20/2^16 V/counts * IMC oplev calibration: * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3175 |
|
Line 46: | Line 64: |
* (UPDATE SVN) | * https://granite.phys.s.u-tokyo.ac.jp/svn/LCGT/trunk/kagranoisebudget/FSS/IOO.slx * https://granite.phys.s.u-tokyo.ac.jp/svn/LCGT/trunk/kagranoisebudget/FSS/IOO_params.mat * https://granite.phys.s.u-tokyo.ac.jp/svn/LCGT/trunk/kagranoisebudget/FSS/mnNb_Lib.slx |
Line 51: | Line 71: |
* Roughly measured beam profiles for Gouy telescope * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3156 * http://klog.icrr.u-tokyo.ac.jp/osl/?r=3218 |
Back to CommissioningWorkshop2017
IMC Information
- Current IMC servo topology:
- Common mode servo board configuration:
SLOW (K1:IMC-SERVO_SLOW_CON_IN1) goes to AOM (and IMC length)
- IMC sideband frequency: 13.78 MHz
- Measured IMC modulation depth: ~0.1 rad
- Measured IMC finesse: 540 +/- 70 (NOTE: 458 from cavity pole, not klogged)
- Measured IMC round-trip length: 53.3 m (Could be slightly different right now since we had re-installation work after this measurement)
- Measured IMC mode matching ratio: 86.2 % (Could be fairly worse right now since we had re-installation work after this measurement)
- Measured MCe RoC: 37.33(9) m
- Measured AOM actuation efficiency: 5.064(7) MHz/V (x2 for double-pass)
- Measured laser PZT actuation efficiency: 1.48 MHz/V (NOTE: 1.26 MHz/V using AOM, not klogged)
- Laser temperature actuation efficiency: not measured yet
- Measured mirror actuation efficiency: 17.2 MHz/V(MCe), 15.9 MHz/V(MCi), 14.9 MHz/V(MCo) @DC (could be wrong; inconsistent between measurements)
http://gwclio.icrr.u-tokyo.ac.jp/lcgtsubgroup/inoutoptics/2015/06/mce625-26.html#comments
- MCe pendulum frequency: 0.955 Hz
- Measured IMC optical gain: 18.6 V/MHz (cavity pole=5.6kHz) (NOT UP TO DATE; we changed laser power, detection optics since then)
- Reference cavity parameters: finesse 1.1e4, length 10 cm (spec), cavity pole 68 kHz
- Counts to volts, volts to counts:
- ADC: 2^16/40 counts/V
- DAC: 20/2^16 V/counts
- IMC oplev calibration:
- Calibrated IMC WFS noise level, and some noise hunting:
- Accelerometer on the PSL table:
IMC length NoiseBudget code:
- Frequency stabilization servo modeling (design):
- Roughly measured beam profiles for Gouy telescope