Differences between revisions 37 and 103 (spanning 66 versions)
Revision 37 as of 2019-11-13 15:25:37
Size: 3645
Comment:
Revision 103 as of 2020-09-18 10:45:40
Size: 12822
Comment:
Deletions are marked like this. Additions are marked like this.
Line 1: Line 1:
Describe KAGRA/Subgroups/VIS/OpsManual/FishingRods here. This page describes the status of the fishing rods and provides characterization information measured at the time of assembly.
Line 3: Line 3:
This page describes the status of the fishing rods. It's important to bear in mind that the reported range of the stepper motor in number of steps is not expected to be accurate and it's useful only as a guideline:

 * Whenever the stepper driver is reset the posion of the motor (measured in number of steps from a zero) is lost and begins counting from zero wherever its physical position is.
 * Inaccuracies may arise when a stepper motor reaches the physical end of range because the driver still considers valid the requested position even when its beyond the physcal limit.

With respecto to the LVDT range limits, those may change with temperature.
||'''Suspension''' ||'''Stage''' ||'''Available''' ||'''Polarity (+steps)''' ||'''Range''' ||'''Others''' ||
||<style="text-align:center" |5>ITMX ||F0 ||Stacked[?] ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F1 ||OK [1] ||Up [1] ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F2 ||OK [1] ||Down [1] ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F3 ||NO [1] ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||BF ||Stacked[?] ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||<style="text-align:center" |5>ETMX ||F0 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F1 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F2 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F3 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||BF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||<style="text-align:center" |5>ITMY ||F0 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F1 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F2 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F3 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||BF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||<style="text-align:center" |5>ETMY ||F0 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F1 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F2 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||F3 ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||BF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||'''Suspension''' ||'''Stage''' ||'''Available''' ||'''Polarity (+steps)''' ||'''Range''' ||'''Others''' ||
||<style="text-align:center" |3>BS ||F0 ||? ||? ||'''LVDT''' : (-2500,+4900) um [2]<<BR>>'''Stepper''' : (-1.4e6, +1.4e6 ) steps [2] || ||
||F1 ||? ||? ||'''LVDT''' : ( -700 , +700 ) µm [3]<<BR>>'''Stepper''' : ( -870,000 , +835,000 ) steps [3] || ||
||BF ||? ||? ||'''LVDT''' : ( -1000 , +1800 ) µm [4]<<BR>>'''Stepper :''' ( -746,000 , +746,000 ) steps [4] || ||
||<style="text-align:center" |3>SRM ||F0 ||? ||? ||'''LVDT:''' : ( -865 , +626 ) µm [7]<<BR>>'''Stepper''' : ( -1.6e6 , +1.6e6 ) steps [7] || ||
||F1 ||? ||? ||'''LVDT:''' : ( -1700 , +1400 ) µm [8]<<BR>>'''Stepper''' : ( -7e5 , +1.1e6 ) steps [8] || ||
||BF ||? ||? ||'''LVDT:''' : ( -3000 , +2800 ) µm [9]<<BR>>'''Stepper''' : ( -6e5 , +8e5 ) steps [9] || ||
||<style="text-align:center" |3>SR2 ||F0 ||? ||? ||'''LVDT:''' : ( -990 , 903 ) µm [5]<<BR>>'''Stepper''' : ( -1,700,000 , 1,500,000 ) steps [5] || ||
||F1 ||? ||? ||'''LVDT''' : ( -750 , 1000 ) µm [6]<<BR>>'''Stepper''' : ( 300,000 , 1,000,00 ) steps [6] || ||
||BF ||? ||? ||'''LVDT:''' : ( -500 , +600 ) µm [10]<<BR>>'''Stepper''' : [ -5.5e5 , -2.5e5 ] steps [10] || ||
||<style="text-align:center" |3>SR3 ||F0 ||? ||? ||'''LVDT:''' : ( +885 , -1299 ) µm [11]<<BR>>'''Stepper''' : ( -5e5 , +1.1e6 ) steps [11] || ||
||F1 ||? ||? ||'''LVDT:''' : ( -2000 , +2000 ) µm [12]<<BR>>'''Stepper''' : ( -7e5 , +1.1e6 ) steps [12] || ||
||BF ||? ||? ||'''LVDT:''' : ( -1600 , +1000 ) µm [13]<<BR>>'''Stepper''' : ( -2e5 , +2.5e5 ) steps [13] || ||
||'''Suspension''' ||'''Stage''' ||'''Available''' ||'''Polarity (+steps)''' ||'''Range''' ||'''Others''' ||
||<style="text-align:center" |2>PRM ||SF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||BF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||<style="text-align:center" |2>PR2 ||SF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||BF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||<style="text-align:center" |2>PR3 ||SF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||
||BF ||? ||? ||'''LVDT:''' : ( , ) µm []<<BR>>'''Stepper''' : ( , ) steps [] || ||




