Line 17: Line 17:
X-end PCal = X-end PCal =
Line 19: Line 19:
 * TxPD1 repairing (THS4631 seems broken)
  * Asked Bin-Hua to bring parts.
 * OFS2 closed loop functionality repairing (relay is suspected)
  * Asked Bin-Hua to bring parts.
 * Interlock equippement
  * Switch is equipped to modules.
  * Connector between chassis and laser system is repaired.
 * Unstable D-sub connection at Tx module
  * Some connectors are dead.
 * Cable arrangement
 * Laser remote control
  * PC will be setup.
  * Same script as in Y-end may be applicable.
Urgent
 * Excitation port noise

2nd-urgent

 * Stray light
  * Tx PDs detect light even if the AOM input is almost 0.
  * Put dumpers

~11/3 first test to inject PCal laser to the X-end test mass

 * Rx module alignment
  * QPD alignment
 * OL/CL TF measurement
  * OFS board gain seems strange. Saturation around 13 dB.
  * Strange bump in AA/AI (~6 kHz)

~X-arm commissioning line injection
Line 33: Line 39:
  * Picomoter setting
 * Periscope install
  * Wall between duct and modules
   * Ordered. Will arrive Kamioka on 9/10.
  * Set duct and wall.
  * Green laser preparation
   *
 * OL/CL TF measurement
 * Preparation for X-arm commissioning calibration line injection
  * Guardian

~O3
Line 44: Line 47:
 * VGA noise measurement
 * Excitation noise measurement
  * DGS part seems not noisy, but cause large noise when connected to the OFS.
 * Peak stability test
 * Shot noise
 * Offset input noise
 * Noise model

 * Unstable D-sub connection at Tx module
  * Some connectors are dead.
 * OFS upgrade
 * Operation tool preparation

Bonus work

 * AOM current
  * Sometimes over 0.5 V ?
 * Take care about grounding

Pre-categorization


= Y-end PCal =

 * Periscope re-installation

= Laser power calibration setting in Toyama Univ. =
 
~Y-arm commissioning

 * Prepare cross calibration setup and calibrate an integrating sphere

Detailed task list of PCal is listed here.

X-end PCal

Urgent

  • Excitation port noise

2nd-urgent

  • Stray light
    • Tx PDs detect light even if the AOM input is almost 0.
    • Put dumpers

~11/3 first test to inject PCal laser to the X-end test mass

  • Rx module alignment
    • QPD alignment
  • OL/CL TF measurement
    • OFS board gain seems strange. Saturation around 13 dB.
    • Strange bump in AA/AI (~6 kHz)

~X-arm commissioning line injection

  • BPM test
  • Preparation for X-arm commissioning calibration line injection
    • Guardian

~O3

  • RPN measurement
  • Laser noise measurement
  • Harmonics noise measurement
  • VGA noise measurement
  • Excitation noise measurement
    • DGS part seems not noisy, but cause large noise when connected to the OFS.
  • Peak stability test
  • Shot noise
  • Offset input noise
  • Noise model
  • Unstable D-sub connection at Tx module
    • Some connectors are dead.
  • OFS upgrade
  • Operation tool preparation

Bonus work

  • AOM current
    • Sometimes over 0.5 V ?
  • Take care about grounding

Pre-categorization

Y-end PCal

  • Periscope re-installation

Laser power calibration setting in Toyama Univ.

~Y-arm commissioning

  • Prepare cross calibration setup and calibrate an integrating sphere

PcalTask (last edited 2018-11-22 09:41:33 by chihiro.kozakai)