Differences between revisions 74 and 81 (spanning 7 versions)
Revision 74 as of 2024-04-04 11:35:45
Size: 6228
Editor: miyoki
Comment:
Revision 81 as of 2024-04-05 11:52:39
Size: 7095
Editor: miyoki
Comment:
Deletions are marked like this. Additions are marked like this.
Line 94: Line 94:
Type-A
  * A02 Porality Check Type-A/C-Payload: OK Eddy Current Damping もOK
  * A03 Rubbing Check Type-A:
   *A
   *C
   *
  * A02 Porality Check Type-A/C-Payload
*
  * A03 Rubbing Check
   *
Line 100: Line 100:
   *A
   *C
   *
Line 104: Line 103:
   * Done: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28780
Line 107: Line 107:
  * A09: Spectrum
   *A
   *C
  * A09: Spectrum(spectra of IM/MN photosensors.)
   * OK https://klog.icrr.u-tokyo.ac.jp/osl/?r=29102
Line 111: Line 110:
   *A
   *C
   * Some GASs are strange? : https://klog.icrr.u-tokyo.ac.jp/osl/?r=29100
Line 114: Line 112:
  *   * MN: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28905
  * TM:
  * PF:
 * Green flash at REFL and Trans PD
  * REFL: OK
  * Trans: No beam on PD.

 * X-Pcal homeworks
  * Loss of power of ~ 5%: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29081
  * Cleaning: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29091

 *Windows Cleaning
  * Done: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28976,
  * Blanck ports:
Line 116: Line 128:
Line 117: Line 130:
Line 118: Line 132:
Line 119: Line 134:
  * NAB: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28757
  * WAB: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28758, https://klog.icrr.u-tokyo.ac.jp/osl/?r=28627
Line 120: Line 138:
  * tie rods adjustment between EXC and TMSX to avoid stress when the only EXC side is pumped down.
  * Replaced flange
Line 121: Line 142:
  * All temp signals are obtained?


IMC recovery(20240322)



one of templates for IMC closing by Akutsu-kun in 2021.


Summary


  • TF of MCe had a shift of peak frequency on pitch. Ushiba-kun will measure it again with the well-centered oplev. It will be reported at the recovery meeting or later.
  • All others should be OK for closing IMC chambers.
    • The resistance of the Coil of MCo H4 looked high. Washimi-kun will measure it again on Monday, but it won't stop closing anyway.
    • Washimi-kun will measure inductances again on Monday.
    • Check no dust inside, especially on the bottom before closing.
    • Check the leak points are fastened tightly when closing.


X-Pcal Recovery(20240402)


  • Dan's presentation files

    • Path2 drift is concerns. However, we expect we can compensate for the drift by using pico-mirrors.
    • 5% power loss is concerned. It should be checked on 3~4 April. --> klog#29065

    • Pcal-X noise in the Cal line is OK? --> klog#29075


EXC Recovery(202404??)



*** Recovery(202404??)



*** Recovery(202404??)


KAGRA/IFOIntegration2024/Recovery20240322 (last edited 2024-04-05 16:16:04 by miyoki)