  1. In the corresponding directory for the suspension you are working with, create a new directory with the date as a name. For example,
    • /kagra/Dropbox/Subsystems/VIS/TypeBData/SRM/OPLEV/measurements/20191208/
  2. Into the newly created directory copy the corresponding template files from
    • /kagra/Dropbox/Subsystems/VIS/TypeBData/SRM/OPLEV/templates/ The templates are different for each suspension, therefore make sure you use the appropriate directory
  3. Align the optic into a convenient orientation.
    • Use Guardian to set the suspension in IM_DAMP mode.
    • Align the oplev beam to the centre of the QPDs. Use the IM test channels to move the marionette such that the readout of the oplev in Normalized Displacement Units (ndu) are close to zero.
  4. Use diaggui to open the file SRM_TM_OPLEV_L_PREDIAG.xml, execute it, wait until it finishes and then save the results. Mind the name, it follows the suspension.
  5. Use diaggui to open the file SRM_TM_OPLEV_P_PREDIAG.xml, execute it, wait until it finishes and then save the results. Mind the name, it follows the suspension.
  6. Use diaggui to open the file SRM_TM_OPLEV_Y_PREDIAG.xml, execute it, wait until it finishes and then save the results. Mind the name, it follows the suspension.
  7. Open the file SRM_TM_OPLEV_L_PREDIAG.xml.