Differences between revisions 106 and 107
Revision 106 as of 2019-12-17 15:30:01
Size: 4581
Comment:
Revision 107 as of 2019-12-17 15:33:04
Size: 4417
Comment:
Deletions are marked like this. Additions are marked like this.
Line 30: Line 30:
= How to check Guardian =
[[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|how to use Guardian]]

For collaborator shifts


Collaboration shift routine tasks

  1. Write "Log book" (JGW-E1911124) on the desk in the control room.

    • Check Guardian status for 5 items, LSC, VIS, IO, Green laser and ALS. Instruction is in below.
    • Put check mark every 2 hours.
    • Write down everything you noticed.
      • Example: lock lost, earthquake, somebody entered mine etc.
  2. When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state
    • ---> call operation shift members

      • contact person on the day is listed with PHONE NUMBER around white board in the control room or see the link JGW-M1910914 (no phone number in this list)

  3. Please write klog: http://klog.icrr.u-tokyo.ac.jp/osl/ before you leave control room.

    • Writing klog is very important. This is only one tool which can show the status in the control room to outside.

    • Account name and password are shown on the white board in the control room.
  4. Leave message on "Takeover note" (JGW-T1605032) on the desk in the control room if you have something to tell to the next shift.

  5. Improve this wiki page if you had something you did not understand well and you found what it is.
    1. Search "KAGRA wiki" in google and access to KAGRA top page.
    2. Go to ERun top page on the second line of KAGRA Top page, and go to "Collaborator shift manuals". That page is this manual.
    3. Login KAGRA wiki, and press "Edit" button around the top left.
      • Account name and password are shown on the white board in the control room.



Accessing to the guardian screen

  • On the desktop, you can click the sitemap.
  • 112.png

  • In the sitemap, you can click on the link for the guardian overview screen.
  • 116.png

  • Meaning of the background color
  • guardian_explanation.png

Check list of the interferometer

Check if the LSC_LOCK Guardian at OBSERVATION state

LCSLOCK.png

Check if all of VIS Guardians are at ALIGNED state

Look at Guardian Overview

  • VIS guardian.png

OR you can see the VIS status on the Monitor

  • VIS guardian 2.png

  • If the Guardian is at TRIPPED state, you should reset Watch Dogs (WDs), OR call operation shift members (contact person TBD).

Check the IO Guardian

IO guardian.png

IO Guardian should at PROVIDING_STABLE_LIGHT state, if not call operation shift members ASAP!!!(contact person TBD).

Check the Green Laser status

  • See if the shutter is on or not. If the background color is green, the shutter is ok/open (red means the shutter is closed).

shutter.png

  • see if K1: ALS-X (Y)_ARM_INPUT_OUT is more than 0.7

0.7.png

  • Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state

PLLXY1.png


Control Room Overview

  • IFO Monitor: shows the status of the Main Interferometer
  • IFO1.png

  • CCD Monitor:
  • CCD.jpg

  • DAQ/VIS status:
  • vis status.png

  • PEM Monitor: Room temperature/ Vacuum pressure
  • Pem.jpg

  • Time: show JST/UTC/GPS time
  • GPS t ime.jpg

KAGRA/EngineeringRun/ER191217_24/Shift (last edited 2020-04-07 12:03:16 by OsamuMiyakawa)