References

 * [1] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=14820|14820]]
 * [2] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7694|7694]]
 * [3] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=3558|3558]]
 * [4] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=3280|3280]]
 * [5] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8082|8082]]
 * [6] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4431|4431]]
 * [7] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7857|7857]]
 * [8] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]]
 * [9] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]]
 * [10] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4430|4430]]
 * [11] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7758|7758]]
 * [12] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4720|4720]]
 * [13] [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4654|4654]]
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=
 * [] http://klog.icrr.u-tokyo.ac.jp/osl/?r=

<<TableOfContents>>
Line 17: Line 93:
  * '''LVDT:''' [ -700 , +700 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=3358|3558]]).
  * '''Stepper motor:''' [ -870,000 to +835,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=3358|3558]]).
  * '''LVDT:''' [ -700 , +700 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7738|7738]], [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=3358|3358]]).
  * '''Stepper motor:''' [ -870,000 to +835,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7738|7738]], [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=3358|3358]]).
Line 34: Line 110:
  * '''LVDT:''' [ -1,700,000 , 1,500,000 ] µm
  * Stepper motor:
  * '''LVDT:''' [ -990 , 903 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8082|8082]]).
  * '''Stepper motor:''' [ -1,700,000 , 1,500,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8082|8082]]).
Line 37: Line 113:
 * Observations:  * '''Observations:''' This fishing rod is not known to have problems. See klog entry [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=13930|13930]].
Line 44: Line 120:
 * Observations:  * '''Observations:''' This fishing rod is not known to have problems. See klog entry [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=13930|13930]].
Line 51: Line 127:
 * Observations:  * '''Observations:''' This fishing rod is not known to have problems.
Line 54: Line 130:
[[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8082|8082]]  * Calculated temperature correction: [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8082|8082]].
 * Recovery after pump down: [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8160|8160]].
Line 58: Line 135:
([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7758|7758]])
Line 61: Line 136:
  * '''LVDT:''' µm
  * Stepper motor:
  * '''LVDT:''' [ 885 , -1299 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7758|7758]]).
  * '''Stepper motor:''' [ -500,000 , 1,100,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7758|7758]]).
Line 64: Line 139:
 * Observations:  * '''Observations:''' This fishing rod doesn't seem to have any problems ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8135|8135]], [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8658|8658]], [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=14890|14890]]) (The previouly reported status was wrong: "After pump down it was not possible to move the keystone to zero using the fishing rod. It went as high as -750 µm at the time". However, I didn't find any klog reporting a problem with this fishing rod.)
Line 67: Line 142:
([[http://p://klog.icrr.u-tokyo.ac.jp/osl/?r=4720|4720]])
Line 70: Line 143:
  * '''LVDT:''' µm
  * Stepper motor:
  * '''LVDT:''' [ -2000 , 2000 ]''' '''µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4720|4720]]).
  * '''Stepper motor:''' [ -700,000 , 1,100,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4720|4720]]).
Line 73: Line 146:
 * Observations:  * '''Observations:'''This fishing rod of not working at all ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8135|8135]]). It's a real shame because it had a large range of 4 mm total.
Line 76: Line 149:
([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4654|4654]])
Line 79: Line 150:
  * '''LVDT:''' µm
  * Stepper motor:
  * '''LVDT:''' [ -1600 , +1000 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4654|4654]]).
  * '''Stepper motor:''' [ -200,000 , 250,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4654|4654]]).
Line 82: Line 153:
 * Observations:  * '''Observations:'''
  * After pump down it was not possible to move the keystone to zero using the fishing rod. It went as high as -265 µm at the time [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8658|8658]].
  * The number of steps for the stepper motor seem relatively small. In principle there's nothing wrong with this but let's point this out.
Line 89: Line 162:
([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7857|7857]])  * '''Range:'''
  * '''LVDT:''' [ -865 , 626 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7857|7857]]).
  * '''Stepper motor:''' [ -1,600,000 , 1,600,00 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=7857|7857]]).
  * Temperature when measured:
 * '''Observations: '''This fishing rod is not known to have problems.
Line 92: Line 169:
([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]])  * '''Range:'''
  * '''LVDT:''' [ -1,700 , 1,400 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]]) .
  * '''Stepper motor:''' [ -700,000 , 1,100,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]]) .
  * Temperature when measured:
 * '''Observations: '''This fishing rod is not known to have problems.
Line 95: Line 176:
([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]])  * '''Range:'''
  * '''LVDT:''' [ -3,000 , 2,800 ] µm ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]]).
  * '''Stepper motor:''' [ -600,000 , 800,000 ] steps ([[http://klog.icrr.u-tokyo.ac.jp/osl/?r=4721|4721]]).
  * Temperature when measured:
 * '''Observations: '''This fishing rod is not known to have problems.

