3178
Comment:
|
3505
|
Deletions are marked like this. | Additions are marked like this. |
Line 3: | Line 3: |
==== Schedule ==== | === 1. Schedule === |
Line 9: | Line 9: |
==== Disassembling at ATC Mitaka [Ohishi] (12th-19th Jan.) ==== * Two nuts stuck. Hirata-san removed them. * A Cu screw stuck. Hirata-san removed it. ==== Washing and transportation to Kamioka [Hirata-san] ==== * The ultra sonic washer at ATC was broken on 20th Jan. |
=== 2. Disassembling in the ATC clean booth at Mitaka [Ohishi] (12th-19th Jan.) === * Two nuts stuck. Hirata-san removed them. * A Cu screw stuck. Hirata-san removed it. === 3. Washing and transportation to Kamioka [Hirata-san] === * The ultra sonic washer at ATC was broken on 20th Jan. |
Line 15: | Line 15: |
* Eight screws for adjusting the IM holding plate was dirty. | * Eight screws for adjusting the IM holding plate was dirty. |
Line 17: | Line 17: |
==== New parts ==== * Coil bobbin [Takahashi-san, Tatsumi-san] |
=== 4. New parts preparation === ==== Coil bobbin [Takahashi-san, Tatsumi-san] ==== |
Line 20: | Line 20: |
* The outer diameter of the coil has been changed from 14.2mm to 14.5mm. * Due to this change, the number of winding becomes 39 from 41. * Delivered to Kamioka on 4th Feb. * Coil winding [Kuwahara-san, and Ushiba-san from U-Tokyo, Lee-san from Okazaki, and Ohishi](8th-16th Feb.) * We used coil winding machine. * Lead wires from the bobbin to Dsub connector were too short (20cm). The length must be 40cm or more. * At the end of the lead wires, D-sub pins are swaged. We have run short of pins and waiting for delivery on 29th Feb(Mon). * Peeling off the coating of wires longer than the length of those pins make the wire broken easily. * D-sub connector plates [Takahashi-san] |
* The outer diameter of the coil has been changed from 14.2mm to 14.5mm. * Due to this change, the number of winding becomes 39 from 41. * Delivered to Kamioka on 4th Feb. ==== Coil winding [Kuwahara-san, and Ushiba-san from U-Tokyo, Lee-san from Okazaki, and Ohishi](8th-16th Feb.) ==== * We used coil winding machine. * Lead wires from the bobbin to Dsub connector were too short (20cm). The length must be 40cm or more. * At the end of the lead wires, D-sub pins are swaged. We have run short of pins and waiting for delivery on 29th Feb(Mon). * Peeling off the coating of wires longer than the length of those pins make the wire broken easily. ==== D-sub connector plates [Takahashi-san] ==== |
Line 30: | Line 30: |
* M3 holes were changed to phi3.1 holes. Yamamoto-san used drilling machine at Kamioka surface office. *15pin-9pin changer for coil driver output [Ushiba-san, Ohishi] * Mirror container [Tatsumi-san] [[attachment:MirrorContainer.pdf|{{attachment:MirrorContainer.pdf||width=600}}]]<<BR>> * some plates once used for transportation of MC mirrors were used * * Magnets for dummy mirror (4 ; 2-phi-L-5mm Nd magnets) [Tomaru, Yamamoto] |
* M3 holes were changed to phi3.1 holes. Yamamoto-san used drilling machine at Kamioka surface office. * One must be careful about the pillar position in order to attach this plate to the suspension. ==== 15pin-9pin changer for coil driver output [Ushiba-san, Ohishi] ==== ==== Screws in replace with Cu screws [Ohishi] ==== * Replaced all Cu screws to stainless screws. Magnetism problem was indicated. === Jigs === * We used CLIO magnet fixture jigs. ==== Attach magnets to dummy mirror (4 ; with the diameter of 2mm and the length of 5mm NdFeB magnets) [Tomaru, Yamamoto] |
Line 38: | Line 39: |
* Screws in replace with Cu screws [Ohishi] [problem] replaced all Cu screws to stainless. Magnetism problem was indicated. |
=== Some parts === |
Line 43: | Line 43: |
==== Coil driver ==== | ==== Coil drivers ==== |
Line 47: | Line 47: |
