Layout related issues

Location of PR2 tank

Is it possible to put the PR2 and SR2 vacuum chambers further away from the BS chamber ? The minimum separation is 2.7m (center-to-center). However, in order to improve the workability around the BS chamber, we want increase the separation.

MIF Diagram

Comparison Table

L_BS_PR2 is the separation between the center of BS chamber and the center of PR2 chamber. The following table shows the the Folding parameters with three different cases of L_BS_PR2. Case 1 is the default plan. Case 2 adds 2m to the BS-PR2 separation, and Case 3 adds 3m.

L_BS_PR2

Folding Angle

Lp1

Lp2

Lp3

PRM ROC

PRM Beam Size

PR2 ROC

PR2 Beam Size

PR3 ROC

PR3 Beam Size

Case 1

2.7m

0.63deg

14.76m

12.07m

14.76m

310.42m

4.05mm

-3.13m

4.05mm

27.01m

35.43mm

Case 2

4.7m

0.67deg

14.09m

11.40m

16.10m

295.3m

3.95mm

-2.88m

3.95mm

25.45m

35.46mm

Case 3

5.7m

0.69deg

13.76m

11.06m

16.76m

287.75m

3.91mm

-2.75m

3.91mm

24.67m

35.48mm