Related entries

Recovery after pump down: [[http://klog.icrr.u-tokyo.ac.jp/osl/?r=8197|8197]].

This page describes the status of the fishing rods and provides characterization information measured at the time of assembly.

It's important to bear in mind that the reported range of the stepper motor in number of steps is not expected to be accurate and it's useful only as a guideline:

  • Whenever the stepper driver is reset the posion of the motor (measured in number of steps from a zero) is lost and begins counting from zero wherever its physical position is.
  • Inaccuracies may arise when a stepper motor reaches the physical end of range because the driver still considers valid the requested position even when its beyond the physcal limit.

With respecto to the LVDT range limits, those may change with temperature.

Suspension

Stage

Available

Polarity (+steps)

Range

Others

ITMX

F0

Stacked[?]

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F1

OK [1]

Up [1]

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F2

OK [1]

Down [1]

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F3

NO [1]

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

BF

Stacked[?]

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

ETMX

F0

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F1

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F2

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F3

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

BF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

ITMY

F0

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F1

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F2

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F3

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

BF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

ETMY

F0

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F1

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F2

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

F3

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

BF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

Suspension

Stage

Available

Polarity (+steps)

Range

Others

BS

F0

?

?

LVDT : (-2500,+4900) um [2]
Stepper : (-1.4e6, +1.4e6 ) steps [2]

F1

?

?

LVDT : ( -700 , +700 ) µm [3]
Stepper : ( -870,000 , +835,000 ) steps [3]

BF

?

?

LVDT : ( -1000 , +1800 ) µm [4]
Stepper : ( -746,000 , +746,000 ) steps [4]

SRM

F0

?

?

LVDT: : ( -865 , +626 ) µm [7]
Stepper : ( -1.6e6 , +1.6e6 ) steps [7]

F1

?

?

LVDT: : ( -1700 , +1400 ) µm [8]
Stepper : ( -7e5 , +1.1e6 ) steps [8]

BF

?

?

LVDT: : ( -3000 , +2800 ) µm [9]
Stepper : ( -6e5 , +8e5 ) steps [9]

SR2

F0

?

?

LVDT: : ( -990 , 903 ) µm [5]
Stepper : ( -1,700,000 , 1,500,000 ) steps [5]

F1

?

?

LVDT : ( -750 , 1000 ) µm [6]
Stepper : ( 300,000 , 1,000,00 ) steps [6]

BF

?

?

LVDT: : ( -500 , +600 ) µm [10]
Stepper : [ -5.5e5 , -2.5e5 ] steps [10]

SR3

F0

?

?

LVDT: : ( +885 , -1299 ) µm [11]
Stepper : ( -5e5 , +1.1e6 ) steps [11]

F1

?

?

LVDT: : ( -2000 , +2000 ) µm [12]
Stepper : ( -7e5 , +1.1e6 ) steps [12]

BF

?

?

LVDT: : ( -1600 , +1000 ) µm [13]
Stepper : ( -2e5 , +2.5e5 ) steps [13]

Suspension

Stage

Available

Polarity (+steps)

Range

Others

PRM

SF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

BF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

PR2

SF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

BF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

PR3

SF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

BF

?

?

