Differences between revisions 5 and 57 (spanning 52 versions)
Revision 5 as of 2018-04-28 16:09:19
Size: 4595
Comment:
Revision 57 as of 2018-04-28 18:56:37
Size: 4822
Comment:
Deletions are marked like this. Additions are marked like this.
Line 1: Line 1:
Describe KAGRA/Commissioning/Phase1/Operation/ShiftManual here. TEST  . <<TableOfContents(2)>>
Line 3: Line 3:
=== Control Room Overview ===
pic
== Control Room Overview ==
 {{attachment:image1.jpeg||height="172",width="430"}}
 {{attachment:IMG_4929.jpg||width="9800"}}
== How to use Guardian ==
=== Accessing to the guardian screen ===
 . In the sitemap, you can click on the link for the guardian overview screen.
 {{attachment:sitemap_guardian.png||height="172",width="430"}} <<BR>> It pops up the following screen, <<BR>> {{attachment:guardian_howto3.png||height="281",width="434"}}
Line 6: Line 11:
=== How to use Guardian ===
pics
=== Knowing the current state and requested state ===
 . {{attachment:guardian_howto4.png||height="281",width="434"}}
Line 9: Line 14:
=== How to know the status of the interferometer ===
 * Look at the monitor here. If "ON", it is in OPERATION mode. If "OFF", it is not; something is strange
=== How to request a state ===
 . Click on the gray bar in the middle.<<BR>> {{attachment:guardian_howto.png||height="298",width="433"}} <<BR>> Then, choose the desired state.<<BR>> {{attachment:guardian_howto2.png||height="299",width="434"}}
Line 12: Line 17:
screen shots == How to know the status of the interferometer ==
 . Look at one of the TV monitor in the control room. If "ON", it is in OPERATION mode.
 {{attachment:obs_state.png||height="233",width="465"}}
 <<BR>>If "OFF", check the guardian overview screen<<BR>>
 {{attachment:guardian_howto3.png||height="281",width="434"}}
 <<BR>>Check the state of LSC_MICH. First check if the request state of LSC_MICH is OPERATION. If it is not, request OPERATION.<<BR>>
 {{attachment:guardian_howto4.png||height="281",width="434"}}
 . If the state is ''READY, ENGAGE_SERVO, ENGAGE_BOOST, MICH_LOCKED, CALIBRATION_INJECTION'', or ''ACTUATORCAL_INJECTION'', the interferometer is now on the way to be OPERATION state. It will be ready soon.
 <<BR>> /!\ ''' If it does not enter OPERATION state in a few mins. Call the interferometer experts.'''<<BR>>
Line 14: Line 27:
 * If "OFF", check this screen  Check the status of ''IMC_LOCK, VIS_BS, VIS_PR2, VIS_PR3''. (Note that ''VIS_PRM'' is not used for now, so just ignore it)
 <<BR>>For IMC_LOCK, every thing should be green and IMC should be back automatically. <<BR>>
Line 16: Line 30:
screen shots  {{attachment:IMC_guardian.png||height="293",width="453"}}<<BR>>
 For example, this is how the IMC_LOCK guardian should normally look like.<<BR>>
 {{attachment:IMC_guardian_status.png||height="281",width="407"}}<<BR>>
 /!\ '''If it does not come back in ~ 10 mins, call interferometer/IOO experts.'''<<BR>>
 For VIS_PR2 and VIS_BS, the state should be ALIGNED. If inot, a notification "please reset Watch Dog!" will show up, and the state is in TRIPPED. In that case, go to the next section to know how to reset Watch Dog.
 <<BR>> For VIS_PR3, the state should be DAMPED.
 If it is not, a notification "please reset Watch Dog!" will show up, and the state is in TRIPPED. In that case, go to the next section to know how to reset Watch Dog.
Line 18: Line 38:
 1. Check the state of LSC_MICH. First check if the request state of LSC_MICH is OPERATION. If it is not, request OPERATION. == How to reset Watch Dogs ==
 0. Request LSC_MICH to be READY by Guardian.
Line 20: Line 41:
Screen shot  1. Open Sitemap (from Terminal, type "mk" and press return).
Line 22: Line 43:
 2. If the state is READY, ENGAGE_SERVO, ENGAGE_BOOST, MICH_LOCKED, CALIBRATION_INJECTION, or ACTUATORCAL_INJECTION, the interferometer is now on the way to be OPERATION state. It will be ready soon. If it will not enter OPERATION state in a few mins, something is wrong. Call interferometer experts.
  . If it is in DOWN state, go to 3.
 2. Click VIS
Line 25: Line 45:
 2. Check the status of IMC_LOCK, VIS_BS, VIS_PR2, VIS_PR3. (Note that VIS_PRM is not used for now, so just ignore it)  3. Click the suspension of which you want to reset Watch Dog
Line 27: Line 47:
 -- For IMC_LOCK, every thing should be green and IMC should be back automatically. If it will not come back in ~ 10 mins, call interferomter/IOO experts.  {{attachment:VISALL.png||width="407"}}<<BR>>
Line 29: Line 49:
 -- For VIS_PR2 and VIS_BS, the state should be ALIGNED. If it is not, a notification "please reset WatchDog!" will show up, and the state is in TRIPPED. In that case, go to the next section to know how to reset WatchDog.  4. Go to PRs case or BS case
Line 31: Line 51:
 -- For VIS_PR3, the state should be DAMPED. If it is not, a notification "please reset WatchDog!" will show up, and the state is in TRIPPED. In that case, go to the next section to know how to reset WatchDog. === PRs case ===
Line 33: Line 53:
=== basic tasks ===
 * you have to take the logbook to report the incident during the observation shift (see [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Observation/iKAGRA/Manual/MIF|IFO Operational Manual]])
 4. Request SAFE by Guardian
Line 36: Line 55:
 * you have to check the check list to avoid mistake during the observation shift ([[http://gwdoc.icrr.u-tokyo.ac.jp/DocDB/0049/T1604975/005/CheckList_20160324.pdf|Check List]])  5. Open "DETAILS" at bottom right,
Line 38: Line 57:
ここから先は観測シフトの仕事を簡単にまとめたものです。まだ暫定的なメモのような状態なので、適宜アップデートして下さい。ここは分かりやすさのために日本語で書きます。余裕のある方は英語版も作って下さい。  {{attachment:PRTOP.png||width="300"}}<<BR>>
  and you will find PRs overview screen
Line 40: Line 60:
[General]  {{attachment:PR.png||width="800"}}<<BR>>
 
