Differences between revisions 209 and 273 (spanning 64 versions)
Revision 209 as of 2018-12-26 18:38:40
Size: 15018
Editor: KousekiMiyo
Comment:
Revision 273 as of 2018-12-30 09:13:28
Size: 14941
Comment:
Deletions are marked like this. Additions are marked like this.
Line 8: Line 8:
== PSL(-2018/12/10-) == == Accelerometer ==
Line 15: Line 15:
|| --( magnetometer)-- || MAG13 MCL100 || --( K1:PEM-PSL_MAG_LASERRACK_(X|Y|Z) )-- || --( 2048 )-- || [[attachment:PEM_pic_MAG_pslrack.pdf|pic]] || T || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7535|MAG13]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8181|WF]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8519|Distributor]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8518|Power]] removed temporary||
|| microphone || (B&K) 4188-A-021 || K1:PEM-PSL_MIC_CENTER || 2048 || [[attachment:PEM_pic_MIC_center.pdf|pic]] || Pa || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8142|MIC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8142|amp]]||
Line 19: Line 17:
|| accelerometer || TEAC706 || K1:PEM-IMC_ACC_REFL_TABLE || 2048 || [[attachment:PEM_pic_ACC_IMC_refl.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9062|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7649|amp]]||
|| accelerometer || TEAC710 || K1:PEM-POP_ACC_TABLE || 2048 || [[attachment:PEM_pic_ACC_MIC_POP.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9326|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9330|amp]]||
|| accelerometer || TEAC710 || K1:PEM-POS_ACC_TABLE || 2048 || [[attachment:PEM_pic_ACC_POS.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9325|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9329|amp]]||
|| accelerometer || TEAC710 || K1:PEM-OMC_ACC_TABLE || 2048 || [[attachment:PEM_pic_ACC_TEAC_OMC.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9327|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9331|amp]]||
Line 20: Line 22:
== CS(-2018/12/10-) ==
|| PEM || Sensor || Channel name (_OUT_DQ) || sampling rate[Hz || Picture || Calibration(unit) || Memo(S number) ||
== Microphone ==
|| microphone || (B&K) 4188-A-021 || K1:PEM-PSL_MIC_CENTER || 2048 || [[attachment:PEM_pic_MIC_center.pdf|pic]] || Pa || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8142|MIC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8142|amp]]||
Line 23: Line 25:
|| accelerometer || TEAC706 || K1:PEM-IMC_ACC_REFL_TABLE || 2048 || [[attachment:PEM_pic_ACC_IMC_refl.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9062|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7649|amp]]||
Line 26: Line 27:
|| accelerometer || TEAC710 || K1:PEM-IFI_ACC_REFL_TABLE || 2048 || [[attachment:PEM_pic_ACC_MIC_IFI.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9328|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9332|amp]]||
Line 28: Line 28:
|| accelerometer || TEAC710 || K1:PEM-POP_ACC_TABLE || 2048 || [[attachment:PEM_pic_ACC_MIC_POP.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9326|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9330|amp]]||
Line 30: Line 29:
|| accelerometer || TEAC710 || K1:PEM-POS_ACC_TABLE || 2048 || [[attachment:PEM_pic_ACC_POS.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9325|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9329|amp]]||
Line 32: Line 30:
|| accelerometer || TEAC710 || K1:PEM-OMC_ACC_TABLE || 2048 || [[attachment:PEM_pic_ACC_TEAC_OMC.pdf|pic]] || m/s^2 || [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9327|ACC]], [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9331|amp]]||
Line 36: Line 33:
== EX(-201X/XX/XX) ==
 * PCal Tx table (TEAC710)
 * PCal Rx table (TEAC710)
 * EXA area (ACO mic)
 * EXC/EXV area (ACO mic)
 * EXV talbe (TEAC706)
 * EXC area (magnetometer)

