Differences between revisions 35 and 36
Revision 35 as of 2016-03-24 11:49:06
Size: 3017
Comment:
Revision 36 as of 2016-03-24 11:50:48
Size: 3023
Comment:
Deletions are marked like this. Additions are marked like this.
Line 4: Line 4:
 * Record K1:PSL-PMC_TRANS_DC_OUTPUT  * Record K1:PSL-PMC_TRANS_DC_OUTPUT.
Line 6: Line 6:
 * Record K1:IMC-MCL_TRANS_OUTPUT  * Record K1:IMC-MCL_TRANS_OUTPUT.
Line 8: Line 8:
 * Record IMC oplev values.
 * Record IMC oplev values.{{{
Line 16: Line 15:
}}}

IFO Operation Manual

CHECK LIST

  • Record K1:PSL-PMC_TRANS_DC_OUTPUT.
    • If it is ~6, OK. If not, ask Miyakawa, Kawamura, or Enomoto to manually lock PMC.
  • Record K1:IMC-MCL_TRANS_OUTPUT.
    • If it is ~0.4, OK. If not, check IMC alignment or guardian (see detail below).
  • Record IMC oplev values.

     K1:VIS-MCE_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCE_TM_OPLEV_YAW_OUTPUT
     K1:VIS-MCI_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCI_TM_OPLEV_YAW_OUTPUT
     K1:VIS-MCO_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCO_TM_OPLEV_YAW_OUTPUT

General

  • Type "mk" on terminal and you will open a "sitemap" screen

PSL

Lock PMC

  • Open the "PSL PMC" medm screen from sitemap

PMC medm

  • Open StripTool /users/striptool/pmc.stp (link is put on the PMC screen)

  • Find PMC_TRANS Camera
  • Open the Guardian screen (link on the PMC screen)

PMC Guardian

  • 1. Request "PSL_LOCKED"
  • 2. Execute (Choose "EXEC")
  • Make sure that PMC is locked with the StripTool, PMC_TRANS camera, and the Guardian. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 6, PMC is locked. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 0, unlocked.

IMC

Lock IMC

  • Check MCI, MCO and MCE (link to VIS page. DC alignment values, Watchdog, Dackill)
  • Open IMC medm from sitemap

IMC medm

IMC Guardian

  • The usage of the IMC Guardian is almost same as one of PMC. Request "IMC_LOCKED" and execute.
  • NOTE: If PMC is unlocked, you will see a message "PSL_is_unlocked!" in USERMSG line in the IMC Guardian. When you see it, check if PMC is locked or not
  • Open StripTool /users/stiptool/IMC.stp

  • Find MCE_TRANS camera
  • Make sure IMC is locked with the StripTool, MCE_TRANS camera, and the Guardian. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.40 - 0.46 , IMC is locked. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.020, unlocked.

Main IFO

Check VIS

  • Check PR3, BS, IX and IY (link to VIS page)

Initial Alignment

Lock Michelson IFO

  • Establish the lock procedure
  • Prepare Guardian and put the manual here
  • Prepare StripTool (list of check channels)

Check DAQ

  • MonitCheck Check if all the DAQ system is working properly

  • DaqCheck1 Check if the real time models are sending data to DAQ

  • DaqCheck2 Check if the GW channel is properly recorded in the frames

  • Check Detchar screen?
  • Check glitch rate (prepare a script?)
  • Check noise lines (prepare a script?)

In the case of IFO/subsystem failure or emergency

KAGRA/Observation/iKAGRA/Manual/MIF (last edited 2016-03-25 10:56:26 by YutaMichimura)