|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7554|(Manual)]] Trillium 120QA|| seismic || || KAGRA CDS ||<|2>Seismic noise on the end and corner station || | || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7554|(Manual)]] Trillium 120QA|| seismic || || KAGRA CDS ||<|2>Seismic noise on the '''both ends and Center''' station || |
Line 13: | Line 13: |
|| [[https://www.passcal.nmt.edu/webfm_send/220|(Manual)]] Trillium 240 || seismic || || cRIO ||<|2>Seismic noise on the Arm || | || [[https://www.passcal.nmt.edu/webfm_send/220|(Manual)]] Trillium 240 || seismic || || cRIO ||<|2>Seismic noise on the '''X Arm''' || |
Environmental Sensors
Overview
ひずみ計に対する各センサーをプロジェクションしたASDを載せる。
Sensors List
Sensor |
Use |
Spec |
DAQ1 |
Purpose |
(Manual) HMP155(Vaisala) |
Temperature |
|
cRIO, KAGRA CDS |
Air Envrionment |
Humidity |
|
cRIO, KAGRA CDS |
||
(Manual) PTB110(Vaisala) |
Pressure |
|
cRIO, KAGRA CDS |
|
(Manual) Trillium 120QA |
seismic |
|
KAGRA CDS |
Seismic noise on the both ends and Center station |
(Manual) Trillium Compact |
seismic |
|
KAGRA CDS |
|
(Manual) Trillium 240 |
seismic |
|
cRIO |
Seismic noise on the X Arm |
(Manual) CMG3T |
seismic |
|
cRIO |
Deployment
配置図を載せる。
Circuit
アンプの回路図、Dsubの配線図等のJGWのリンクを載せる。
Calibration
cRIO01 500m, y[V]がADの電圧値。上からAD00,01,02,03
t[℃]=(y[V]/10/(1+3300/430)/1.e-3-107.79)/.388+20.0 h[%]=-y[V]/5*100/2 p[hPa]=-y[V]/2/5*(1100-800)+800 vac[Pa]=10^[2*y[V]-13.0]
cRIO02 2000m
t[℃]=(y[V]/10/(1+3300/430)/1.e-3-107.79)/.388+20.0 h[%]=y[V]/5*100/2 p[hPa]=y[V]/2/5*(1100-800)+800 vac[Pa]=10^[2*y[V]-13.0]
Reference