Line 37: Line 37:
  * Updated shot noise calculation https://klog.icrr.u-tokyo.ac.jp/osl/?r=12772

ISC Meeting on 2020/03/09 14:00 - 15:00

Participants:

Zoom meeting: https://zoom.us/j/6676627462

Next meeting

TBD

back to Meetings page

Minutes

Agenda

  • We want to know what to do for O4 with birefringence and ITM transmission asymmetry
    • SRM reflectivity: 0% or 70% or 85%
    • Polarizers in PRC and SRC?
    • ITM transmission asymmetry: re-coating necessary?
  • Summary of measurements so far
  • Summary of calculations
  • List of measurements to be done before the vent for O4
    • Try DRFPMI locking for one more week.
    • Updated shot noise calculation https://klog.icrr.u-tokyo.ac.jp/osl/?r=12772

    • X-arm and Y-arm reflectivity/loss measurements with polarization considered
    • Re-measurement of CMRR for frequency noise https://klog.icrr.u-tokyo.ac.jp/osl/?r=13069

    • Measure power recycling gain for sidebands (POP 2f with FPMI locked and PRFPMI locked)
    • Measure MICH contrast defect with MICH locked and FPMI locked
    • OMC cavity scan to scan mode content of AS beam
    • Scattered light measurement in PRC by sweeping PRM and PR2/3 deferentially.
  • List of calculations to be done
    • DRFPMI LSC and ASC sensing matrix with birefringence and ITM transmission asymmetry
    • Frequency noise and intensity noise coupling with birefringence and ITM transmission asymmetry
    • MICH/PRCL/SRCL to DARM coupling with birefringence and ITM transmission asymmetry
  • O4 Schedule https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=9209

  • Executive summary (tentative)
    • We are not sure yet whether birefringence is limiting our sensitivity and stability. DRFPMI locking probably possible with current interferometer.
    • Putting polarizers inside PRC and SRC might improve the sensitivity and stability, but it might introduce some extra issue (thermal lens, extra loss, scattered light etc).
    • May be better to focus on characterization of the interferometer and the preparation of better ITMs rather than the preparation of polarizers.
    • It is always good to prepare for the unexpected if we have manpower and money. Preparing polarizers now might save our lives in the future.
    • Measurement listed above must be performed before the vent for O4 to characterize the current situation and to make a plan for O4.