= Collaborator shift manual = ---- == What you should do == 1. Write "Log book" on the desk in the control room. * Put check mark every 2 hours. * Write down everything you noticed. * Example: lock lost, earthquake, somebody entered mine etc. * When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state ---> call operation shift members * contact person on the day is listed around white board in the control room or see the ling below; * https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=10914 * Write down "Takeover note" on the desk in the control room. * If you have something which should be known to the next shift, please leave messages in the notebook. * Please write klog before you leave control room. * Writing klog is very important. This is only one tool which can show the status in the control room to outside. ---- |
|
Line 2: | Line 20: |
* [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=10914| Operator Shift]] | |
Line 3: | Line 22: |
* When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state ---> call operation shift members. | |
Line 9: | Line 27: |
== Control Room Overview == * IFO Monitor: shows the status of the Main Interferometer |
= Control Room Overview = * IFO Monitor: shows the status of the Main Interferometer |
Line 12: | Line 30: |
* CCD Monitor: . {{attachment:ccd monitor.png||height="681"}} * DAQ/VIS status: |
* CCD Monitor: . {{attachment:CCD.jpg||height="881"}} * DAQ/VIS status: |
Line 16: | Line 34: |
* PEM Monitor: Room temperature/ Vacuum pressure | * PEM Monitor: Room temperature/ Vacuum pressure |
Line 18: | Line 36: |
* Time: show JST/UTC/GPS time | * Time: show JST/UTC/GPS time |
Line 22: | Line 40: |
== How to use Guardian == | = How to check Guardian = [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|how to use Guardian]] |
Line 24: | Line 44: |
. In the desktop, you can click on the sitemap. . {{attachment:111.png||width="330"}} . {{attachment:112.png||width="430"}} |
. On the desktop, you can click the sitemap. . {{attachment:112.png||width="630"}} |
Line 28: | Line 48: |
. {{attachment:114.png||width="730"}} | . {{attachment:116.png||width="730"}} . Meaning of the background color |
Line 36: | Line 57: |
== Check list of the interferometer == . Look at the Guardian overview, check if all of VIS Guardians are in ALIGNED state. |
= Check list of the interferometer = === Check if the LSC_LOCK Guardian at OBSERVATION state === {{attachment:LCSLOCK.png||height="",width="880"}} === Check if all of VIS Guardians are at ALIGNED state === Look at Guardian Overview |
Line 39: | Line 64: |
OR you can see the VIS status on the Monitor | OR you can see the VIS status on the Monitor |
Line 41: | Line 67: |
*If the Guardian is in TRIPPED state, you should [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|reset Watch Dogs (WDs)]], OR call operation shift members. | * If the Guardian is at TRIPPED state, you should [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|reset Watch Dogs (WDs)]], OR call operation shift members (contact person TBD). === Check the IO Guardian === {{attachment:IO guardian.png||height="",width="680"}} IO Guardian should at PROVIDING_STABLE_LIGHT state, if not call operation shift members '''ASAP!!!'''(contact person TBD). === Check the Green Laser status === * See if the shutter is on or not. If the background color is green, the shutter is ok/open (red means the shutter is closed). {{attachment:shutter.png||height="",width="680"}} |
Line 44: | Line 80: |
* see if K1: ALS-X (Y)_ARM_INPUT_OUT is more than 0.7 | |
Line 45: | Line 82: |
<<BR>>If "OFF", check the guardian overview screen<<BR>> {{attachment:guardian_howto3.png||height="281",width="434"}} <<BR>>Check the state of LSC_MICH. First check if the request state of LSC_MICH is OPERATION. If it is not, request OPERATION.<<BR>> {{attachment:guardian_howto4.png||height="281",width="434"}} . If the state is ''READY, ENGAGE_SERVO, ENGAGE_BOOST, MICH_LOCKED, CALIBRATION_INJECTION'', or ''ACTUATORCAL_INJECTION'', the interferometer is now on the way to be OPERATION state. It will be ready soon. <<BR>> /!\ ''' If it does not enter OPERATION state in a few mins. Call the interferometer experts.'''<<BR>> Check the status of ''IMC_LOCK, VIS_BS, VIS_PR2, VIS_PR3''. (Note that ''VIS_PRM'' is not used for now, so just ignore it) <<BR>>For IMC_LOCK, every thing should be green and IMC should be back automatically. <<BR>> {{attachment:IMC_guardian.png||height="293",width="453"}} <<BR>> For example, this is how the IMC_LOCK guardian should normally look like.<<BR>> {{attachment:IMC_guardian_status.png||height="281",width="407"}} <<BR>> /!\ '''If it does not come back in ~ 10 mins, call interferometer/IOO experts.'''<<BR>> For VIS_PR2 and VIS_BS, the state should be ALIGNED. If inot, a notification "please reset Watch Dog!" will show up, and the state is in TRIPPED. In that case, go to the next section to know how to reset Watch Dog. <<BR>> For VIS_PR3, the state should be DAMPED. If it is not, a notification "please reset Watch Dog!" will show up, and the state is in TRIPPED. In that case, go to the next section to know how to reset Watch Dog. |
{{attachment:0.7.png||height="",width="680"}} * Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state {{attachment:PLLXY1.png||height="",width="880"}} |
Collaborator shift manual
What you should do
- Write "Log book" on the desk in the control room.
- Put check mark every 2 hours.
- Write down everything you noticed.
- Example: lock lost, earthquake, somebody entered mine etc.
- When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state
---> call operation shift members
- contact person on the day is listed around white board in the control room or see the ling below;
- Write down "Takeover note" on the desk in the control room.
- If you have something which should be known to the next shift, please leave messages in the notebook.
- Please write klog before you leave control room.
- Writing klog is very important. This is only one tool which can show the status in the control room to outside.
- Collaboration shift routine tasks
Control Room Overview
- IFO Monitor: shows the status of the Main Interferometer
- CCD Monitor:
- DAQ/VIS status:
- PEM Monitor: Room temperature/ Vacuum pressure
- Time: show JST/UTC/GPS time
How to check Guardian
Accessing to the guardian screen
- On the desktop, you can click the sitemap.
- In the sitemap, you can click on the link for the guardian overview screen.
- Meaning of the background color
Knowing the current state and requested state
How to request a state
Click on the gray bar in the middle.
Then, choose the desired state.
Check list of the interferometer
Check if the LSC_LOCK Guardian at OBSERVATION state
Check if all of VIS Guardians are at ALIGNED state
Look at Guardian Overview
OR you can see the VIS status on the Monitor
If the Guardian is at TRIPPED state, you should reset Watch Dogs (WDs), OR call operation shift members (contact person TBD).
Check the IO Guardian
IO Guardian should at PROVIDING_STABLE_LIGHT state, if not call operation shift members ASAP!!!(contact person TBD).
Check the Green Laser status
- See if the shutter is on or not. If the background color is green, the shutter is ok/open (red means the shutter is closed).
- see if K1: ALS-X (Y)_ARM_INPUT_OUT is more than 0.7
- Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state