2883
Comment:
|
3849
|
Deletions are marked like this. | Additions are marked like this. |
Line 1: | Line 1: |
= For collaborator shifts = |
|
Line 2: | Line 4: |
---- /!\ '''Edit conflict - other version:''' ---- |
|
Line 6: | Line 6: |
---- /!\ '''Edit conflict - your version:''' ---- | ---- == Collaboration shift routine tasks == |
Line 8: | Line 9: |
---- /!\ '''End of edit conflict''' ---- | 1. Write "Log book" on the desk in the control room. * Put check mark every 2 hours. * Write down everything you noticed. * Example: lock lost, earthquake, somebody entered mine etc. * When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state ---> call operation shift members * contact person on the day is listed around white board in the control room or see the ling below; * https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=10914 * Write down "Takeover note" on the desk in the control room. * If you have something which should be known to the next shift, please leave messages in the notebook. * Please write klog before you leave control room. * Writing klog is very important. This is only one tool which can show the status in the control room to outside. ---- |
Line 10: | Line 24: |
* When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state ---> call operation shift members (contact person TBD). * Collaboration shift routine tasks |
|
Line 16: | Line 27: |
== Control Room Overview == | = Control Room Overview = |
Line 20: | Line 31: |
. {{attachment:ccd monitor.png||height="681"}} | . {{attachment:CCD.jpg||height="881"}} |
Line 29: | Line 40: |
== How to use Guardian == | = How to check Guardian = [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|how to use Guardian]] |
Line 31: | Line 44: |
. In the desktop, you can click on the sitemap. | . On the desktop, you can click the sitemap. |
Line 34: | Line 48: |
. {{attachment:114.png||width="730"}} | . {{attachment:116.png||width="730"}} . Meaning of the background color |
Line 42: | Line 57: |
== Check list of the interferometer == ==== Look at the Guardian overview, check if all of VIS Guardians are in ALIGNED state ==== |
= Check list of the interferometer = === Check if the LSC_LOCK Guardian at OBSERVATION state === {{attachment:LCSLOCK.png||height="",width="880"}} === Check if all of VIS Guardians are at ALIGNED state === Look at Guardian Overview |
Line 48: | Line 67: |
* If the Guardian is in TRIPPED state, you should [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|reset Watch Dogs (WDs)]], OR call operation shift members (contact person TBD). | * If the Guardian is at TRIPPED state, you should [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|reset Watch Dogs (WDs)]], OR call operation shift members (contact person TBD). |
Line 50: | Line 69: |
==== Check the IO Guardian ==== |
=== Check the IO Guardian === |
Line 54: | Line 72: |
IO Guardian should in PROVIDING_STABLE_LIGHT state, if not call operation shift members '''ASAP!!!'''(contact person TBD). | IO Guardian should at PROVIDING_STABLE_LIGHT state, if not call operation shift members '''ASAP!!!'''(contact person TBD). |
Line 56: | Line 74: |
==== Check the Green Laser status ==== * see if the shutter is on or not |
=== Check the Green Laser status === * See if the shutter is on or not. If the background color is green, the shutter is ok/open (red means the shutter is closed). |
Line 60: | Line 78: |
Line 65: | Line 84: |
* Check if ALS_PLLX/ALS_PLLY Guardian is on PLL_LOCK state | * Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state |
Line 68: | Line 87: |
For collaborator shifts
Collaboration shift routine tasks
- Write "Log book" on the desk in the control room.
- Put check mark every 2 hours.
- Write down everything you noticed.
- Example: lock lost, earthquake, somebody entered mine etc.
- When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state
---> call operation shift members
- contact person on the day is listed around white board in the control room or see the ling below;
- Write down "Takeover note" on the desk in the control room.
- If you have something which should be known to the next shift, please leave messages in the notebook.
- Please write klog before you leave control room.
- Writing klog is very important. This is only one tool which can show the status in the control room to outside.
Control Room Overview
- IFO Monitor: shows the status of the Main Interferometer
- CCD Monitor:
- DAQ/VIS status:
- PEM Monitor: Room temperature/ Vacuum pressure
- Time: show JST/UTC/GPS time
How to check Guardian
Accessing to the guardian screen
- On the desktop, you can click the sitemap.
- In the sitemap, you can click on the link for the guardian overview screen.
- Meaning of the background color
Knowing the current state and requested state
How to request a state
Click on the gray bar in the middle.
Then, choose the desired state.
Check list of the interferometer
Check if the LSC_LOCK Guardian at OBSERVATION state
Check if all of VIS Guardians are at ALIGNED state
Look at Guardian Overview
OR you can see the VIS status on the Monitor
If the Guardian is at TRIPPED state, you should reset Watch Dogs (WDs), OR call operation shift members (contact person TBD).
Check the IO Guardian
IO Guardian should at PROVIDING_STABLE_LIGHT state, if not call operation shift members ASAP!!!(contact person TBD).
Check the Green Laser status
- See if the shutter is on or not. If the background color is green, the shutter is ok/open (red means the shutter is closed).
- see if K1: ALS-X (Y)_ARM_INPUT_OUT is more than 0.7
- Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state