==== Assembling [Tomaru-san, Ohishi, Yamamoto-san] ==== | === Assembling [Tomaru-san, Ohishi, Yamamoto-san] === |
Line 50: | Line 50: |
==== Hanging ==== | === Hanging === |
Line 53: | Line 53: |
==== Transfer Function Measurement ==== | === Transfer Function Measurement === |
Line 55: | Line 55: |
==== Transportation to the End [Prof. Kimura, Sawada-san, Tomaru-san(KEK),]==== | === Transportation to the End [Prof. Kimura, Sawada-san, Tomaru-san(KEK),] === |
Line 60: | Line 60: |
==== Mirror container [Tatsumi-san] ==== [[attachment:MirrorContainer.pdf|{{attachment:MirrorContainer.pdf||width=600}}]]<<BR>> * Some plates once used for transportation of MC mirrors were used * Two places of the container were cut to avoid standoffs. |
|
Line 61: | Line 65: |
==== Stages ==== | === Stages ==== |
Line 64: | Line 68: |
==== Installation ==== | === Installation === |
iKAGRA ETM suspension preparation log
1. Schedule
- 12th Jan- (2 weeks) Disassembling, Washing, and transportation
- 25th Jan - (2 weeks) Assembling and Test Hanging with an Al dummy mirror
- 7th Feb- (1 week) coil winding, measurement of transfer function, transportation to the End area and installation test
- 15th Feb- (1 week)installation
2. Disassembling in the ATC clean booth at Mitaka [Ohishi] (12th-19th Jan.)
- Two nuts stuck. Hirata-san removed them.
- A Cu screw stuck. Hirata-san removed it.
3. Washing and transportation to Kamioka [Hirata-san]
- The ultra sonic washer at ATC was broken on 20th Jan.
- A new washer was ordered on 22th(Fri) and arrived at Mitaka on 25th(Mon).
- Eight screws for adjusting the IM holding plate was dirty.
- Takahashi-san ordered similar screws.
4. New parts preparation
Coil bobbin [Takahashi-san, Tatsumi-san]
- The outer diameter of the coil has been changed from 14.2mm to 14.5mm.
- Due to this change, the number of winding becomes 39 from 41.
- Delivered to Kamioka on 4th Feb.
Coil winding [Kuwahara-san, and Ushiba-san from U-Tokyo, Lee-san from Okazaki, and Ohishi](8th-16th Feb.)
- We used coil winding machine.
- Lead wires from the bobbin to Dsub connector were too short (20cm). The length must be 40cm or more.
- At the end of the lead wires, D-sub pins are swaged. We have run short of pins and waiting for delivery on 29th Feb(Mon).
- Peeling off the coating of wires longer than the length of those pins make the wire broken easily.
D-sub connector plates [Takahashi-san]
- M3 holes were changed to phi3.1 holes. Yamamoto-san used drilling machine at Kamioka surface office.
- One must be careful about the pillar position in order to attach this plate to the suspension.
15pin-9pin changer for coil driver output [Ushiba-san, Ohishi]
Screws in replace with Cu screws [Ohishi]
- Replaced all Cu screws to stainless screws. Magnetism problem was indicated.
Jigs
- We used CLIO magnet fixture jigs.
==== Attach magnets to dummy mirror (4 ; with the diameter of 2mm and the length of 5mm NdFeB magnets) [Tomaru, Yamamoto]
Some parts
Pico-motors and PZTs
- We have 4 pico-motors except for x-axis (the direction of the light ) y, z, pitch, yaw
Coil drivers
- We use KAGRA coil driver; we need to cheek its performance.
Cables
Assembling [Tomaru-san, Ohishi, Yamamoto-san]
- On 21th, suspension frame, security frame, xy stages completed
Hanging
- Hanging jig preparation
Transfer Function Measurement
Transportation to the End [Prof. Kimura, Sawada-san, Tomaru-san(KEK),]
- KEK
Mirrors
- Two TAMA PRM mirrors (with about 50% and 90% reflectivities each, resultant contrast becomes 96%) were delivered to Kamioka on 12th Feb.
Mirror container [Tatsumi-san]
- Some plates once used for transportation of MC mirrors were used
- Two places of the container were cut to avoid standoffs.
=== Stages ====
- Condensation
- Vacuum compatibility?