Differences between revisions 8 and 9
Revision 8 as of 2016-04-20 16:37:29
Size: 2523
Comment:
Revision 9 as of 2016-04-20 16:47:01
Size: 2551
Comment:
Deletions are marked like this. Additions are marked like this.
Line 25: Line 25:
||MCO || 12.5 (FM1 16.34) || 12.5 (FM1 15.1976)||13144 ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=569|569]] (strange; calibration not reflected?) || ||MCO || 12.5 (FM1 16.34 is turned off) || 12.5 (FM1 15.1976 is turned off)||13144 ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=569|569]] (strange; calibration not reflected?) ||

iKAGRA Oplev Calibration

Here's summary of calibration of optical levers. K1:VIS-$(OPTIC)_TM_OPLEV_(PIT|YAW)_GAIN are the calibration factors which convert counts (normalized with quadrant sum) to urad. For IMC mirrors, FM1 "cal" in K1:VIS-$(OPTIC)_TM_OPLEV_(PIT|YAW) filter bank is also used for calibration

March Test Run (Mar 25 9:00 - Mar 31 17:00)

$(OPTIC)

K1:VIS-$(OPTIC)_TM_OPLEV_PIT_GAIN

K1:VIS-$(OPTIC)_TM_OPLEV_YAW_GAIN

Quadrant sum

Related klog

EX

988 (after collimator change)

EY

1031 (after collimator change)

BS

949,952 (EX, EY values in the klog are wrong because we changed collimators)

PR3

932

MCI

569

MCO

569

MCE

569

  • Please put March values

April Test Run (Apr 11 9:00 - Apr 25 17:00)

$(OPTIC)

K1:VIS-$(OPTIC)_TM_OPLEV_PIT_GAIN

K1:VIS-$(OPTIC)_TM_OPLEV_YAW_GAIN

Quadrant sum

Related klog

EX

128.6

-131.1

1497

1436 (after folding)

EY

109.5

-103.2

10066

1374 (after folding)

BS

168.0

-195.0

12110

949,952 (EX, EY values in the klog are wrong because we changed collimators and configuration to folded)

PR3

244.9

229.4

37882

932 (quadrant sum seems low compared with the value in the klog)

MCI

1.0 (FM1 58.1395)

1.0 (FM1 56.1798)

85345

569 (strange; calibration not reflected?)

MCO

12.5 (FM1 16.34 is turned off)

12.5 (FM1 15.1976 is turned off)

13144

569 (strange; calibration not reflected?)

MCE

1.0 (FM1 29.5858)

1.0 (FM1 25.641)

21806

569 (strange; calibration not reflected?)

  • April values are noted by T. Akutsu on Apr 19, 2016
  • FM1 gains for IMC mirrors are noted by A. Shoda on Apr 20, 2016

KAGRA/Observation/iKAGRA/OplevCalibration (last edited 2016-04-20 16:47:01 by YutaMichimura)