Line 7: Line 7:
|| '''Sensor''' || '''Use''' || '''Spec''' ||
||<|2>[[https://www.vaisala.com/sites/default/files/documents/HMP155_User_Guide_in_English.pdf|(Manual)]] HMP155(Vaisala) || Temperature || ||
|| Humidity || ||
|| [[https://www.vaisala.com/sites/default/files/documents/PTB110%20User%20Guide%20in%20English.pdf|(Manual)]] PTB110(Vaisala) || Pressure || ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7554|(Manual)]] Trillium 120QA|| seismic || ||
|| [[https://www-iuem.univ-brest.fr/pops/attachments/1489|(Manual)]] Trillium Compact || seismic || ||
|| [[https://www.passcal.nmt.edu/webfm_send/220|(Manual)]] Trillium 240 || seismic || ||
|| [[https://www.guralp.com/documents/MAN-030-0001.pdf|(Manual)]] CMG3T || seismic ||  ||
|| '''Sensor''' || '''Use''' || '''Spec''' || '''DAQ''' ||
||<|2>[[https://www.vaisala.com/sites/default/files/documents/HMP155_User_Guide_in_English.pdf|(Manual)]] HMP155(Vaisala) || Temperature || || cRIO, KAGRA CDS ||
|| Humidity ||  || cRIO, KAGRA CDS ||
|| [[https://www.vaisala.com/sites/default/files/documents/PTB110%20User%20Guide%20in%20English.pdf|(Manual)]] PTB110(Vaisala) || Pressure || cRIO, KAGRA CDS ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7554|(Manual)]] Trillium 120QA|| seismic || KAGRA CDS ||
|| [[https://www-iuem.univ-brest.fr/pops/attachments/1489|(Manual)]] Trillium Compact || seismic || || KAGRA CDS ||
|| [[https://www.passcal.nmt.edu/webfm_send/220|(Manual)]] Trillium 240 || seismic ||  || cRIO ||
|| [[https://www.guralp.com/documents/MAN-030-0001.pdf|(Manual)]] CMG3T || seismic || cRIO ||

Environmental Sensors

Overview

  • {*} ひずみ計に対する各センサーをプロジェクションしたASDを載せる。

Sensors List

Sensor

Use

Spec

DAQ

(Manual) HMP155(Vaisala)

Temperature

cRIO, KAGRA CDS

Humidity

cRIO, KAGRA CDS

(Manual) PTB110(Vaisala)

Pressure

cRIO, KAGRA CDS

(Manual) Trillium 120QA

seismic

KAGRA CDS

(Manual) Trillium Compact

seismic

KAGRA CDS

(Manual) Trillium 240

seismic

cRIO

(Manual) CMG3T

seismic

cRIO

Deployment

  • {*} 配置図を載せる。

Circuit

  • {*} アンプの回路図、Dsubの配線図等のJGWのリンクを載せる。

Calibration

cRIO01 500m, y[V]がADの電圧値。上からAD00,01,02,03

t[℃]=(y[V]/10/(1+3300/430)/1.e-3-107.79)/.388+20.0
h[%]=-y[V]/5*100/2
p[hPa]=-y[V]/2/5*(1100-800)+800
vac[Pa]=10^[2*y[V]-13.0]

cRIO02 2000m

t[℃]=(y[V]/10/(1+3300/430)/1.e-3-107.79)/.388+20.0
h[%]=y[V]/5*100/2
p[hPa]=y[V]/2/5*(1100-800)+800
vac[Pa]=10^[2*y[V]-13.0]