LVDT: : ( , ) µm []
Stepper : ( , ) steps []

References

Beam splitter

Top filter

  • Range:

    • LVDT: [ -2500 , +4900 ] µm (7694).

    • Stepper motor: [ -1,400,000 , +1,400,000 ] steps (7694).

    • Temperature when measured:
  • Observations:

    • At 25 ºC it was possible to use the fishing rod to set the keystone at nominal position (8125).

    • at 24.5 ºC or colder it was not (8645, 11465).

Standard Filter

  • Range:

    • LVDT: [ -700 , +700 ] µm (7738, 3358).

    • Stepper motor: [ -870,000 to +835,000 ] steps (7738, 3358).

    • Temperature when measured:
  • Observations:

    • It's been at its lowest position since the temperature was 25 ºC (8125, 8645).

    • We set the keystone to 488 µm and couldn't move it further down (8125, 8645).

Bottom filter

  • Range:

    • LVDT: [ -1000 µm , +1800 ] µm (3280).

    • Stepper motor: [ -746,000 , +746,000 ] steps (3280).

    • Temperature when measured:
  • Observations: this fishing rod doesn't seem to have any problems, but we haven't used it at every temperature condition we've had.

SR2

Top filter

  • Range:

    • LVDT: [ -990 , 903 ] µm (8082).

    • Stepper motor: [ -1,700,000 , 1,500,000 ] steps (8082).

    • Temperature when measured:
  • Observations: This fishing rod is not known to have problems. See klog entry 13930.

Stardard filter

  • Range:

    • LVDT: [ -750 , 1000 ] µm (4431).

    • Stepper motor: [ 300,000 , 1,000,00 ] steps (4431).

    • Temperature when measured:
  • Observations: This fishing rod is not known to have problems. See klog entry 13930.

Bottom filter

  • Range:

    • LVDT: [ -500 , 600 ] µm (4430).

    • Stepper motor: [ -550,000 , -250,000 ] steps (4430).

    • Temperature when measured:
  • Observations: This fishing rod is not known to have problems.

  • Calculated temperature correction: 8082.

  • Recovery after pump down: 8160.

SR3

Top filter

  • Range:

    • LVDT: [ 885 , -1299 ] µm (7758).

    • Stepper motor: [ -500,000 , 1,100,000 ] steps (7758).

    • Temperature when measured:
  • Observations: This fishing rod doesn't seem to have any problems (8135, 8658, 14890) (The previouly reported status was wrong: "After pump down it was not possible to move the keystone to zero using the fishing rod. It went as high as -750 µm at the time". However, I didn't find any klog reporting a problem with this fishing rod.)

Standard Flter

  • Range:

    • LVDT: [ -2000 , 2000 ] µm (4720).

    • Stepper motor: [ -700,000 , 1,100,000 ] steps (4720).

    • Temperature when measured:
  • Observations:This fishing rod of not working at all (8135). It's a real shame because it had a large range of 4 mm total.

Bottom Filter

  • Range:

    • LVDT: [ -1600 , +1000 ] µm (4654).

    • Stepper motor: [ -200,000 , 250,000 ] steps (4654).

    • Temperature when measured:
  • Observations:

    • After pump down it was not possible to move the keystone to zero using the fishing rod. It went as high as -265 µm at the time 8658.

    • The number of steps for the stepper motor seem relatively small. In principle there's nothing wrong with this but let's point this out.

SRM

Top Filter

  • Range:

    • LVDT: [ -865 , 626 ] µm (7857).

    • Stepper motor: [ -1,600,000 , 1,600,00 ] steps (7857).

    • Temperature when measured:
  • Observations: This fishing rod is not known to have problems.

Standard Filter

  • Range:

    • LVDT: [ -1,700 , 1,400 ] µm (4721) .

    • Stepper motor: [ -700,000 , 1,100,000 ] steps (4721) .

    • Temperature when measured:
  • Observations: This fishing rod is not known to have problems.

Bottom Filter

  • Range:

    • LVDT: [ -3,000 , 2,800 ] µm (4721).

    • Stepper motor: [ -600,000 , 800,000 ] steps (4721).

    • Temperature when measured:
  • Observations: This fishing rod is not known to have problems.

Related entries

Recovery after pump down: 8197.

KAGRA/Subgroups/VIS/OpsManual/FishingRods (last edited 2020-09-18 10:45:40 by fabian.arellano)