== EY(-201X/XX/XX) ==
 * PCal Tx table (TEAC710)
 * PCal Rx table (TEAC710)
 * EYA area (ACO mic)
 * EYC/EYV area (ACO mic)
 * EYV talbe (TEAC706)
 * EYC area (magnetometer)
Line 54: Line 36:
||<#F0F0FF> Snumber ||<#F0F0FF>Installed Place (ADC port) ||<#F0F0FF> Name ||<#F0F0FF> Channel name (OUT_DQ) ||<#F0F0FF> Unit ||<#F0F0FF> fs [Hz] ||<#F0F0FF> klog ||
||[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7552|JGW-S1707552]] || EXV (IX1, ADC0, 24-26) || Trillium 12QA || K1:PEM-IXV_GND_TR120Q_(X/Y/Z)_OUT_DQ || um/s || 512|| [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
||[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7551|JGW-S1707551]] || EXV (EX1, ADC0, 24-26) || Trillium 120QA || K1:PEM-EXV_GND_TR120Q_(X/Y/Z)_OUT_DQ || um/s || 512 || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
||--([[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7553|JGW-S1707553]])-- || --(EYV (EY1, ADC0, 24-26) )-- || --(Trillium 120QA)-- || --(K1:PEM-EYV_GND_TR120Q_(X/Y/Z)_OUT_DQ)-- || --(um/s)-- || 512 || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
||[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7553|JGW-S1707553]] || IXV (IX1, ADC0, 28-30) || Trillium 120QA|| K1:PEM-IXV_GND_TR120QTEST_(X/Y/Z)_OUT_DQ || um/s || 512 || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=xxxx|JGW-S170xxxx]] || MCe (IOO1, ADC1, 4-6) || Trillium compact 120-SV1 || K1:PEM-IMC_GND_TR120C_MCE_(X/Y/Z)_OUT_DQ || um/s || 512 || ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=xxxx|JGW-S170xxxx]] || MCi (IOO1, ADC1, 0-2) || Trillium compact 120-SV1|| K1:PEM-IMC_GND_TR120C_MCI_(X/Y/Z)_OUT_DQ || um/s || 512 || ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7528|JGW-S1707528]] || Fukuoka Univ. || Trillium compact 120-SV1|| ??? || --- || 512|| ||
||<#F0F0FF> Snumber ||<#F0F0FF>Installed ADC port ||<#F0F0FF> Name ||<#F0F0FF> Channel name (OUT_DQ) ||<#F0F0FF> Unit ||<#F0F0FF> fs [Hz] ||<#F0F0FF> klog ||
||[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7552|JGW-S1707552]] || IX1, ADC0, 24-26 || Trillium 12QA || K1:PEM-IXV_GND_TR120Q_(X/Y/Z)_OUT_DQ || um/s || 512|| [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
||[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7551|JGW-S1707551]] || EX1, ADC0, 24-26 || Trillium 120QA || K1:PEM-EXV_GND_TR120Q_(X/Y/Z)_OUT_DQ || um/s || 512 || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
||--([[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7553|JGW-S1707553]])-- || --(EY1, ADC0, 24-26)-- || --(Trillium 120QA)-- || --(K1:PEM-EYV_GND_TR120Q_(X/Y/Z)_OUT_DQ)-- || --(um/s)-- || 512 || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
||[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7553|JGW-S1707553]] || IX1, ADC0, 28-30 || Trillium 120QA|| K1:PEM-IXV_GND_TR120QTEST_(X/Y/Z)_OUT_DQ || um/s || 512 || [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7186|7186]] ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=xxxx|JGW-S170xxxx]] || IOO1, ADC1, 4-6 || Trillium compact 120-SV1 || K1:PEM-IMC_GND_TR120C_MCE_(X/Y/Z)_OUT_DQ || um/s || 512 || ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=xxxx|JGW-S170xxxx]] || IOO1, ADC1, 0-2 || Trillium compact 120-SV1|| K1:PEM-IMC_GND_TR120C_MCI_(X/Y/Z)_OUT_DQ || um/s || 512 || ||
|| [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=7528|JGW-S1707528]] || Fukuoka Univ. || Trillium compact 120-SV1|| --- || --- || 512|| ||
Line 97: Line 79:
== EX(-201X/XX/XX) ==
 * PCal Tx table (TEAC710)
 * PCal Rx table (TEAC710)
 * EXA area (ACO mic)
 * EXC/EXV area (ACO mic)
 * EXV talbe (TEAC706)
 * EXC area (magnetometer)

