* PRFPMI with OMC DC readout * 1.2 W output from IMC (7 W when O3GK) * Noise hunting * (ongoing) MICH/PRCL feed forward tuning * (ongoing) Turning off the instruments (Especially vacuum pumps(TMP)) * Investigating the noise from the vibration * OMC leg * IMC * IFI * etc * We opened the IFI chamber and remove the material which was connected between IFI chamber and IFI optical table in vac * Motivation was to extract the thermal from PRM reflection beam * After this work, the sensitivity is improved * Latest sensitivity [[https://dac.icrr.u-tokyo.ac.jp/KAGRA/DAWG/Detchar/klog/klog24301|wiki]] * Silent run * Latest one [[ https://klog.icrr.u-tokyo.ac.jp/osl/?r=24301|klog24301]] * Future work * Evaluate the noise budget * Turned off the instruments * Check the scattering light * Hammering test/Acustic injection * Suspension control optimization * Alignment sensing and control * ADS and ASC * Interferometer characterization * Increased the laser power