Differences between revisions 11 and 12
Revision 11 as of 2019-11-11 19:26:08
Size: 879
Comment:
Revision 12 as of 2019-11-11 19:27:54
Size: 922
Comment:
Deletions are marked like this. Additions are marked like this.
Line 8: Line 8:
||Standard Filter ||It's been at its lowest position since the temperature was 25 ºC.|| || ||Standard Filter ||It's been at its lowest position since the temperature was 25 ºC. We're using coil actuation to compensate. || ||

Describe KAGRA/Subgroups/VIS/OpsManual/FishingRods here.

This page describes the status of the fishing rods.

Beam splitter

Stage

Description

Relevant log entries

Top Filter

At 25 ºC it's possible to use it to set the keystone at nominal (8125), but at 24.5 ºC or lower it's not (8645, 11465).

8125, 8645, 11465

Standard Filter

It's been at its lowest position since the temperature was 25 ºC. We're using coil actuation to compensate.

Bottom Filter

So far it's been working, but we haven't tried using it at all every temperature condition.

KAGRA/Subgroups/VIS/OpsManual/FishingRods (last edited 2020-09-18 10:45:40 by fabian.arellano)