= IFO Operation Manual = == Take logbook if == * PSL PMC looses lock * IMC does not lock even if PSL PMC is locking * IMC keeps locking and unlocking for more than an hour * Main interferometer does not lock even if IMC is locked * Main interferometer does not lock for more than 20 mins * Each mirror’s oplev feedback signals go more than 6000 cts. == CHECK LIST == * Record K1:PSL-PMC_TRANS_DC_OUTPUT. * If it is ~6, OK. * If not, ask Miyakawa, Kawamura, or Enomoto to manually lock PMC. * Check the state of IMC_LOCK guardian. * If it is IMC_LOCKED, OK. Record K1:IMC-MCL_TRANS_OUTPUT. * If it is ~0.4, OK. * If not, check IMC alignment. * If not, check IMC alignment. * Record IMC oplev values below. Check if they are drifting or not.{{{ K1:VIS-MCE_TM_OPLEV_PIT_OUTPUT K1:VIS-MCE_TM_OPLEV_YAW_OUTPUT K1:VIS-MCI_TM_OPLEV_PIT_OUTPUT K1:VIS-MCI_TM_OPLEV_YAW_OUTPUT K1:VIS-MCO_TM_OPLEV_PIT_OUTPUT K1:VIS-MCO_TM_OPLEV_YAW_OUTPUT }}} * Check the state of MICH_LOCK guardian. * If it is OPERATION, OK. Check if K1:LSC-REFL_PDA1_DC_OUTPUT is fluctuating or not. * If it is kind of constant, OK. * If not, check Michelson alignment (see [[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5027|JGW-T1605027]] for alignment procedure). * If not, check Michelson alignment (see [[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5027|JGW-T1605027]] for alignment procedure). * Record oplev values for Michelson mirrors. Check if they are drifting or not.{{{ K1:VIS-PR3_TM_OPLEV_PIT_OUTPUT K1:VIS-PR3_TM_OPLEV_YAW_OUTPUT K1:VIS-BS_TM_OPLEV_PIT_OUTPUT K1:VIS-BS_TM_OPLEV_YAW_OUTPUT K1:VIS-ETMX_TM_OPLEV_PIT_OUTPUT K1:VIS-ETMX_TM_OPLEV_YAW_OUTPUT K1:VIS-ETMY_TM_OPLEV_PIT_OUTPUT K1:VIS-ETMY_TM_OPLEV_YAW_OUTPUT }}} ---- == General == * Type "mk" on terminal and you will open a "sitemap" screen == PSL (no PSL remote control in March) == === Lock PMC === * Open the "PSL PMC" medm screen from sitemap [[attachment:PMC_medm.png|{{attachment:PMC_medm.png|PMC medm|width="600"}}]] * Open StripTool /users/striptool/pmc.stp (link is put on the PMC screen) * Find PMC_TRANS Camera * Open the Guardian screen (link on the PMC screen) [[attachment:PMC_Guardian.png|{{attachment:PMC_Guardian.png|PMC Guardian|width="600"}}]] * 1. Request "PSL_LOCKED" * 2. Execute (Choose "EXEC") * Make sure that PMC is locked with the StripTool, PMC_TRANS camera, and the Guardian. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 6, PMC is locked. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 0, unlocked. == IMC == === Lock IMC === * Check MCI, MCO and MCE (link to VIS page. DC alignment values, Watchdog, Dackill) * Open IMC medm from sitemap [[attachment:スクリーンショット 2016-03-08 10.37.45.png|{{attachment:スクリーンショット 2016-03-08 10.37.45.png|IMC medm|width="600"}}]] * Open StripTools and Guardian from IMC medm [[attachment:IMC_guard.png|{{attachment:IMC_guard.png|IMC Guardian|width="600"}}]] * The usage of the IMC Guardian is almost same as one of PMC. Request "IMC_LOCKED" and execute. * NOTE: If PMC is unlocked, you will see a message "PSL_is_unlocked!" in USERMSG line in the IMC Guardian. When you see it, check if PMC is locked or not * Open StripTool /users/stiptool/IMC.stp * Find MCE_TRANS camera * Make sure IMC is locked with the StripTool, MCE_TRANS camera, and the Guardian. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.40 - 0.46 , IMC is locked. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.020, unlocked. == Main IFO == === Lock Michelson IFO === * Open /users/striptool/REFLPD_RMS.stp. You can see REFL PDA1 * Control panel for MICH length-sensing control (LSC) is sitemap => LSC => iKAGRA LSC * make sure MICH_LOCK Guardian showing "operation" [[attachment:MICH_guardian.png|{{attachment:MICH_guardian.png|MICH Guardian|width="600"}}]]]] === Check VIS === * Check PR3, BS, IX and IY (link to VIS page) === Initial Alignment === * [[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5027]] * Educate Experts === Lock Michelson IFO === * Establish the lock procedure * Prepare Guardian and put the manual here * Prepare StripTool (list of check channels) === Check DAQ === * MonitCheck Check if all the DAQ system is working properly * DaqCheck1 Check if the real time models are sending data to DAQ * DaqCheck2 Check if the GW channel is properly recorded in the frames * Check Detchar screen? * Check glitch rate (prepare a script?) * Check noise lines (prepare a script?) = In the case of IFO/subsystem failure or emergency = *