Differences between revisions 7 and 19 (spanning 12 versions)
Revision 7 as of 2022-11-18 13:42:23
Size: 1035
Editor: miyoki
Comment:
Revision 19 as of 2022-11-18 15:14:44
Size: 2190
Editor: miyoki
Comment:
Deletions are marked like this. Additions are marked like this.
Line 5: Line 5:
 * OMC (Aso-kun, NAOJ)
  *
 * [[ https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=14515 | OMC (Aso, NAOJ)]]
  * Higher mode rejection -> OK with good mode matching.
  * New optical board to remove additional structure for resonance dumping.
  * New high-quality mirror. -> 97% T_FP.
  * Replacement of the broken PD
  * 800 Hz resonance still remains. -> unstable lock in air. Pumping down OMC are will mitigate this resonance.
   * FP_tra estimation in air ~ 0.89 derived by subtracting HOMs power sum from the total power.
  * OMC alignment. OSTM should not be moved by pico to avoid trouble.
  * Dark noise: same btw both PDs.
Line 8: Line 15:
 * OMMT1, OMMT2, OSTM Suspension (Takahasi-san, NAOJ)
* [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=14522 | OMMT1, OMMT2, OSTM Suspension (Takahasi, NAOJ)]]
Line 10: Line 18:
  * OSEM Values   * OSEM output values
  * Space btw magnets and plates
  * Pico motion
Line 12: Line 22:
 * Stray light dumping (Akutstu-kun, NAOJ)  * [[ https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=14523 | Installation Acceptance Document for modified sub-optics in the OMM and OMC chambers (Akutstu, NAOJ) ]]
Line 14: Line 24:
  * [[ https://klog.icrr.u-tokyo.ac.jp/osl/?r=22942 | klog#22942 ]]   * [[https://klog.icrr.u-tokyo.ac.jp/osl/?r=22942 | klog#22942 ]]
Line 16: Line 26:
 * OMMT2T optical setting (Ushiba, Akutsu, Miyoki, Sato)  * OMMT2T optical setting for sensing OMMT1/SR2 motion separately (Ushiba, Akutsu, Miyoki, Sato, Takahashi)
Line 26: Line 36:
  * Clamp preparation -> OK, f300 -> metal, f800 -> elastomer,
Line 27: Line 38:
  * Trouble in a -Y side Hatch   * [[https://klog.icrr.u-tokyo.ac.jp/osl/?r=22948 | Trouble in a -Y side Hatch]]
Line 30: Line 41:
  * Leak in GV again??   * Leak in GV_srm-OMM again??
  * Slow pumping as IFI-IMMT/IMC -> Suspension Monitoring(Takahashi-san + alpha)


OMC Closing Review (2022/11/18)