== EY(-201X/XX/XX) ==
 * PCal Tx table (TEAC710)
 * PCal Rx table (TEAC710)
 * EYA area (ACO mic)
 * EYC/EYV area (ACO mic)
 * EYV talbe (TEAC706)
 * EYC area (magnetometer)


---------------
Line 98: Line 99:
Definition of the PEM channel name
Line 100: Line 102:
K1:PEM-(SENSOR)_(PLACE1)_(PLACE2)_(DOF) K1:PEM-(SENSOR)_(AREA)_(LOCATION)_(DOF)
Line 103: Line 105:
== SENSOR ==
||<#FBEFF8> Name ||<#FBEFF8> Purpose ||
|| TEMPERATURE || Ondotori temperature ||
|| HUMIDITY || Ondotori humidity ||
||<#58FA82> Name ||<#58FA82> Purpose ||
||<#A9F5E1 :99% -2> SENSOR ||
|| ONDTR || Ondotori sensor ||
Line 112: Line 113:
|| SENSOR(or SIGNAL; すぐに決めてどちらかにする) || -- ||

 * 現状(2018/12/26)は以上だが、今後PEMが足されるにつれて追加する。(yokozawa)
 * オンドトリのTEMPERATUREと、気象計(Weather Station)のTEMPが被る。

== PLACE1 ==
||<#FBEFF8> Name ||<#FBEFF8> Purpose ||
|| SENSOR || This name is not for specific sensor ||
||<#A9F5E1 :99% -2> AREA ||
Line 120: Line 116:
|| IXC || around the IXC area ||
|| IYC || around the IYC area ||
|| EXC || around the EXC area ||
|| EYC || around the EYC area ||
|| IX || around the IX area ||
|| IY || around the IY area ||
|| EX || around the EX area ||
|| EY || around the EY area ||
Line 132: Line 128:
|| PR|| around the PR area? ||
|| BS|| around the BS area? ||
|| SR|| around the SR area? ||
|| RACK || in the DGS Rack ||
|| IX0|| around the IX0 DGS rack ||
|| IY0|| around the IY0 DGS rack ||
|| IMC|| around the MCi chamber?||
|| IFI|| around the IFI chamber? ||
|| OMC|| around the OMC chamber? ||
|| POP|| around the POP table? ||
|| POS|| around the POS table? ||
|| FIELD || on the field?||
|| BOOTH|| in the "any" clean booth ||
|| PR|| around the PR area including chamber||
|| BS|| around the BS area including chamber||
|| SR|| around the SR area including chamber||
|| IMC|| around the MCi, MCe, MCo area||
|| IFI|| around the IFI area including table, chamber ||
|| OMC|| around the OMC area including table, chamber ||
|| POP|| around the POP area including table, chamber ||
|| POS|| around the POS area including talbe, chamber ||
|| --(RACK)-- || --(in the DGS Rack)-- ||
|| --(FIELD)-- || --(on the field)-- ||
|| --(BOOTH)-- || --(in the "any" clean booth)-- ||
Line 146: Line 140:
|| XARM || x arm ||
|| YARM || y arm ||
||<#A9F5E1 :99% -2> LOCATION ||
|| CHAMBER_(NAME) || on the chamber. Name : [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=1111| here]] ||
|| TABLE_(NAME) || on the table. Name : [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Subgroups/IOO/OutputTables| here]] ||
|| BOOTH_(NAME) || in the clean booth. Name : [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=1662|here?]] ||
|| BEAMDUCT_(NAME) || on the beamduct. Name : [[http://google.com|???]] ||
|| PERI_(NAME) || on the periscope. Name : [[http://google.com|???]] ||
|| FIELD || It means other places. ||
||<#A9F5E1 :99% -2> DOF ||
|| X/Y/Z || X,Y,Z axis as KAGRA cordinate ||
|| 18V/24V || type of the DC power supply ||
|| TEMPERATURE/HUMIDITY/PRESSSURE || signals of the both ondotori and weather station ||
Line 147: Line 154:
  * (A chamberを加えるかどうかは議論が必要だが、今の所置く余裕がない)
Line 149: Line 155:
== PLACE2 ==
||<#FBEFF8> Name ||<#FBEFF8> Purpose ||
|| GND || on the ground ||
|| PRERI || on the periscope ||
|| CHAMBER || on the chamber ||
|| BEAMDUCT || on the beamduct ||
|| TABLE || on the table ||
|| ONTABLE || Microphone on the table??? ||
 * SENSOR
  * PEMが管理しているセンサーの種類がここに該当する。
 * AREA
  * エリア名が該当する。
  * 三代木さんが定義。[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=1110| here]]
 * LOCATION
  * センサーが設置されている構造物の名前が該当する。たとえば、クリーンブースや光学定盤の名前。
  * クリーンブースは三代木さんが定義。[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=1662|here]]
  * 光学定盤はIOOグループが定義。[[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Subgroups/IOO/OutputTables| here]]
 * DOF
  * 1つのセンサーが複数個の出力信号を持つ場合、それぞれの名前が該当する。
Line 158: Line 167:
  * No specific list
  * GND, TABLE, PERI(PERISCOPE), CHAMBER, BEAMDUCT, ONTABLE(MIC), etc

