Angular Sensing and Control (Oct, 2022)

WFS ports

REFL (17, 45 MHz)

POP (17, 45 MHz)

AS (17, 28 MHz)

DC QPD ports

TMSX

TMSY

POP Forward

  • There are two DC QPDs on POP Forward, used for the initial alignment

DC PDs

  • There are POP_P and POP_S DC PDs on POP.

WFS Commissioning

WFS DC centering

  • Open the REFL shutter for REFL WFS
  • Beam spots on the quadrant diodes must be centered (DC centering)
    • PZTs and picomotors are available

Single Y arm

Single X arm

PRMI

PRFPMI

DRFPMI

Next Steps

Infrastructure

  • Guardian implementation
    • DC centering loops on REFL, POP, and AS
    • Single Y, X arms, and PRMI guardstates for initial alignment
  • Add trigger for the main ASC switch (K1:ASC-WFS_GAIN) similarly to the IMC switch (K1:IMC-WFS_GAIN)

Simulation

  • Compare the measured sensing matrices against the simulation
    • Koyama-kun's simulation GUI tool is available on the workstations
      • sitemap > Commissioning Top > finesse GUI

Integration

  • Implement single-Y, single-X, and PRMI controls for the initial alignment
    • If the alignment is too off, first ADS then WFS control?
  • PRFPMI/DRFPMI angular controls
  • Noise budget

Known issues

  • TMSX PIT-YAW coupling
    • The beam is slightly not centered in yaw on the second lens on the in-air table. Adjusting the beam position on the lens might fix the problem.