= Magnetic Field Measurement in X-end = | = Magnetic Field Measurement in X-end = == Days and Members == * 2019/1/24 (Thu.) Washimi, Mori, Kaihotu, + mTakahashi(ICRR) * 2019/1/28 (Mon.) Washimi, + Ushiba(CRY) * 2019/1/31 (Thu.) Washimi, tTanaka, + hTanaka(CRY) == Apparatus and Settings == * Setup : Mag. -> WF -> AA -> ADC -> diaggui * Magnetic Sensors : Mag-13 (#141, #142, #192) * Whitening Filter : 00000000 10001100 00001100 00000000 * Digital Filter and Diaggui : see (by Mori) == Measured Position == [[attachment:EXCmap.png]] == Data List == directory in k1ctr : /users/PEM/MAG/EXC/ ||'''File Name'''||'''time'''||'''comment'''|| || EXC_inCRY_1.xml || 1/24 || 鏡の中心 || || EXC_inCRY_2.xml || 1/24 || 鏡の中心 || || EXC_inCRY_3.xml || 1/24 || _1 からX方向に+8cm (中心のアクチュエータマグネットの位置) || || EXC_inCRY_4.xml || 1/24 || _3 からY方向に-10cm (左側のアクチュエータマグネットの位置) || || EXC_inCRY_5.xml || 1/24 || _3 からY方向に+10cm (右側のアクチュエータマグネットの位置) || || EXC_inCRY_6.xml || 1/24 || _3 と同じ位置 (再現性確認) || || EXC_inCRY_7.xml || 1/24 || _6 で+90度回転 || || EXC_inCRY_8.xml || 1/24 || _7 で+90度回転 || || EXC_inCRY_21.xml || 1/28 14:40 || 冷凍機off, 空調on || || EXC_inCRY_22.xml || 1/28 15:13 || 冷凍機(P53~P56)のモーターon, 空調on || || EXC_inCRY_23.xml || 1/28 15:24 || 冷凍機(P53~P56)のモーターon, 空調off || || EXC_inCRY_24.xml || 1/28 15:30 || 冷凍機off, 空調off || || EXC_inCRY_25.xml || 1/28 15:35 || 冷凍機off, 空調on || || EXC_inCRY_31.xml || 1/31 14:10 || 冷凍機off, 空調on || || EXC_inCRY_32.xml || 1/31 14:10 || 鏡用冷凍機on, ダクト用冷凍機off, 空調on || || EXC_inCRY_33.xml || 1/31 14:26 || 鏡用冷凍機on, ダクト用冷凍機off, 空調on (_32と同じ) || || EXC_inCRY_34.xml || 1/31 14:33 || 冷凍機on, 空調on || || EXC_inCRY_35.xml || 1/31 14:41 || 冷凍機on, 空調off || || EXC_inCRY_36.xml || 1/31 14:48 || 冷凍機off, 空調off || || EXC_inCRY_37.xml || 1/31 15:01 || 冷凍機off, 空調off, + 電磁波遮蔽シート || || EXC_inCRY_41.xml || 1/31 17:21 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_42.xml || 1/31 17:33 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_43.xml || 1/31 19:00 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_44.xml || 1/31 20:11 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_45.xml || 1/31 21:04 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_46.xml || 1/31 21:59 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_47.xml || 1/31 23:13 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_48.xml || 2/1 0:00 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_49.xml || 2/1 2:23 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_50.xml || 2/1 5:51 || 冷凍機off, 空調off, 長期測定 || || EXC_inCRY_51.xml || 2/1 7:23 || 冷凍機off, 空調off, 長期測定 || || EXC_inoutCRY_1.xml || 1/24 || 冷凍機P53の真上 || || EXC_inoutCRY_2.xml || 1/24 || 冷凍機P54の真上 || || EXC_inoutCRY_3.xml || 1/24 || 冷凍機P55の真上 || || EXC_inoutCRY_4.xml || 1/24 || 冷凍機P56の真上 || || EXC_inoutCRY_5.xml || 1/24 || FFU(center側)の前 || || EXC_inoutCRY_6.xml || 1/24 || FFUの電源パネル中心の前 || || EXC_inoutCRY_7.xml || 1/24 || FFU(end側)の前 || || EXC_inoutCRY_8.xml || 1/24 || ビームダクト(ベローズ)の下 || || EXC_inoutCRY_9.xml || 1/24 || エレキラックEX0の前 || || EXC_inoutCRY_10.xml || 1/24 || エレキラックcryoconの前 || || EXC_inoutCRY_11.xml || 1/24 || エレキラックTMSXの前 || || EXC_inoutCRY_12.xml || 1/24 || 冷凍機P52の下 || || EXC_inoutCRY_31.xml || 1/31 15:12 || 冷凍機off, 空調off, Mag②をP53の上に || || EXC_inoutCRY_32.xml || 1/31 15:18 || P53のみon, 空調off, Mag②をP53の上に || || EXC_inoutCRY_33.xml || 1/31 15:23 || 冷凍機on, 空調off, Mag②をP53の上に || |