== DOF ==
||<#FBEFF8> Name ||<#FBEFF8> Purpose ||
|| X/Y/Z || X,Y,Z axis as KAGRA cordinate ||
|| 1/2/3 || ??? ||
|| 18V/24V || type of the DC power supply ||
|| TEMP/HUMI/PRES || signals of the weather station ||
---------------

KAGRA PEM Channel List

Ondotori

Accelerometer

PEM

Sensor

Channel name (_OUT_DQ)

sampling rate[Hz]

Picture

Calibration(unit)

Memo(S number)

accelerometer

TEAC 710

K1:PEM-PSL_ACC_TABLE1_Z

2048

pic

m/s^2

ACC, Amp

accelerometer

TEAC 710

K1:PEM-PSL_ACC_TABLE2_Z

2048

pic

m/s^2

ACC, Amp

accelerometer

TEAC 710

K1:PEM-PSL_ACC_TABLE3_Z

2048

pic

m/s^2

ACC, Amp

accelerometer

KISTLER 8640A5

K1:PEM-PSL_ACC_PERI_EXIT

2048

pic

m/s^2

ACC, Amp

accelerometer

KISTLER 8640A5

K1:PEM-PSL_ACC_PERI_REFCAV

2048

pic

m/s^2

ACC, Amp

accelerometer

KISTLER

K1:PEM-PSL_ACC_SIGNAL1

2048

pic

m/s^2

ACC, Amp, portable accelerometer

accelerometer

KISTLER

K1:PEM-PSL_ACC_SIGNAL2

2048

pic

m/s^2

ACC, Amp, portable accelerometer

accelerometer

TEAC706

K1:PEM-IMC_ACC_REFL_TABLE

2048

pic

m/s^2

ACC, amp

accelerometer

TEAC710

K1:PEM-POP_ACC_TABLE

2048

pic

m/s^2

ACC, amp

accelerometer

TEAC710

K1:PEM-POS_ACC_TABLE

2048

pic

m/s^2

ACC, amp

accelerometer

TEAC710

K1:PEM-OMC_ACC_TABLE

2048

pic

m/s^2

ACC, amp

Microphone

microphone

(B&K) 4188-A-021

K1:PEM-PSL_MIC_CENTER

2048

pic

Pa

MIC, amp

microphone

AT-VD6 No.1

K1:PEM-IMC_MIC_REFL

2048

pic

Pa

MIC, amp

microphone

AT-VD6 No.2

K1:PEM-IMC_MIC_TRANS

2048

[[|pic]]

