Differences between revisions 7 and 8
Revision 7 as of 2018-06-23 22:40:31
Size: 1082
Editor: YukiInoue
Comment:
Revision 8 as of 2018-06-23 22:40:55
Size: 958
Editor: YukiInoue
Comment:
Deletions are marked like this. Additions are marked like this.
Line 5: Line 5:
=== Systematic error estimation by Free swinging Michelson === === Free swinging Michelson ===
Line 7: Line 7:
=== Systematic error estimation by Photon calibrator === === Photon calibrator ===
Line 9: Line 9:
=== Systematic error estimation by Gravity field calibrator === === Gravity field calibrator ===
Line 11: Line 11:
=== Systematic error estimation by Gcal + Pcal === === Gcal + Pcal ===

Verification of Calibration

Comparison of calibration method

Free swinging Michelson

Photon calibrator

Gravity field calibrator

Gcal + Pcal

Differential-mode and Common-mode Actuation

  • Normally, the differential length response of the detector is calibrated using one Pcal system, varying the length of only one arm. The KAGRA interferometers, however, have Pcal systems installed at both end stations. They can be used simultaneously to produce either pure differential arm length variations, where the two arms of the interferometer stretch and contract out of phase or pure common arm-length variations, where the arms stretch and contract in phase. Comparing differential and common excitations, enables diagnosing systematic differences between the two arms and quantifying the coupling between common-arm motion and differential-arm motion.

Measuring Time Delays and Signs

KAGRA/Subgroups/CAL/Pcal_verification (last edited 2018-12-26 23:52:16 by YukiInoue)