Differences between revisions 3 and 20 (spanning 17 versions)
Revision 3 as of 2024-04-09 09:49:46
Size: 1091
Editor: miyoki
Comment:
Revision 20 as of 2024-04-12 16:02:45
Size: 2642
Comment:
Deletions are marked like this. Additions are marked like this.
Line 5: Line 5:
  * A01: Cable check
   *
  * A01: Cable check -> accept
   * Re-cabling of BF damper
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=28993
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=29177
Line 8: Line 10:
   *    * Magnet gluing
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=28831
   * According to Ushiba-kun's picture, they show that all the poraries are correct -> accept
Line 10: Line 14:
   *    * judge by the transfer functions
Line 14: Line 18:
  * A06: Height Measurement
   *
  * A06: Height Measurement -> accept
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=29064
Line 17: Line 21:
  * A09: Spectrum(spectra of IM/MN photosensors.)
   *
  * A09: Spectrum(spectra of IM/MN photosensors.) -> accept
   * ITMY IM/MN photosensors: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29140
   * ITMY tower sensor: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29141
     * Moving peaks exist, but amplitudes are not so high and frequency is far from controls. -> accept
     * 2nd and 3rd graph shows it can be removed when evacuated. -> accept, we should check it after the pump down.
Line 20: Line 27:
   *
 * Oplev Check (Yokozawa, Akutsu)
  *
 * Green flash at REFL and Trans PD
  *
   * Transsfer function : https://klog.icrr.u-tokyo.ac.jp/osl/?r=29138
    * F0 is close to the edge, so the gain is high. -> accept
    * F1 is within the range enough, we have no explanation for why the gain is so high. The reason for the dip of 8Hz depends on coupling, the earthquake might change the coupling. 1.6Hz peak showed in the new data, but we do not know why. However, no control is applied, so -> accept
   * EQ stop: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29087
    
 * Oplev Check (Yokozawa, Akutsu) -> accept
   * path check: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29098
 * Green flash at REFL and Trans PD : https://klog.icrr.u-tokyo.ac.jp/osl/?r=29135
  * Flash was seen -> accept
Line 27: Line 38:
  *    * https://klog.icrr.u-tokyo.ac.jp/osl/?r=29128 -> accept
Line 36: Line 47:
  *    * NAB: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28998 -> accept
   * WAB: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28984 -> accept
Line 42: Line 54:
  * N2 purge is necessary for EXV.   * N2 purge is necessary for IYV.


IYC Recovery(20240407)