Line 9: | Line 9: |
* Found that the flanges | * All * Found that flanges of MCF and MCE chambers were not protected by white tapes although there were viton o-rings on. We protected these by white tapes. -> Informed this to SEO already. * MCi * Relocated the connector plate. * Attached adapter structures for black shields onto the pillars of the MCi suspension. (But not yet the black plates.) * MCo * Found the connector plate on the MCo suspension was '''hard''' to relocate on the optical table in the chamber due to short wires from actuator coils. * MCe * Relocated the connector plate. |
Line 12: | Line 20: |
* Wash some screws. * Have VIS to modify the MCi connector. * Have VIS to consider how to tame/fix the electric wires from the coil actuators so that they won't cross the laser beams. * Check the ghost beams. |
|
Line 14: | Line 26: |
* Metal racks in the clean booth (MCF and MCE); the height must be lower than 180 cm. * |
* 2 metal rack shelves in the clean booth (MCF and MCE); the height must be lower than 180 cm due to lower ceiling of the relevant clean booths. * 1 sticky mat to get the dirt off our feet. |
20210203
klog
- xxx
- xxx
Today's activities
Workers: Akutsu, N. Sato, K. Tanaka, Nakano
- All
Found that flanges of MCF and MCE chambers were not protected by white tapes although there were viton o-rings on. We protected these by white tapes. -> Informed this to SEO already.
- MCi
- Relocated the connector plate.
- Attached adapter structures for black shields onto the pillars of the MCi suspension. (But not yet the black plates.)
- MCo
Found the connector plate on the MCo suspension was hard to relocate on the optical table in the chamber due to short wires from actuator coils.
- MCe
- Relocated the connector plate.
Next plan
- Wash some screws.
- Have VIS to modify the MCi connector.
- Have VIS to consider how to tame/fix the electric wires from the coil actuators so that they won't cross the laser beams.
- Check the ghost beams.
Wanted!
- 2 metal rack shelves in the clean booth (MCF and MCE); the height must be lower than 180 cm due to lower ceiling of the relevant clean booths.
- 1 sticky mat to get the dirt off our feet.