= Transmission Monitor System (TMS)= | = Transmission Monitor System (TMS) = |
Line 20: | Line 20: |
=== Required channels === | === Requirements === For each TMS, * 2QPDs for 1064nm are required to monitor off-centering of the beam; one is for high gain, the other is for low gain. * 2QPDs for 1064nm are required to monitor pitch/yaw misalignments of the beam; one is for high gain, the other is for low gain. * 2QPDs for 532nm are required. * Remote stirring mirrors for QPDs are required at least in the initial commissioning phase. * Remote actuators for TMS itself would be required. * CCD monitors for monitoring the beam mode transmitting the ETM are required. * Optical levers (or equivalent local monitors) are required. * Suspension system will require more signal channels? * Shutters for high gain QPDs are required. ||<#FFFF00>Chamber Name||<#FFFF00>Type||<#FFFF00>Quantity||<#FFFF00>Purpose||<#FFFF00>Used By|| ||PRM||1064nm QPD||4||REFL port||MIF|| |
Transmission Monitor System (TMS)
Related subgroups: IOO, MIF, VAC, CRY, VIS...
Also called "beam reducing telescope (BRT)" in LCGT.
Contents
Scope
Transmission monitors for IR/Green beams.
General Description
Functions
- Collect 1064nm and 532nm light transmitting ETMs.
- Monitor the off-centering and pitch/yaw fluctuations of the ITMs/ETMs relative to the preselected optical path.
- Monitor the shape of the beam modes that transmit the arm cavity.
Requirements
For each TMS,
- 2QPDs for 1064nm are required to monitor off-centering of the beam; one is for high gain, the other is for low gain.
- 2QPDs for 1064nm are required to monitor pitch/yaw misalignments of the beam; one is for high gain, the other is for low gain.
- 2QPDs for 532nm are required.
- Remote stirring mirrors for QPDs are required at least in the initial commissioning phase.
- Remote actuators for TMS itself would be required.
- CCD monitors for monitoring the beam mode transmitting the ETM are required.
- Optical levers (or equivalent local monitors) are required.
- Suspension system will require more signal channels?
- Shutters for high gain QPDs are required.
Chamber Name |
Type |
Quantity |
Purpose |
Used By |
PRM |
1064nm QPD |
4 |
REFL port |
MIF |
Naming Conventions
TBD
Locations
- At the dedicated chambers in back of ETM chambers