Line 2: Line 2:
<<TableOfContents(3)>>
Line 10: Line 10:

L_BS_PR2 is the separation between the center of BS chamber and the center of PR2 chamber.
The following table shows the the Folding parameters with three different cases of L_BS_PR2.
Case 1 is the default plan. Case 2 adds 2m to the BS-PR2 separation, and Case 3 adds 3m.

Note that the arm cavity g-factor is assumed to be negative (1.5km-1.9km), which usually gives slightly smaller beam sizes on the PRM.

Layout related issues

Location of PR2 tank

Is it possible to put the PR2 and SR2 vacuum chambers further away from the BS chamber ? The minimum separation is 2.7m (center-to-center). However, in order to improve the workability around the BS chamber, we want increase the separation.

MIF Diagram

Comparison Table

L_BS_PR2 is the separation between the center of BS chamber and the center of PR2 chamber. The following table shows the the Folding parameters with three different cases of L_BS_PR2. Case 1 is the default plan. Case 2 adds 2m to the BS-PR2 separation, and Case 3 adds 3m.

Note that the arm cavity g-factor is assumed to be negative (1.5km-1.9km), which usually gives slightly smaller beam sizes on the PRM.

L_BS_PR2

Folding Angle

Lp1

Lp2

Lp3

PRM ROC

PRM Beam Size

PR2 ROC

PR2 Beam Size

PR3 ROC

PR3 Beam Size

Case 1

2.7m

0.63deg

14.76m

12.07m

14.76m

310.42m

4.05mm

-3.13m

4.05mm

27.01m

35.43mm

Case 2

4.7m

0.67deg

14.09m

11.40m

16.10m

295.3m

3.95mm

-2.88m

3.95mm

25.45m

35.46mm

Case 3

5.7m

0.69deg

13.76m

11.06m

16.76m

287.75m

3.91mm

-2.75m

3.91mm

24.67m

35.48mm