3173
Comment:
|
← Revision 115 as of 2022-01-24 19:46:38 ⇥
6587
|
Deletions are marked like this. | Additions are marked like this. |
Line 2: | Line 2: |
Line 4: | Line 3: |
* FY2012: 1 pc delivered. | * FY2012: 1 pc delivered. (Just for a test) |
Line 9: | Line 8: |
* Current PRM (serial?): degrading. | * 20191218: three free-exchanges are delivered: S/N 242767, 242768, 242769. * 20220121: one new? light source (S/N 268205) and a repaired one (S/N 203248) arrived. |
Line 13: | Line 13: |
|| || || || || || 203245 || || ||IMMT1 (Mail by Uehara-san on 20180726); degrading, halved in 7200 hours.|| || 203246 || || ||IMMT2 (Mail by Uehara-san on 20180726); degrading, halved in 7200 hours.|| || 203247 || || |||| || 203248 || || ||Stocked in a box at Yend; had been used for iKAGRA-Yend. || || 203249 || || ||SR2; degrading; halved in 4800 hours.|| || '''--(203250)--''' || || ||[WAS] MCo (halved at ~8400 hours)-> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4319|Detached (20180301)]] -> Repaired one is delivered (20190411) || || 203251 || || || || || 203252 || || ||PR2; degrading. Halved in 7200 hours. || || '''--(203253)--''' || || ||[WAS] MCi (halved at ~6000 hours) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Detached (20181212)]] -> Repaired one is delivered (20190411) || || 203255 || || || EY Test Mass oplev|| || '''--(203256)--'''|| || ||[WAS] MCe (halved at ~4800 hours??) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Detached (20181212)]] -> Repaired one is delivered (20190411) || || 203258 || || ||EX Marionette oplev || || 203259 || || ||BS; degrading. Halved in 10800 hours || || 203262 || || ||Stocked at Mitaka (as of 20190729) || || '''--(203263)--''' || || ||old PR3; detached.|| |
|| || || || || ||'''--(203245)--''' || || ||IMMT1 (confirmed by Akutsu on 20190801); degrading, halved in 7200 hours. -> [[https://klog.icrr.u-tokyo.ac.jp/osl/?r=19346|Dead on 20220104]]|| ||203246 || || ||PRM (confirmed by Akutsu on 20190801); degrading, halved in 7200 hours. --> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=10586|Detached (20190918)]]; but it is just halved, so|among these sources it can be still usable??? Anyway it is now '''stocked''' tetatively.|| ||'''--(203247)--''' || || ||Stocked in a box at Xend; had been used for iKAGRA-Xend. (20190731 confirmed) --> '''Took it out''' to send back to the company according to our agreement (20200109)|| ||203248 || || ||'''Stocked''' in a box at Yend; had been used for iKAGRA-Yend. (20190731 confirmed) --> Took it back; to be sent to the manufacture (20210531) --> Repaired one came back (20220121)|| ||203249 || || ||SR2; degrading; halved in 4800 hours. || ||'''--(203250)--''' || || ||[Detached] MCo (halved at ~8400 hours)-> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4319|Detached (20180301)]] -> Repaired one is delivered (20190411) || ||'''--(203251)--''' || || ||[Detached (on 20190801) and replaced] old IMMT2; degrading... || ||'''--(203252)--''' || || ||[Detached (on 20190801) and replaced] old PR2; degrading. Halved in 7200 hours. || ||'''--(203253)--''' || || ||[Detached] MCi (halved at ~6000 hours) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Detached (20181212)]] -> Repaired one is delivered (20190411) || ||'''--(203255)--''' || || ||EY Test Mass oplev; suspect degrading. -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=12440|Detached and stocked in the center (20200109)]] || ||'''--(203256)--''' || || ||[Detached] MCe (halved at ~4800 hours??) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Detached (20181212)]] -> Repaired one is delivered (20190411) || ||'''--(203258)--''' || || ||EX Marionette oplev; suspect degrading. -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=12440|Detached and stocked in the center (20200109)]] || ||203259 || || ||BS; degrading. Halved in 10800 hours || ||203262 || || ||'''Stocked''' at Mitaka (as of 20190729) || ||'''--(203263)--''' || || ||[Detached] old PR3; detached. See [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8707|here]] and [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8706|here]] || |
Line 32: | Line 33: |
|| 225022 || || ||IY Test Mass oplev || || 228141 || || || Stocked (20190730 confirmed)|| || 228142 || || || Stocked (unpacked) at NAB@IYA (20190730 confirmed)|| || 228143 || || || Stocked (20190730 confirmed)|| || 228144 || || || Stocked (20190730); firstly I (Akutsu) happened to find it was left at the entrance of the IXC booth for some reason...|| || 228145 || || || IX Marionette oplev || || 228146 || || ||SR3; degrading. Halved in 7200 hours.|| || 228147 || || || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4319|Installed at MCo (20180301)]]. So far it is ok (20190718).|| || 228148 || || ||IX Test Mass opLev|| || 228151 || || ||EY Marionette oplev|| || 228153 || || ||IY Marionette oplev || || 228154 || || ||EX Test Mass oplev || || 228155 || || || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Installed at MCe (20181212)]]. Not so clear as that of MCi, but this is also getting degraded.|| || 228158 || || ||SRM; degrading. Halved in 7200 hours.|| || 228160 || || || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Installed at MCi (20181212)]]. Note: judging from the trend monitor, even though it was replaced with those newly purchased in FY2016, it got degraded.|| |
