Differences between revisions 3 and 5 (spanning 2 versions)
Revision 3 as of 2019-11-05 17:01:40
Size: 2786
Comment:
Revision 5 as of 2019-11-05 18:00:51
Size: 4751
Comment:
Deletions are marked like this. Additions are marked like this.
Line 4: Line 4:
 * Attend:Yokozaqwa, TNTanaka, Fujikawa, Ohkawa, Washimi, Mori, Kozakai, Yamada  * Attend:Yokozaqwa, TNTanaka, Fujikawa, Ohkawa, Washimi, Mori, Kozakai, Yamada, Akutsu, Kaihotsu
Line 64: Line 64:
    * 週末のFPMIとPEMのcoherenceを見ていった。
    * 11HzのEXCにmagとの相関が見えたのですが、みに行った時間にはなかった、lockとの関係がある?
    * Pcal RxPD 1.3HzはETMYのgasが励起していて、それが影響していた可能性がある。->long oplevとして見えた。
    * 24Hzはrack noise
    * 44,45Hz Compressor or FFU?
    * Compressor sound make44Hz, 46Hz and 56Hz -> stop during observation, check the coherence with FPMI signal
    * 60Hz EXC magが大きい、GNDが汚れている可能性が高い
    * PSL acousticは複数micを置いて場所を見る。
    * ~110Hzはlaser amp
    * 840Hz IMC refl
    * Priorityを決める、Identifyとhuntingはちがう。
Line 66: Line 77:
   * IXC vibration    * IXC vibration -> See the Yamada-san's slide
     * Identify the origin of the frequency noise
     * 42Hzパイロンの共振
     * 44Hz 気まぐれcompressor
     * 46.125Hz
     * 54.125Hz きまぐれcompressor
     * 58音?
     * acoustic injectionで盛り上がるpeakがある。散乱光の影響か?
     * IXCとDARMのcoherenceを取れるdataが木曜日の夜にすこしだけある。
     * center areaのmicとDARMに7Hz coherence、エンドには出ない
     * ITMX MN Pとcoherence <20Hz
     * TM oplev定盤に置いたACCとDARMにcoherenceがある、46Hz、いいヒント
     * MICとITMX PEMのcoherence、46Hzにmic vs TM oplev、lengthにあるかも
     * プシュー有り無しで比較したい。
     * compressor, 2分静か、30秒かたかた、10分、プシュー
Line 70: Line 95:
   * glitch plotがautoで走ったので、Yuzu summary pageを見てみてください。
   * glitch plot catalog
   * 磁束計で見えているのと地震計で見えているものがある、何処のsuspensionが影響が大きいかなどもわかると思う。
   * Fabianがコメントしていた、oplev jumpは影響あるか?->あとでみてみる

Minute of PEM meeting

  • Date: 2019/11/05 17:00-18:00(JST)
  • Attend:Yokozaqwa, TNTanaka, Fujikawa, Ohkawa, Washimi, Mori, Kozakai, Yamada, Akutsu, Kaihotsu

Agenda

  • General issue
    • Interferometer status
      • Now, we can lock the interferometer stably
      • ISS(Intensity Stabilization System) was implemented, better sensitivity
      • Long silent run at 2nd Nobember (klog11312)

      • See the detail in url

      • Silent run -> Some excitation(including calibration line) -> PEM injection -> Hardware injection

  • PTEP paper
    • [kagra 03133]
    • KAGRA Overview papers
    • Overview of KAGRA : (3) Detchar, calibration, physical environment, data quality
  • Next F2F meeting
    • RESCEU
    • 4th and 5th Dec., 2019
    • 1 slot Washimi-san will have a talk
    • If you plan to write master/doctor thesis, please let me know
  • Chrome Book project (Portable PEM researcher)
  • PEM injection
    • Shaker injection1 (TTanaka, Yokozawa)
    • Shaker injection2 (Okutomi, Fabian)
    • acoustic injection (TTanaka, Yokozawa)
    • magnetic field injection (Niigata Tanaka, Fujikawa, Yokozawa)
    • RF injection (TTanaka, Washimi)
  • Weekly PEM measurement
    • Friday is weekly PEM measurement
    • Understanding the effect of the instruments, search the origin of noise, and so on
    • Tentative : New vacuum pump, 18V DC noise(24Hz), Cryochamber vibration(46Hz), FFU, Noise search
  • Database for the KAGRA characteristic lines
  • Visitors from LV
    • Anamaria from Livingston : 18 - 29 Nov.
    • Federico, Irene, Camilla (input optics) + Francesca from Virgo : 11/18-12/7 (2 weeks@KAGRA, 1 weeks@NAOJ)
  • Each progress
    • Yokozawa
      • Current serious noises
        • 24Hz noise (100V AC + AA?)
        • MICH control noise
        • Coherent with ITMX MN YAW
        • Coherence with ITMY TM PIT
      • Daily summary page
    • Washimi
      • Summary of coherence study between PEMs and DARM displacement
        • 週末のFPMIとPEMのcoherenceを見ていった。
        • 11HzのEXCにmagとの相関が見えたのですが、みに行った時間にはなかった、lockとの関係がある?
        • Pcal RxPD 1.3HzはETMYのgasが励起していて、それが影響していた可能性がある。->long oplevとして見えた。

        • 24Hzはrack noise
        • 44,45Hz Compressor or FFU?
        • Compressor sound make44Hz, 46Hz and 56Hz -> stop during observation, check the coherence with FPMI signal

        • 60Hz EXC magが大きい、GNDが汚れている可能性が高い
        • PSL acousticは複数micを置いて場所を見る。
        • ~110Hzはlaser amp
        • 840Hz IMC refl
        • Priorityを決める、Identifyとhuntingはちがう。
    • Washimi, Yamada, TTanaka
      • IXC vibration -> See the Yamada-san's slide

        • Identify the origin of the frequency noise
        • 42Hzパイロンの共振
        • 44Hz 気まぐれcompressor
        • 46.125Hz
        • 54.125Hz きまぐれcompressor
        • 58音?
        • acoustic injectionで盛り上がるpeakがある。散乱光の影響か?
        • IXCとDARMのcoherenceを取れるdataが木曜日の夜にすこしだけある。
        • center areaのmicとDARMに7Hz coherence、エンドには出ない
        • ITMX MN Pとcoherence <20Hz

        • TM oplev定盤に置いたACCとDARMにcoherenceがある、46Hz、いいヒント
        • MICとITMX PEMのcoherence、46Hzにmic vs TM oplev、lengthにあるかも
        • プシュー有り無しで比較したい。
        • compressor, 2分静か、30秒かたかた、10分、プシュー
    • Kozakai
      • Trigger analysis with silent run
      • glitch plotがautoで走ったので、Yuzu summary pageを見てみてください。
      • glitch plot catalog
      • 磁束計で見えているのと地震計で見えているものがある、何処のsuspensionが影響が大きいかなどもわかると思う。
      • Fabianがコメントしていた、oplev jumpは影響あるか?->あとでみてみる

    • TTanaka(ICRR), Washimi, TTanaka(Niigata), Yamada
      • Acoustic injection with FPMI
    • Mori
      • Vacuum pump vibration
    • Fujikawa, Mori
    • Fujikawa, Washimi
      • Lock loss study
    • TTanaka(Niigata)
      • Preparation of the coil
      • 24Hz noise situation
    • etc

KAGRA/Subgroups/PEM/Meeting/191105/Minute (last edited 2019-11-05 18:36:36 by TakaakiYokozawa)