Differences between revisions 5 and 22 (spanning 17 versions)
Revision 5 as of 2024-04-10 17:02:26
Size: 1135
Comment:
Revision 22 as of 2024-04-12 16:11:16
Size: 2775
Comment:
Deletions are marked like this. Additions are marked like this.
Line 5: Line 5:
  * A01: Cable check
   *
  * A02 Porality Check
   *
  * A01: Cable check -> accept
   * Re-cabling of BF damper
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=28993
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=29177
  * A02 Porality Check -> accept
   * Magnet gluing
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=28831
   * According to Ushiba-kun's picture, they show that all the poraries are correct -> accept
Line 10: Line 14:
   *    * judge by the transfer functions
Line 14: Line 18:
  * A06: Height Measurement
   *
  * A06: Height Measurement -> accept
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=29064
Line 17: Line 21:
  * A09: Spectrum(spectra of IM/MN photosensors.)
   *
  * A10: Transfer Function
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=29138
 * Oplev Check (Yokozawa, Akutsu)
  *
 * Green flash at REFL and Trans PD
  *
  * A09: Spectrum(spectra of IM/MN photosensors.) -> accept
   * ITMY IM/MN photosensors: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29140
   * ITMY tower sensor: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29141
     * Moving peaks exist, but amplitudes are not so high and frequency is far from controls. -> accept
     * 2nd and 3rd graph shows it can be removed when evacuated. -> accept, we should check it after the pump down.
  * A10: Transfer Function -> accept
   * Transsfer function : https://klog.icrr.u-tokyo.ac.jp/osl/?r=29138
    * F0 is close to the edge, so the gain is high. -> accept
    * F1 is within the range enough, we have no explanation for why the gain is so high. The reason for the dip of 8Hz depends on coupling, the earthquake might change the coupling. 1.6Hz peak showed in the new data, but we do not know why. However, no control is applied, so -> accept
   * EQ stop: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29087 -> accept
    
 * Oplev Check (Yokozawa, Akutsu) -> accept
   * path check: https://klog.icrr.u-tokyo.ac.jp/osl/?r=29098
 * Green flash at REFL and Trans PD : https://klog.icrr.u-tokyo.ac.jp/osl/?r=29135
  * Flash was seen -> accept
Line 26: Line 37:
 *Windows Cleaning
  *
 *Windows Cleaning -> accept
   * https://klog.icrr.u-tokyo.ac.jp/osl/?r=29128
Line 36: Line 47:
  *    * NAB: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28998 -> accept
   * WAB: https://klog.icrr.u-tokyo.ac.jp/osl/?r=28984 -> accept

-------
 * There were some discussions written in each item, but all items were accepted.
-------


IYC Recovery(20240407)



  • There were some discussions written in each item, but all items were accepted.


  • Vacuum system (Kimura-san / Uchiyama-kun and their selected persons)
    • Nasty air parts were replaced.
    • Vacuum alert recovery after pumping down.
    • GV interlock will be tried after pumping down.
    • N2 purge is necessary for IYV.
  • Cryo-system (Kimura-san and his selected persons)
    • All temp signals are obtained?
    • Cryocon alert to slack after cooling.
    • No interlock for TMP around the Cryostat.

KAGRA/Recovery2024/IYC (last edited 2024-04-15 12:31:12 by TakashiUchiyama)