Differences between revisions 48 and 49
Revision 48 as of 2016-03-25 09:29:10
Size: 4344
Comment:
Revision 49 as of 2016-03-25 10:34:59
Size: 4683
Comment:
Deletions are marked like this. Additions are marked like this.
Line 2: Line 2:

Take logbook if
 * PSL PMC looses lock
 * IMC does not lock even if PSL PMC is locking
 * IMC keeps locking and unlocking for more than an hour
 * Main interferometer does not lock even if IMC is locked
 * Main interferometer does not lock for more than 20 mins
 * Each mirror’s oplev feedback signals goe more than 6000 cts.

IFO Operation Manual

Take logbook if

  • PSL PMC looses lock
  • IMC does not lock even if PSL PMC is locking
  • IMC keeps locking and unlocking for more than an hour
  • Main interferometer does not lock even if IMC is locked
  • Main interferometer does not lock for more than 20 mins
  • Each mirror’s oplev feedback signals goe more than 6000 cts.

CHECK LIST

  • Record K1:PSL-PMC_TRANS_DC_OUTPUT.
    • If it is ~6, OK.
    • If not, ask Miyakawa, Kawamura, or Enomoto to manually lock PMC.
  • Check the state of IMC_LOCK guardian.
    • If it is IMC_LOCKED, OK. Record K1:IMC-MCL_TRANS_OUTPUT.
      • If it is ~0.4, OK.
      • If not, check IMC alignment.
    • If not, check IMC alignment.
  • Record IMC oplev values below. Check if they are drifting or not.

     K1:VIS-MCE_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCE_TM_OPLEV_YAW_OUTPUT
     K1:VIS-MCI_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCI_TM_OPLEV_YAW_OUTPUT
     K1:VIS-MCO_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCO_TM_OPLEV_YAW_OUTPUT
  • Check the state of MICH_LOCK guardian.
    • If it is MICH_LOCKED, OK. Check if K1:LSC-REFL_PDA1_DC_OUTPUT is fluctuating or not.
      • If it is kind of constant, OK.
      • If not, check Michelson alignment (see JGW-T1605027 for alignment procedure).

    • If not, check Michelson alignment (see JGW-T1605027 for alignment procedure).

  • Record oplev values for Michelson mirrors. Check if they are drifting or not.

     K1:VIS-PR3_TM_OPLEV_PIT_OUTPUT
     K1:VIS-PR3_TM_OPLEV_YAW_OUTPUT
     K1:VIS-BS_TM_OPLEV_PIT_OUTPUT
     K1:VIS-BS_TM_OPLEV_YAW_OUTPUT
     K1:VIS-ETMX_TM_OPLEV_PIT_OUTPUT
     K1:VIS-ETMX_TM_OPLEV_YAW_OUTPUT
     K1:VIS-ETMY_TM_OPLEV_PIT_OUTPUT
     K1:VIS-ETMY_TM_OPLEV_YAW_OUTPUT


General

  • Type "mk" on terminal and you will open a "sitemap" screen

PSL (no PSL remote control in March)

Lock PMC

  • Open the "PSL PMC" medm screen from sitemap

PMC medm

  • Open StripTool /users/striptool/pmc.stp (link is put on the PMC screen)

  • Find PMC_TRANS Camera
  • Open the Guardian screen (link on the PMC screen)

PMC Guardian

  • 1. Request "PSL_LOCKED"
  • 2. Execute (Choose "EXEC")
  • Make sure that PMC is locked with the StripTool, PMC_TRANS camera, and the Guardian. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 6, PMC is locked. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 0, unlocked.

IMC

Lock IMC

  • Check MCI, MCO and MCE (link to VIS page. DC alignment values, Watchdog, Dackill)
  • Open IMC medm from sitemap

IMC medm

IMC Guardian

  • The usage of the IMC Guardian is almost same as one of PMC. Request "IMC_LOCKED" and execute.
  • NOTE: If PMC is unlocked, you will see a message "PSL_is_unlocked!" in USERMSG line in the IMC Guardian. When you see it, check if PMC is locked or not
  • Open StripTool /users/stiptool/IMC.stp

  • Find MCE_TRANS camera
  • Make sure IMC is locked with the StripTool, MCE_TRANS camera, and the Guardian. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.40 - 0.46 , IMC is locked. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.020, unlocked.

Main IFO

Lock Michelson IFO

  • Open /users/striptool/REFLPD_RMS.stp. You can see REFL PDA1
  • Control panel for MICH length-sensing control (LSC) is sitemap => LSC => iKAGRA LSC

  • make sure MICH_LOCK Guardian showing "operation"

    MICH Guardian]]

Check VIS

  • Check PR3, BS, IX and IY (link to VIS page)

Initial Alignment

Lock Michelson IFO

  • Establish the lock procedure
  • Prepare Guardian and put the manual here
  • Prepare StripTool (list of check channels)

Check DAQ

  • MonitCheck Check if all the DAQ system is working properly

  • DaqCheck1 Check if the real time models are sending data to DAQ

  • DaqCheck2 Check if the GW channel is properly recorded in the frames

  • Check Detchar screen?
  • Check glitch rate (prepare a script?)
  • Check noise lines (prepare a script?)

In the case of IFO/subsystem failure or emergency

KAGRA/Observation/iKAGRA/Manual/MIF (last edited 2016-03-25 10:56:26 by YutaMichimura)