Pa

MIC, amp

microphone

(B&K) 4188-A-021

K1:PEM-IFI_MIC_REFL

2048

pic

Pa

MIC, amp

microphone

(B&K) 4188-A-021

K1:PEM-POP_MIC

2048

pic

Pa

MIC, amp

microphone

ACO 7146NL/4152

K1:PEM-POS_MIC

2048

pic

Pa

MIC, amp

microphone

ACO 7146/4152

K1:PEM-OMC_MIC

2048

pic

Pa

MIC, amp

microphone

ACO 7146/4152

K1:PEM-BS_MIC

2048

pic

Pa

MIC, amp

microphone

ACO 7146/4152

K1:PEM-SR_MIC

2048

pic

Pa

MIC, amp

Seismometer

Snumber

Installed ADC port

Name

Channel name (OUT_DQ)

Unit

fs [Hz]

klog

JGW-S1707552

IX1, ADC0, 24-26

Trillium 12QA

K1:PEM-IXV_GND_TR120Q_(X/Y/Z)_OUT_DQ

um/s

512

7186

JGW-S1707551

EX1, ADC0, 24-26

Trillium 120QA

K1:PEM-EXV_GND_TR120Q_(X/Y/Z)_OUT_DQ

um/s

512

7186

JGW-S1707553

EY1, ADC0, 24-26

Trillium 120QA

K1:PEM-EYV_GND_TR120Q_(X/Y/Z)_OUT_DQ

um/s

512

7186

JGW-S1707553

IX1, ADC0, 28-30

Trillium 120QA

K1:PEM-IXV_GND_TR120QTEST_(X/Y/Z)_OUT_DQ

um/s

512

7186

JGW-S170xxxx

IOO1, ADC1, 4-6

Trillium compact 120-SV1

K1:PEM-IMC_GND_TR120C_MCE_(X/Y/Z)_OUT_DQ

um/s

512

JGW-S170xxxx

IOO1, ADC1, 0-2

Trillium compact 120-SV1

K1:PEM-IMC_GND_TR120C_MCI_(X/Y/Z)_OUT_DQ

um/s

512

JGW-S1707528

Fukuoka Univ.

Trillium compact 120-SV1

---

---

512

{*} Selfnoiseのグラフを載せる

Air Environment Sensors

Snumber

Installed Place (ADC port)

Name

Channel name (xxx_OUT_DQ)

Unit

Sample rate [Hz]

JGW-S1709339

IXV (IX1, ADC1, 24-27)

No.5

K1-IXV_WEATHER_{TEMP,HUMD,PRES}_

Degree, %, Pa

128

JGW-S170XXX

IYV (IY1, ADC1, 24-27)

128

JGW-S170xxxx

EXV (EX1, ADC1, 24-27)

---

K1-EXV_WEATHER_{TEMP,HUMD,PRES}_

Degree, %, Pa

128

JGW-S170xxxx

EYV (EY1, ADC1, 24-27)

---

K1-EYV_WEATHER_{TEMP,HUMD,PRES}_

Degree, %, Pa

128

JGW-S170xxxx

IXC (IX0, ADC0, 16-27?)

128

JGW-S1709340

IYC (IY0, ADC0, 20-23)

No6

K1-IYC_WEATHER_{TEMP,HUMD,PRES}_

Degree, %, Pa

128

Acording to JGW-T1809269, calibration factor from output voltage to each unit is given with

temp = ( Vout/78/0.001-100)/0.388  [deg_c]
humd = (Vout/2)/5*100 [%]
pressure = (Vout/2)/5*300 + 800 [hPa]

Then, "_GAIN" and "_OFFSET" parameters in cds filter of each channels are calculated by above equation with count value not voltage,

Name

Value

Unit

_TEMP_OFFSET

-1.278e4

count

_HUMD_OFFSET

0.0

count

_PRES_OFFSET

4.369e4

count

_TEMP_GAIN

2.017e-2

deg_c/count

_HUMD_GAIN

6.104e-3

%/count

_PRES_GAIN

1.831e-2

hPa/count

,where ransformation coefficient from voltage to adc count is given with c2V = 2.0 * 10/2**15 [V/count], because output signal from sensor is measured with single-end not differential. These parameters are managed by SDF with observe.snap. So, plase do not change this file.

