Differences between revisions 40 and 45 (spanning 5 versions)
Revision 40 as of 2024-03-22 13:14:55
Size: 3337
Comment:
Revision 45 as of 2024-03-22 13:39:23
Size: 3742
Comment:
Deletions are marked like this. Additions are marked like this.
Line 9: Line 9:
   * Gaps between magnets and coils    * Gaps between magnets and coils -> MCo uppper is the minimum, but this is the best effort.
Line 14: Line 14:
   * Coil resistances and inductances
    * MCo: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28668
    *MCi/e: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28696
   * Coil resistances and inductances -> inductances shoulg refer the past data if coil did not change, or it will be measured again (Wasimi, 3/24(Monday morning)
    * MCo: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28668 -> MCo H4 looks high, it could be disconnection. It should be measured again (Wasimi, 3/24(Monday morning).
    *MCi/e: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28696 -> One of the MCi resistance is low, but low may has no problem. If it is high, it may have some disconnection.


IMC recovery(20240322)



one of templates for IMC closing by Akutsu-kun in 2021.


??? Recovery(202404??)



??? Recovery(202404??)


KAGRA/IFOIntegration2024/Recovery20240322 (last edited 2024-04-05 16:16:04 by miyoki)