 6. Click WATCHDOG
Line 42: Line 64:
 * まずコントロールルームのPCのapplication→terminal→mk→sitemapが開かれる   If something is red, press here to reset.
 {{attachment:PRWD.png||width="400"}}<<BR>>
Line 44: Line 67:
[DGS]  7. Click DACKILL (see the picture above, PRs overview sceen)
Line 46: Line 69:
 * sitemap→CDS→DAQが全部緑になっているかどうかを確認。緑じゃなければその項目のボタンを押し、statusが0×0になっているか確認。なっていなれけば宮川さんか苔山さんに連絡。   If it is not green, press reset.
 {{attachment:DACKILL.png||width="200"}}<<BR>>
Line 48: Line 72:
[MIF]  8. Click Coil Driver WD (picture above, PRs overview sceen)
Line 50: Line 74:
 * sitemap→guardian→MICH_LOCKとIMC_LOCKを確認。PSLは3月中は触らない。(IMCがロックしていればIMC_MCI_TRANS_OUTPUTは0.35くらい)  Check if everything is green. If there are red signs, push corresponding WD Pulse ON button and then OFF button
Line 52: Line 76:
[VIS]  {{attachment:PR_CDWD.png||width="800"}}<<BR>>
Line 54: Line 78:
 * sitemap→VIS→PR3の画面右側のWATCHDOG、DACKILL IOPが緑かどうかを確認。  9. Request DAMPED. For PR2, ~ 10 seconds after the state becomes DAMPED, request ALIENED.
Line 56: Line 80:
[DMG]  10. ~ 10 seconds after the suspension gets in a nominal state, request LSC_MICH to be OPERATION.
Line 58: Line 82:
 * DMG Check-Sheetを2時間おきに書く。コントロールルーム隣のサーバールームのPCを見て、Check-SheetのIn case of normalを満たしているか確認。Is the log outputs normal?は、PC画面に赤か黄のメッセージが出てこなければOK。 === BS case ===
 4. Request SAFE by Guardian
Line 60: Line 85:
[klog]  5. Open "Payload" at bottom center,
Line 62: Line 87:
 * klogに干渉計の状況を簡単に報告。  {{attachment:BSTOP.png||width="250"}}<<BR>>
  and you will find BS payload overview screen
Line 64: Line 90:
=== useful information ===
 * terminology
 {{attachment:BSpayload.png||width="800"}}<<BR>>
Line 67: Line 92:
||<#F0F0FF>Abbreviation / Acronym ||<#F0F0FF>Meaning / Explanation ||
||BS ||Beam Splitter ||
||CDS || ||
||DAQ ||Data Acquisition ||
||DAS ||Data Analysis ||
||DC ||Direct Current ||
||DET/Detchar ||Detector Characterization ||
||DGS ||Digital System ||
||DMG ||Data Management ||
||EPICS ||Experimental Physics and Industrial Control System ||
||EX/ETMX ||X Arm End Test Mass ||
||EY/ETMY ||Y Arm End Test Mass ||
||FCL ||Facility ||
||GAS ||Geometric Anti-Spring ||
||GIF ||Geophysics Interferometer ||
||Guardian ||Automation system for interferometers ||
||IFO ||Interferometer Optics ||
||IMC ||Input Mode Cleaner ||
||LSC ||Length Sensing and Control, especially for MIF ||
||MCi/e/o ||Name of three mirrors of IMC; Input, End, and Output mirrors ||
||Medm ||GUI for dealing with EPICS ||
||MI ||Michealson Interferometer ||
||MIF ||Main Interferometer ||
||OpLev ||Optical Lever; measures angle of a mirror ||
||OSEM ||Optical Sensor and ElectroMagnet ||
||PMC ||Pre-Mode Cleaner ||
||PR3 ||Type Bp’ SAS. works as the last steering mirror for MI ||
||PSL ||Pre-Stabilized Laser ||
||SAS ||Seismic Attenuation System ||
||TM ||Test Mass ||
||VAC ||Vacuum ||
||VIS ||Vibration Isolation System ||
 6. Click WATCHDOG
 If something is red, press here to reset.
 {{attachment:PRWD.png||width="400"}}<<BR>>
Line 100: Line 96:
 7. Click DACKILL (see the picture above for BS overview screen)
Line 101: Line 98:
 * how to search JGW document  If it is not green, press reset.
 {{attachment:DACKILL.png||width="200"}}<<BR>>