||225022 || || ||IY Test Mass oplev (checked by Yamada on 20190719) --> But it was installed at IX!! (checked by Akutsu on 20190719) --> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=9855|Detached on 20190807]] and stocked. || ||228141 || || ||'''Stocked''' (20190730 confirmed) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=9855|Installed at ITMX (20190807) ]]|| || '''--(228142)--''' || || ||Stocked at the field out of the box in the IYA clean booth (20190730 confirmed) -> '''Took it out''' to send back to the company according to our agreement (20200108)|| || '''--(228143)--''' || || ||Stocked (20190730 confirmed) -> '''Took it out''' to send it back to the company according to our agreement (20200108)|| ||228144 || || ||Stocked (20190730); firstly I (Akutsu) happened to find it was left at the entrance of the IXC booth for some reason... --> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=10586|Installed at PRM (20190918)]]|| ||228145 || || ||IX Marionette oplev (but Maybe IY MN oplev)-> IXC clean booth(Used for MN oplev) -> IXC area|| ||228146 || || ||SR3; degrading. Halved in 7200 hours. || ||228147 || || ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4319|Installed at MCo (20180301)]]. So far it is ok (20190718). || ||228148 || || ||IX Test Mass opLev (but may be IY TM oplev); suspect degrading! || ||228151 || || ||EY Marionette oplev -> EYC clean booth(Used for MN oplev) -> IXC area -> [[https://klog.icrr.u-tokyo.ac.jp/osl/?r=19343|Moved from IXC area to EYC TM oplev]]|| ||228153 || || ||IY Marionette oplev (but maybe IX MN oplev) -> IYC clean booth(Used for MN oplev) -> IXC area|| || '''--(228154)--''' || || ||EX Test Mass oplev; suspect degrading! --> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=12440|Detached and stocked in the center (20200109)]] || ||228155 || || ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Installed at MCe (20181212)]]. Not so clear as that of MCi, but this is also getting degraded. || ||228158 || || ||SRM; degrading. Halved in 7200 hours. || ||228160 || || ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7353|Installed at MCi (20181212)]]. Note: judging from the trend monitor, even though it was replaced with those newly purchased in FY2016, it got degraded. || |
Line 50: | Line 51: |
=== 2019 April === | |
Line 51: | Line 53: |
|| 242770 || || || || || 242771 || || || || || 242772 || || ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8712|Installed at PR3 (20190422)]] || |
||242770 || || ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=9758|Installed at IMMT2 (20190801)]] || ||242771 || || ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=9759|Installed at PR2 (20190801)]] || ||242772 || || ||[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8712|Installed at PR3 (20190422)]] || === 2019 December === The company agreed to provide some new light source for free. ||Serial# ||ICRR asset track# || ||Status || ||'''--(242767)--''' || (Free exchange)|| ||'''Stocked''' at Mozumi (20191218) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=12440|Installed at EY test mass (20200109)]]->[[https://klog.icrr.u-tokyo.ac.jp/osl/?r=19314|Moved to IXC]]-> [[https://klog.icrr.u-tokyo.ac.jp/osl/?r=19361|Sadly broken by an accident]] || ||242768 || (Free exchange)|| ||'''Stocked''' at Mozumi (20191218) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=12440|Installed at EX test mass (20200109)]] || ||242769 || (Free exchange)|| ||'''Stocked''' at Mozumi (20191218) -> [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=12440|Installed at EX marionette (20200109)]] -> IXC area -> [[https://klog.icrr.u-tokyo.ac.jp/osl/?r=19346|Moved from IXC area to IMMT1 TM oplev]] || == FY2021 == === 2022 Jan === ||Serial# ||ICRR asset track# || ||Status || ||268205 || Free delivered || ||'''Stocked''' at Mozumi (20220121) || |
Light sources for oplevs
Superlum: SLD-MS-261-MP1-FC/APC-PMF
- FY2012: 1 pc delivered. (Just for a test)
- FY2013: 15 pcs delivered.
- FY2016: 15 pcs delivered.
- Note: after repaired, it seems the company gave new serial numbers. So the old serial numbers will become unused.
- 20190411: renewed three are delivered: S/N 242770, 242771, 242772.
- 20191218: three free-exchanges are delivered: S/N 242767, 242768, 242769.
- 20220121: one new? light source (S/N 268205) and a repaired one (S/N 203248) arrived.