Temporary channels in IX0

Place

Sensor

Channel name

picture

Memo

IX0

Temporary

K1:PEM-IX0_SENSOR(1-12)

pic

Will be removed

Temporary channels in IY0

Place

Sensor

Channel name

picture

Memo

IY0

Temporary

K1:PEM-IY0_SENSOR(1-12)

pic

Will be removed

EX(-201X/XX/XX)

  • PCal Tx table (TEAC710)
  • PCal Rx table (TEAC710)
  • EXA area (ACO mic)
  • EXC/EXV area (ACO mic)
  • EXV talbe (TEAC706)
  • EXC area (magnetometer)

EY(-201X/XX/XX)

  • PCal Tx table (TEAC710)
  • PCal Rx table (TEAC710)
  • EYA area (ACO mic)
  • EYC/EYV area (ACO mic)
  • EYV talbe (TEAC706)
  • EYC area (magnetometer)


Channel Name Definition

Definition of the PEM channel name

K1:PEM-(SENSOR)_(AREA)_(LOCATION)_(DOF)

Name

Purpose

SENSOR

ONDTR

Ondotori sensor

ACC

Accelerometer

MIC

Microphone

MAG

Magnetometer

SEIS

Seismometer

WEATHER

Weather station

SENSOR

This name is not for specific sensor

AREA

PSL

in the PSL Room area

IX

around the IX area

IY

around the IY area

EX

around the EX area

EY

around the EY area

IXV

in the IXV area

IYV

in the IYV area

EXV

in the EXV area

EYV

in the EYV area

TMSX

around the TMSX area

TMSY

around the TMSY area

PCALX

around the PCALX area

PCALY

around the PCALY area

PR

around the PR area including chamber

BS

around the BS area including chamber

SR

around the SR area including chamber

IMC

around the MCi, MCe, MCo area

IFI

around the IFI area including table, chamber

OMC

around the OMC area including table, chamber

POP

around the POP area including table, chamber

POS

around the POS area including talbe, chamber

RACK

in the DGS Rack

FIELD

on the field

BOOTH

in the "any" clean booth

MOZUMI

in the mozumi building

XARM

x arm

YARM

y arm

LOCATION

CHAMBER_(NAME)

on the chamber. Name : here

TABLE_(NAME)

on the table. Name : here

BOOTH_(NAME)

in the clean booth. Name : here?

BEAMDUCT_(NAME)

on the beamduct. Name : ???

PERI_(NAME)

on the periscope. Name : ???

FIELD

It means other places.

DOF

X/Y/Z

X,Y,Z axis as KAGRA cordinate

18V/24V

type of the DC power supply

TEMPERATURE/HUMIDITY/PRESSSURE

signals of the both ondotori and weather station

  • SENSOR
    • PEMが管理しているセンサーの種類がここに該当する。
  • AREA
    • エリア名が該当する。
    • 三代木さんが定義。here

  • LOCATION
    • センサーが設置されている構造物の名前が該当する。たとえば、クリーンブースや光学定盤の名前。
    • クリーンブースは三代木さんが定義。here

    • 光学定盤はIOOグループが定義。here

  • DOF
    • 1つのセンサーが複数個の出力信号を持つ場合、それぞれの名前が該当する。


SDF

latest snapfile

2018/12/03 Update, miyo

Model

snapfile

Managed date

klog

k1pemiy0

observe.snap

Dec 03 09:49:41 2018

7249

k1pemix1

observe.snap

Nov 28 19:27:27 2018

7249

k1pemex1

observe.snap

Nov 28 19:25:46 2018

7249

k1pemey1

observe.snap

Nov 30 17:15:59 2018

7249

KAGRA/Subgroups/PEM/Channels (last edited 2019-05-11 12:31:04 by KousekiMiyo)