4409
Comment:
|
4864
|
Deletions are marked like this. | Additions are marked like this. |
Line 7: | Line 7: |
---- | ------ |
Line 10: | Line 10: |
1. Write "Log book" on the desk in the control room. | 1. When FPMI cannot keep locked state for over 10 minutes or cannot got to lock state ---> call operator shift members * contact person (operator shift) on the day is listed with '''PHONE NUMBER''' around white board in the control room or see the link [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=10914|JGW-M1910914]] (no phone number in this list) * Write "Log book" ([[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=11124|JGW-E1911124]]) on the desk in the control room. * Check Guardian status for 5 items, LSC, VIS, IO, Green laser and ALS. Instruction is in below. |
Line 14: | Line 19: |
* When FPMI cannot keep lock state for over 10 minutes or cannot got to lock state ---> call operation shift members * contact person on the day is listed with '''PHONE NUMBER''' around white board in the control room or see the link [[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=10914|JGW-M1910914]] (no phone number in this list) * Improve this wiki page if you have something you do now understand well. * Account name and password are shown on the white board in the control room. Login KAGRA wiki, access to this page and press "Edit" button around the top left. * Write down "Takeover note" on the desk in the control room. * If you have something which should be known to the next shift, please leave messages in the notebook. |
|
Line 23: | Line 21: |
* Account name and password are shown on the white board in the control room. * Leave message on "Takeover note" ([[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5032|JGW-T1605032]]) on the desk in the control room if you have something to tell to the next shift. * Improve this wiki page if you had something you did not understand well and you found what it is. 1. Search "KAGRA wiki" in google and access to KAGRA top page. * Go to ERun top page on the second line of KAGRA Top page, and go to "Collaborator shift manuals". That page is this manual. * Login KAGRA wiki, and press "Edit" button around the top left. * Account name and password are shown on the white board in the control room. |
|
Line 25: | Line 30: |
------ | |
Line 27: | Line 32: |
== Control Room Overview == * IFO Monitor: shows the status of the Main Interferometer . {{attachment:IFO1.png||width="630"}} * CCD Monitor: . {{attachment:CCD.jpg||height="881"}} * DAQ/VIS status: . {{attachment:vis status.png||height="681"}} * PEM Monitor: Room temperature/ Vacuum pressure . {{attachment:Pem.jpg||width="630"}} * Time: show JST/UTC/GPS time . {{attachment:GPS t ime.jpg||width="430"}} |
== 1. Interferometer locking status screen == * Please check the interferometer locking screen at the front monitor in the control room first. If the status is not "LOCKED", something might happen. * If FPMI cannot keep "LOCKED" state for over 10 minutes or cannot got back to "LOCKED" state, please contact operator shift. {{attachment:IFOstatus_s.png||width="800"}} |
Line 40: | Line 39: |
------ == 2. Check list of the interferometer == |
|
Line 41: | Line 42: |
= How to check Guardian = [[http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Commissioning/Phase1/Operation/ShiftManual#How_to_reset_Watch_Dogs|how to use Guardian]] === Accessing to the guardian screen === . On the desktop, you can click the sitemap. |
=== How to access to the guardian screen === *On the desktop, you can click the sitemap. |
Line 46: | Line 45: |
. {{attachment:112.png||width="630"}} . In the sitemap, you can click on the link for the guardian overview screen. . {{attachment:116.png||width="730"}} . Meaning of the background color . {{attachment:guardian_explanation.png||width=""}} === Knowing the current state and requested state === . {{attachment:guardian_howto4.png||height="281",width="434"}} |
{{attachment:112.png||width="800"}} |
Line 54: | Line 47: |
=== How to request a state === . Click on the gray bar in the middle.<<BR>> {{attachment:guardian_howto.png||height="298",width="433"}} <<BR>> Then, choose the desired state.<<BR>> {{attachment:guardian_howto2.png||height="299",width="434"}} |
*In the sitemap, you can click on the link for the guardian overview screen. {{attachment:116.png||width="800"}} |
Line 57: | Line 50: |
= Check list of the interferometer = === Check if the LSC_LOCK Guardian at OBSERVATION state === {{attachment:LCSLOCK.png||height="",width="880"}} |
=== 2-1. Check if the LSC_LOCK Guardian at OBSERVATION state === |
Line 61: | Line 52: |