FY2013
Serial# |
ICRR asset track# |
|
Status |
|
|
|
|
203245 |
|
|
IMMT1 (confirmed by Akutsu on 20190801); degrading, halved in 7200 hours. -> Dead on 20220104 |
203246 |
|
|
PRM (confirmed by Akutsu on 20190801); degrading, halved in 7200 hours. --> Detached (20190918); but it is just halved, so|among these sources it can be still usable??? Anyway it is now stocked tetatively. |
203247 |
|
|
Stocked in a box at Xend; had been used for iKAGRA-Xend. (20190731 confirmed) --> Took it out to send back to the company according to our agreement (20200109) |
203248 |
|
|
Stocked in a box at Yend; had been used for iKAGRA-Yend. (20190731 confirmed) --> Took it back; to be sent to the manufacture (20210531) --> Repaired one came back (20220121) |
203249 |
|
|
SR2; degrading; halved in 4800 hours. |
203250 |
|
|
[Detached] MCo (halved at ~8400 hours)-> Detached (20180301) -> Repaired one is delivered (20190411) |
203251 |
|
|
[Detached (on 20190801) and replaced] old IMMT2; degrading... |
203252 |
|
|
[Detached (on 20190801) and replaced] old PR2; degrading. Halved in 7200 hours. |
203253 |
|
|
[Detached] MCi (halved at ~6000 hours) -> Detached (20181212) -> Repaired one is delivered (20190411) |
203255 |
|
|
EY Test Mass oplev; suspect degrading. -> Detached and stocked in the center (20200109) |
203256 |
|
|
[Detached] MCe (halved at ~4800 hours??) -> Detached (20181212) -> Repaired one is delivered (20190411) |
203258 |
|
|
EX Marionette oplev; suspect degrading. -> Detached and stocked in the center (20200109) |
203259 |
|
|
BS; degrading. Halved in 10800 hours |
203262 |
|
|
Stocked at Mitaka (as of 20190729) |
203263 |
|
|
FY2016
Serial# |
ICRR asset track# |
|
Status |
225022 |
|
|
IY Test Mass oplev (checked by Yamada on 20190719) --> But it was installed at IX!! (checked by Akutsu on 20190719) --> Detached on 20190807 and stocked. |
228141 |
|
|
Stocked (20190730 confirmed) -> Installed at ITMX (20190807) |
228142 |
|
|
Stocked at the field out of the box in the IYA clean booth (20190730 confirmed) -> Took it out to send back to the company according to our agreement (20200108) |
228143 |
|
|
Stocked (20190730 confirmed) -> Took it out to send it back to the company according to our agreement (20200108) |
228144 |
|
|
Stocked (20190730); firstly I (Akutsu) happened to find it was left at the entrance of the IXC booth for some reason... --> Installed at PRM (20190918) |
228145 |
|
|
IX Marionette oplev (but Maybe IY MN oplev)-> IXC clean booth(Used for MN oplev) -> IXC area |
228146 |
|
|
SR3; degrading. Halved in 7200 hours. |
228147 |
|
|
Installed at MCo (20180301). So far it is ok (20190718). |
228148 |
|
|
IX Test Mass opLev (but may be IY TM oplev); suspect degrading! |
228151 |
|
|
EY Marionette oplev -> EYC clean booth(Used for MN oplev) -> IXC area -> Moved from IXC area to EYC TM oplev |
228153 |
|
|
IY Marionette oplev (but maybe IX MN oplev) -> IYC clean booth(Used for MN oplev) -> IXC area |
228154 |
|
|
EX Test Mass oplev; suspect degrading! --> Detached and stocked in the center (20200109) |
228155 |
|
|
Installed at MCe (20181212). Not so clear as that of MCi, but this is also getting degraded. |
228158 |
|
|
SRM; degrading. Halved in 7200 hours. |
228160 |
|
|
Installed at MCi (20181212). Note: judging from the trend monitor, even though it was replaced with those newly purchased in FY2016, it got degraded. |
FY2019
2019 April
Serial# |
ICRR asset track# |
|
Status |
242770 |
|
|
|
242771 |
|
|
|
242772 |
|
|
2019 December
The company agreed to provide some new light source for free.
Serial# |
ICRR asset track# |
|
Status |
242767 |
(Free exchange) |
|
Stocked at Mozumi (20191218) -> Installed at EY test mass (20200109)->Moved to IXC-> Sadly broken by an accident |
242768 |
(Free exchange) |
|
Stocked at Mozumi (20191218) -> Installed at EX test mass (20200109) |
242769 |
(Free exchange) |
|
Stocked at Mozumi (20191218) -> Installed at EX marionette (20200109) -> IXC area -> Moved from IXC area to IMMT1 TM oplev |
FY2021
2022 Jan
Serial# |
ICRR asset track# |
|
Status |
268205 |
Free delivered |
|
Stocked at Mozumi (20220121) |