Size: 257
Comment:
|
Size: 1311
Comment:
|
Deletions are marked like this. | Additions are marked like this. |
Line 3: | Line 3: |
* 2016/4/19 | === 2016/4/19 === |
Line 6: | Line 6: |
[[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5256|presentation file]] [[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5257|supplementary file]] |
|
Line 7: | Line 9: |
* 2016/6/10 | === 2016/6/10 === |
Line 10: | Line 12: |
[[attachment:Somiya150610part2v2.pptx]] | |
Line 11: | Line 14: |
* 2016/6/24 | === 2016/6/24 === |
Line 14: | Line 17: |
[[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5313|presentation file]] | |
Line 15: | Line 19: |
* 2016/8/26 | === 2016/8/26 === |
Line 18: | Line 22: |
[[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5564|presentation file]] | |
Line 19: | Line 24: |
* 2016/9/5 | === 2016/9/5 Telecon with Yamamoto-san === |
Line 21: | Line 26: |
Telecon with Yamamoto-san | * It is reasonable that the output field is more sensitive to the OMC tilt with the BS map added. * We will not need an FFT calculation as the FINESSE simulation seems to converge. * Additional optics (ITM or OFI) may increase the requirement but will be a factor so we can proceed to design suspensions with the current requirement. |
Line 26: | Line 33: |
* A blade spring for BRT suspension will be useful for the OMC suspension which only needs a single-stage vertical isolation at around 1Hz. * Requirements to OMMTs and OFI should be also given to Akutsu-kun in m/rtHz. |
Meetings
2016/4/19
- OMC seminar at Hongo
2016/6/10
- OMC seminar at Tokyo Tech
2016/6/24
- OFI telecon
2016/8/26
- F2F meeting
2016/9/5 Telecon with Yamamoto-san
- It is reasonable that the output field is more sensitive to the OMC tilt with the BS map added.
- We will not need an FFT calculation as the FINESSE simulation seems to converge.
- Additional optics (ITM or OFI) may increase the requirement but will be a factor so we can proceed to design suspensions with the current requirement.
* 2016/9/14
- Telecon with IOO members
- A blade spring for BRT suspension will be useful for the OMC suspension which only needs a single-stage vertical isolation at around 1Hz.
- Requirements to OMMTs and OFI should be also given to Akutsu-kun in m/rtHz.