Magnetic Field Measurement in X-end
Days and Members
- 2019/1/24 (Thu.) Washimi, Mori, Kaihotu, + mTakahashi(ICRR)
- 2019/1/28 (Mon.) Washimi, + Ushiba(CRY)
- 2019/1/31 (Thu.) Washimi, tTanaka, + hTanaka(CRY)
Apparatus and Settings
Setup : Mag. -> WF -> AA -> ADC -> diaggui
- Magnetic Sensors : Mag-13 (#141, #142, #192)
- Whitening Filter : 00000000 10001100 00001100 00000000
- Digital Filter and Diaggui : see (by Mori)
Measured Position
Data List
directory in k1ctr : /users/PEM/MAG/EXC/
File Name |
time |
comment |
EXC_inCRY_1.xml |
1/24 |
鏡の中心 |
EXC_inCRY_2.xml |
1/24 |
鏡の中心 |
EXC_inCRY_3.xml |
1/24 |
_1 からX方向に+8cm (中心のアクチュエータマグネットの位置) |
EXC_inCRY_4.xml |
1/24 |
_3 からY方向に-10cm (左側のアクチュエータマグネットの位置) |
EXC_inCRY_5.xml |
1/24 |
_3 からY方向に+10cm (右側のアクチュエータマグネットの位置) |
EXC_inCRY_6.xml |
1/24 |
_3 と同じ位置 (再現性確認) |
EXC_inCRY_7.xml |
1/24 |
_6 で+90度回転 |
EXC_inCRY_8.xml |
1/24 |
_7 で+90度回転 |
EXC_inCRY_21.xml |
1/28 14:40 |
冷凍機off, 空調on |
EXC_inCRY_22.xml |
1/28 15:13 |
冷凍機(P53~P56)のモーターon, 空調on |
EXC_inCRY_23.xml |
1/28 15:24 |
冷凍機(P53~P56)のモーターon, 空調off |
EXC_inCRY_24.xml |
1/28 15:30 |
冷凍機off, 空調off |
EXC_inCRY_25.xml |
1/28 15:35 |
冷凍機off, 空調on |
EXC_inCRY_31.xml |
1/31 14:10 |
冷凍機off, 空調on |
EXC_inCRY_32.xml |
1/31 14:10 |
鏡用冷凍機on, ダクト用冷凍機off, 空調on |
EXC_inCRY_33.xml |
1/31 14:26 |
鏡用冷凍機on, ダクト用冷凍機off, 空調on (_32と同じ) |
EXC_inCRY_34.xml |
1/31 14:33 |
冷凍機on, 空調on |
EXC_inCRY_35.xml |
1/31 14:41 |
冷凍機on, 空調off |
EXC_inCRY_36.xml |
1/31 14:48 |
冷凍機off, 空調off |
EXC_inCRY_37.xml |
1/31 15:01 |
冷凍機off, 空調off, + 電磁波遮蔽シート |
EXC_inCRY_41.xml |
1/31 17:21 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_42.xml |
1/31 17:33 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_43.xml |
1/31 19:00 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_44.xml |
1/31 20:11 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_45.xml |
1/31 21:04 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_46.xml |
1/31 21:59 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_47.xml |
1/31 23:13 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_48.xml |
2/1 0:00 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_49.xml |
2/1 2:23 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_50.xml |
2/1 5:51 |
冷凍機off, 空調off, 長期測定 |
EXC_inCRY_51.xml |
2/1 7:23 |
冷凍機off, 空調off, 長期測定 |
EXC_inoutCRY_1.xml |
1/24 |
冷凍機P53の真上 |
EXC_inoutCRY_2.xml |
1/24 |
冷凍機P54の真上 |
EXC_inoutCRY_3.xml |
1/24 |
冷凍機P55の真上 |
EXC_inoutCRY_4.xml |
1/24 |
冷凍機P56の真上 |
EXC_inoutCRY_5.xml |
1/24 |
FFU(center側)の前 |
EXC_inoutCRY_6.xml |
1/24 |
FFUの電源パネル中心の前 |
EXC_inoutCRY_7.xml |
1/24 |
FFU(end側)の前 |
EXC_inoutCRY_8.xml |
1/24 |
ビームダクト(ベローズ)の下 |
EXC_inoutCRY_9.xml |
1/24 |
エレキラックEX0の前 |
EXC_inoutCRY_10.xml |
1/24 |
エレキラックcryoconの前 |
EXC_inoutCRY_11.xml |
1/24 |
エレキラックTMSXの前 |
EXC_inoutCRY_12.xml |
1/24 |
冷凍機P52の下 |
EXC_inoutCRY_31.xml |
1/31 15:12 |
冷凍機off, 空調off, Mag②をP53の上に |
EXC_inoutCRY_32.xml |
1/31 15:18 |
P53のみon, 空調off, Mag②をP53の上に |
EXC_inoutCRY_33.xml |
1/31 15:23 |
冷凍機on, 空調off, Mag②をP53の上に |