 8. Click Coil Driver WD (see the picture above for BS overview screen)

 Check if everything is green. If there are red signs, push corresponding WD Pulse ON button and then OFF button

 {{attachment:PR_CDWD.png||width="800"}}<<BR>>

 9. Request DAMPED. For PR2, ~ 10 seconds after the state becomes DAMPED, request ALIENED.

 10. ~ 10 seconds after the suspension gets in a nominal state, request LSC_MICH to be OPERATION.

Control Room Overview

How to use Guardian

Accessing to the guardian screen

Knowing the current state and requested state

How to request a state

How to know the status of the interferometer

How to reset Watch Dogs

  1. Request LSC_MICH to be READY by Guardian.
  2. Open Sitemap (from Terminal, type "mk" and press return).
  3. Click VIS
  4. Click the suspension of which you want to reset Watch Dog

    VISALL.png

  5. Go to PRs case or BS case

PRs case

  1. Request SAFE by Guardian
  2. Open "DETAILS" at bottom right,

    PRTOP.png

    • and you will find PRs overview screen

    PR.png

  3. Click WATCHDOG
    • If something is red, press here to reset.

    PRWD.png

  4. Click DACKILL (see the picture above, PRs overview sceen)
    • If it is not green, press reset.

    DACKILL.png

  5. Click Coil Driver WD (picture above, PRs overview sceen) Check if everything is green. If there are red signs, push corresponding WD Pulse ON button and then OFF button

    PR_CDWD.png

  6. Request DAMPED. For PR2, ~ 10 seconds after the state becomes DAMPED, request ALIENED.
  7. ~ 10 seconds after the suspension gets in a nominal state, request LSC_MICH to be OPERATION.

BS case

  1. Request SAFE by Guardian
  2. Open "Payload" at bottom center,

    BSTOP.png

    • and you will find BS payload overview screen

    BSpayload.png

  3. Click WATCHDOG If something is red, press here to reset.

    PRWD.png

  4. Click DACKILL (see the picture above for BS overview screen) If it is not green, press reset.

    DACKILL.png

  5. Click Coil Driver WD (see the picture above for BS overview screen) Check if everything is green. If there are red signs, push corresponding WD Pulse ON button and then OFF button

    PR_CDWD.png

  6. Request DAMPED. For PR2, ~ 10 seconds after the state becomes DAMPED, request ALIENED.
  7. ~ 10 seconds after the suspension gets in a nominal state, request LSC_MICH to be OPERATION.

KAGRA/Commissioning/Phase1/Operation/ShiftManual (last edited 2018-04-28 19:04:13 by YutaroEnomoto)