Line 3: Line 3:
Here is a list of important channels for iKAGRA Test Run. Time is given in JST. Units are also given. Here is a list of important channels for iKAGRA Test Run. Time is given in JST. Units are in counts if not specified.
Line 8: Line 8:
|| || || || || ||K1:PSL-PMC_TRANS_DC_OUT_DQ||PMC transmitted power. If this is ~6, PMC is locked.|| Same || Same ||
Line 13: Line 13:
|| || || || || ||K1:IMC-MCL_TRANS_OUT_DQ||IMC transmitted power. If this is ~0.4, IMC is locked.|| Same || Same ||
Line 30: Line 30:
||K1:LSC-ACT_DEMOD1_(I|Q)_OUT||N/A|| N/A || Demodulated signal after(???) 111 Hz calibration line injection for ETMX actuator efficiency monitor. ||
||K1:LSC-ACT_DEMOD2_(I|Q)_OUT||N/A|| N/A || Demodulated signal after(???) 113 Hz calibration line injection for ETMY actuator efficiency monitor. ||
||K1:LSC-ACT_DEMOD3_(I|Q)_OUT||N/A|| N/A || Demodulated signal after(???) 115 Hz calibration line injection for BS actuator efficiency monitor. ||
||K1:LSC-ACT_DEMOD1_(I|Q)_OUT_DQ||N/A|| N/A || Demodulated signal after(???) 111 Hz calibration line injection for ETMX actuator efficiency monitor. ||
||K1:LSC-ACT_DEMOD2_(I|Q)_OUT_DQ||N/A|| N/A || Demodulated signal after(???) 113 Hz calibration line injection for ETMY actuator efficiency monitor. ||
||K1:LSC-ACT_DEMOD3_(I|Q)_OUT_DQ||N/A|| N/A || Demodulated signal after(???) 115 Hz calibration line injection for BS actuator efficiency monitor. ||
||K1:LSC-IMT_QPDA1_(PIT|YAW)_OUT_DQ||QPD signal monitoring IMMT1 transmitted beam, calibrated in um.|| Same || Same ||
Line 37: Line 37:
|| || || || || ||K1:VIS-$(OPTIC)_TM_OPLEV_$(DOF)_OUT_DQ||Oplev signal for suspensions in um. $(OPITC)=MCI,MCO,MCE,PR3,BS,ETMX,ETMY and $(DOF)=PIT,YAW.|| Same || Same ||

Important Channels for iKAGRA

Here is a list of important channels for iKAGRA Test Run. Time is given in JST. Units are in counts if not specified.

Channel Name

Mar 25 9:00 - Mar 31 17:00

Apr 11 9:00 - Apr 12 13:00

Apr 12 16:10 -

K1:GRD-PSL_LOCK_STATE_N

PMC guardian state. If this is 1000(???), PMC is locked.

Same

Same

K1:PSL-PMC_TRANS_DC_OUT_DQ

PMC transmitted power. If this is ~6, PMC is locked.

Same

Same

Channel Name

Mar 25 9:00 - Mar 31 17:00

Apr 11 9:00 - Apr 12 13:00

Apr 12 16:10 -

K1:GRD-IMC_LOCK_STATE_N

IMC guardian state. If this is 1000, IMC is locked.

Same

Same

K1:IMC-MCL_TRANS_OUT_DQ

IMC transmitted power. If this is ~0.4, IMC is locked.

Same

Same

Channel Name

Mar 25 9:00 - Mar 31 17:00

Apr 11 9:00 - Apr 12 13:00

Apr 12 16:10 -

K1:GRD-MICH_LOCK_STATE_N

MICH guardian state. If this is 1000, MICH is locked.

Same

Same

K1:LSC-REFL_PDA1_DC_OUT_DQ

Fringe signal in DC. Used for the error signal.

Same, but not used for the error signal.

Same

K1:LSC-REFL_PDA1_RF17_I_ERR_DQ

N/A

Fringe signal in RF (demodulated in I phase). Used for the error signal.

Same, but not used for the error signal.

K1:LSC-REFL_PDA1_RF17_Q_ERR_DQ

N/A

Fringe signal in RF (demodulated in Q phase).

Same, but used for the error signal.

K1:LSC-MICH_OUT_DQ

Feedback signal to ETMs.

Feedback signal to ETMs before gain adjustment from UGF servo.

Same

K1:LSC-UGF_SERVO_OUT_DQ

N/A

Gain adjustment of the feedback signal for UGF servo.

Same

K1:LSC-MICH_ERR_CAL_OUT_DQ

Roughly calibrated GW signal in m/rtHz.

Same

Same

K1:LSC-MICH_CTRL_CAL_OUT_DQ

N/A

Online-calibrated GW signal in nm/rtHz.

Same

K1:LSC-MICH_LKIN_CAL(1|2)_A_(REAL|IMAG)_OUT_DQ

Demodulated signal after calibration lines injection. CAL1 for 80 Hz (used for UGF servo), CAL2 for 135 Hz.

Same

N/A

K1:LSC-MICH_LKIN_CAL(1|2)_B_(REAL|IMAG)_OUT_DQ

Demodulated signal before calibration lines injection. CAL1 for 80 Hz (used ofr UGF servo), CAL2 for 135 Hz.

Same

N/A

K1:LSC-MICH_CAL_DEMOD1_(I|Q)_OUT_DQ

N/A

N/A

Demodulated signal after(???) 135 Hz(???) calibration line injection.

K1:LSC-MICH_CAL_DEMOD2_(I|Q)_OUT_DQ

N/A

N/A

Demodulated signal after(???) 80 Hz(???) calibration line injection for UGF servo.

K1:LSC-ACT_DEMOD1_(I|Q)_OUT_DQ

N/A

N/A

Demodulated signal after(???) 111 Hz calibration line injection for ETMX actuator efficiency monitor.

K1:LSC-ACT_DEMOD2_(I|Q)_OUT_DQ

N/A

N/A

Demodulated signal after(???) 113 Hz calibration line injection for ETMY actuator efficiency monitor.

K1:LSC-ACT_DEMOD3_(I|Q)_OUT_DQ

N/A

N/A

Demodulated signal after(???) 115 Hz calibration line injection for BS actuator efficiency monitor.

K1:LSC-IMT_QPDA1_(PIT|YAW)_OUT_DQ

QPD signal monitoring IMMT1 transmitted beam, calibrated in um.

Same

Same

Channel Name

Mar 25 9:00 - Mar 31 17:00

Apr 11 9:00 - Apr 12 13:00

Apr 12 16:10 -

K1:VIS-$(OPTIC)_TM_OPLEV_$(DOF)_OUT_DQ

Oplev signal for suspensions in um. $(OPITC)=MCI,MCO,MCE,PR3,BS,ETMX,ETMY and $(DOF)=PIT,YAW.

Same

Same

KAGRA/Observation/iKAGRA/ChannelList (last edited 2017-04-24 12:22:30 by YutaMichimura)