=== Check if all of VIS Guardians are at ALIGNED state === | * Please check the "LSC_LOCK" is in "OBSERVATION" mode. If not, please contact operator shift. {{attachment:116.png||width="800"}} === 2-2. Check if all of VIS Guardians are at ALIGNED state === |
Line 63: | Line 58: |
{{attachment:VIS guardian.png||height="",width="880"}} | {{attachment:VIS guardian.png||height="",width="800"}} |
Line 66: | Line 61: |
{{attachment:VIS guardian 2.png||height="",width="680"}} | {{attachment:VIS guardian 2.png||height="",width="700"}} |
Line 69: | Line 64: |
=== Check the IO Guardian === {{attachment:IO guardian.png||height="",width="680"}} |
=== 2-3. Check the IO Guardian === {{attachment:IO guardian.png||height="",width="800"}} |
Line 74: | Line 69: |
=== Check the Green Laser status === | === 2-4. Check the Green Laser status === |
Line 77: | Line 72: |
{{attachment:shutter.png||height="",width="680"}} | {{attachment:shutter.png||height="",width="800"}} |
Line 84: | Line 79: |
* Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state | === 2-5. Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state === |
Line 87: | Line 82: |
------ == Control Room Overview == * IFO Monitor: shows the status of the Main Interferometer . {{attachment:IFO1.png||width="800"}} * CCD Monitor: . {{attachment:CCD.jpg||height="1000"}} * DAQ/VIS status: . {{attachment:vis status.png||height="800"}} * PEM Monitor: Room temperature/ Vacuum pressure . {{attachment:Pem.jpg||width="800"}} * Time: show JST/UTC/GPS time . {{attachment:GPS t ime.jpg||width="800"}} |
For collaborator shifts
Operator shift table in JGW-M1910914
Some information for collaborator shifts in JGW-M1809380
Collaboration shift routine tasks
- When FPMI cannot keep locked state for over 10 minutes or cannot got to lock state
---> call operator shift members
contact person (operator shift) on the day is listed with PHONE NUMBER around white board in the control room or see the link JGW-M1910914 (no phone number in this list)
Write "Log book" (JGW-E1911124) on the desk in the control room.
- Check Guardian status for 5 items, LSC, VIS, IO, Green laser and ALS. Instruction is in below.
- Put check mark every 2 hours.
- Write down everything you noticed.
- Example: lock lost, earthquake, somebody entered mine etc.
Please write klog: http://klog.icrr.u-tokyo.ac.jp/osl/ before you leave control room.
Writing klog is very important. This is only one tool which can show the status in the control room to outside.
- Account name and password are shown on the white board in the control room.
Leave message on "Takeover note" (JGW-T1605032) on the desk in the control room if you have something to tell to the next shift.
- Improve this wiki page if you had something you did not understand well and you found what it is.
- Search "KAGRA wiki" in google and access to KAGRA top page.
- Go to ERun top page on the second line of KAGRA Top page, and go to "Collaborator shift manuals". That page is this manual.
- Login KAGRA wiki, and press "Edit" button around the top left.
- Account name and password are shown on the white board in the control room.
1. Interferometer locking status screen
- Please check the interferometer locking screen at the front monitor in the control room first. If the status is not "LOCKED", something might happen.
- If FPMI cannot keep "LOCKED" state for over 10 minutes or cannot got back to "LOCKED" state, please contact operator shift.
2. Check list of the interferometer
How to access to the guardian screen
- On the desktop, you can click the sitemap.
- In the sitemap, you can click on the link for the guardian overview screen.
2-1. Check if the LSC_LOCK Guardian at OBSERVATION state
- Please check the "LSC_LOCK" is in "OBSERVATION" mode. If not, please contact operator shift.
2-2. Check if all of VIS Guardians are at ALIGNED state
Look at Guardian Overview
OR you can see the VIS status on the Monitor
If the Guardian is at TRIPPED state, you should reset Watch Dogs (WDs), OR call operation shift members (contact person TBD).
2-3. Check the IO Guardian
IO Guardian should at PROVIDING_STABLE_LIGHT state, if not call operation shift members ASAP!!!(contact person TBD).
2-4. Check the Green Laser status
- See if the shutter is on or not. If the background color is green, the shutter is ok/open (red means the shutter is closed).
- see if K1: ALS-X (Y)_ARM_INPUT_OUT is more than 0.7
2-5. Check if ALS_PLLX/ALS_PLLY Guardian is at PLL_LOCK state
Control Room Overview
- IFO Monitor: shows the status of the Main Interferometer
- CCD Monitor:
- DAQ/VIS status:
- PEM Monitor: Room temperature/ Vacuum pressure
- Time: show JST/UTC/GPS time