* Connection between an LVDT driver and an ADC required a gender changer. Against the general policy of KAGRA electronics, the output connector of the LVDT driver is male (output should be female in the policy). Therefore, a gender changer was necessary. We should add a gender changer at the output of the LVDT driver. |
iKAGRA VIS electronics review
2016/2/4 15:00 -
Participants: Takahashi, Aso, Mark, Shoda, Raffaele, Akutsu, Okutomi, Fujii, Sekiguchi
Agenda
- PR3 electronics diagram (Mark)
- Discussion
- We went through the cables for PR3.
- Overview diagram was drawn on a white board (picture)
- Blue double circles indicates cables to be replaced from the test hanging (because of wrong pin-out).
- Red check marks mean we checked the corresponding cable
Missing parts -> nothing
- Connection between an LVDT driver and an ADC required a gender changer. Against the general policy of KAGRA electronics, the output connector of the LVDT driver is male (output should be female in the policy). Therefore, a gender changer was necessary. We should add a gender changer at the output of the LVDT driver.
- Overview diagram was drawn on a white board (picture)
- Future improvement
- We need an overview schematic like the one drawn in the meeting with links to more detailed documents.
- From the schematic, we should generate a table of necessary cables
- We should use this table to check if we have all the necessary cables
- We went through the cables for PR3.