CLIO full digital for locked Fabry-Perot configuration
- long term measurement追加 ==
- laser temp
- laser current
- temperature at end
- humidity at end
- dust at end
CLIO digital block diagram: Locked-FP
channel number
ADC 16kHz
- MC REFL: (DC, RF) = 2ch
- MC Trans: 1ch
- MC_F :1ch
LSC: (I&Q,DC) x2 = 6ch
- Trans QPD: 4quad x2 = 8ch
- Total 18ch
- (Vertex 10ch, X end 4ch, Y end 4ch)
ADC 2kHz
MC WFS(I&Q,DC)12ch x2 = 24ch
IFO WFS(I&Q,DC)12ch x4 = 48ch
- Total 72ch
- (Vertex 72ch, X end 0ch, Y end 0ch)
ADC 256Hz
- Laser power, temp, current 3
- Seismic 3
- Acoustic 3
- Temperature 3ch
- Humidity 1ch
- Particle 1ch
- Total 14ch
- (Vertex 8ch, X end 3ch, Y end 3ch)
DAC
- Laser thermal 1ch
- Laser steering: 4ch
- MC: 3 SUS x 4 coils = 12ch
- IFO: 5 SUS x 4 coils = 20ch
- total 45ch
- (Vertex 37ch, X end 4ch, Y end 4ch)
Binary I/O
- 8SUS x 4coils = 32ch
- MC fast gain, swirch, boost, test
- Yarm gain, swirch, boost, test