Differences between revisions 5 and 51 (spanning 46 versions)
Revision 5 as of 2016-02-04 09:52:47
Size: 690
Comment:
Revision 51 as of 2016-03-25 10:56:26
Size: 4686
Comment:
Deletions are marked like this. Additions are marked like this.
Line 2: Line 2:
== PSL ==
== Take logbook if ==
 * PSL PMC looses lock
 * IMC does not lock even if PSL PMC is locking
 * IMC keeps locking and unlocking for more than an hour
 * Main interferometer does not lock even if IMC is locked
 * Main interferometer does not lock for more than 20 mins
 * Each mirror’s oplev feedback signals go more than 6000 cts.


== CHECK LIST ==
 * Record K1:PSL-PMC_TRANS_DC_OUTPUT.
  * If it is ~6, OK.
  * If not, ask Miyakawa, Kawamura, or Enomoto to manually lock PMC.
 * Check the state of IMC_LOCK guardian.
  * If it is IMC_LOCKED, OK. Record K1:IMC-MCL_TRANS_OUTPUT.
   * If it is ~0.4, OK.
   * If not, check IMC alignment.
  * If not, check IMC alignment.
 * Record IMC oplev values below. Check if they are drifting or not.{{{
 K1:VIS-MCE_TM_OPLEV_PIT_OUTPUT
 K1:VIS-MCE_TM_OPLEV_YAW_OUTPUT
 K1:VIS-MCI_TM_OPLEV_PIT_OUTPUT
 K1:VIS-MCI_TM_OPLEV_YAW_OUTPUT
 K1:VIS-MCO_TM_OPLEV_PIT_OUTPUT
 K1:VIS-MCO_TM_OPLEV_YAW_OUTPUT
}}}
 * Check the state of MICH_LOCK guardian.
  * If it is OPERATION, OK. Check if K1:LSC-REFL_PDA1_DC_OUTPUT is fluctuating or not.
   * If it is kind of constant, OK.
   * If not, check Michelson alignment (see [[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5027|JGW-T1605027]] for alignment procedure).
  * If not, check Michelson alignment (see [[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5027|JGW-T1605027]] for alignment procedure).
 * Record oplev values for Michelson mirrors. Check if they are drifting or not.{{{
 K1:VIS-PR3_TM_OPLEV_PIT_OUTPUT
 K1:VIS-PR3_TM_OPLEV_YAW_OUTPUT
 K1:VIS-BS_TM_OPLEV_PIT_OUTPUT
 K1:VIS-BS_TM_OPLEV_YAW_OUTPUT
 K1:VIS-ETMX_TM_OPLEV_PIT_OUTPUT
 K1:VIS-ETMX_TM_OPLEV_YAW_OUTPUT
 K1:VIS-ETMY_TM_OPLEV_PIT_OUTPUT
 K1:VIS-ETMY_TM_OPLEV_YAW_OUTPUT
}}}

----

== General ==
 * Type "mk" on terminal and you will open a "sitemap" screen

== PSL (no PSL remote control in March) ==
Line 5: Line 53:
 * Prepare Guardian
 * Prepare StripTool (list of check channels)
 * Open the "PSL PMC" medm screen from sitemap
[[attachment:PMC_medm.png|{{attachment:PMC_medm.png|PMC medm|width="600"}}]]
 * Open StripTool /users/striptool/pmc.stp (link is put on the PMC screen)
 * Find PMC_TRANS Camera
 * Open the Guardian screen (link on the PMC screen)
[[attachment:PMC_Guardian.png|{{attachment:PMC_Guardian.png|PMC Guardian|width="600"}}]]
 * 1. Request "PSL_LOCKED"
 * 2. Execute (Choose "EXEC")
 * Make sure that PMC is locked with the StripTool, PMC_TRANS camera, and the Guardian. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 6, PMC is locked. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 0, unlocked.
Line 11: Line 66:

 * Put Guardian usage here
 * Prepare StripTool (list of check channels)
 * Check MCI, MCO and MCE (link to VIS page. DC alignment values, Watchdog, Dackill)
 * Open IMC medm from sitemap
[[attachment:スクリーンショット 2016-03-08 10.37.45.png|{{attachment:スクリーンショット 2016-03-08 10.37.45.png|IMC medm|width="600"}}]]
 * Open StripTools and Guardian from IMC medm
[[attachment:IMC_guard.png|{{attachment:IMC_guard.png|IMC Guardian|width="600"}}]]
 * The usage of the IMC Guardian is almost same as one of PMC. Request "IMC_LOCKED" and execute.
 * NOTE: If PMC is unlocked, you will see a message "PSL_is_unlocked!" in USERMSG line in the IMC Guardian. When you see it, check if PMC is locked or not
 * Open StripTool /users/stiptool/IMC.stp
 * Find MCE_TRANS camera
 * Make sure IMC is locked with the StripTool, MCE_TRANS camera, and the Guardian. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.40 - 0.46 , IMC is locked. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.020, unlocked.
Line 18: Line 79:

=== Lock Michelson IFO ===
 * Open /users/striptool/REFLPD_RMS.stp. You can see REFL PDA1
 * Control panel for MICH length-sensing control (LSC) is sitemap => LSC => iKAGRA LSC
 * make sure MICH_LOCK Guardian showing "operation"
 [[attachment:MICH_guardian.png|{{attachment:MICH_guardian.png|MICH Guardian|width="600"}}]]]]
 



=== Check VIS ===
 * Check PR3, BS, IX and IY (link to VIS page)
Line 19: Line 93:
 * Establish the alignment procedure
 * Make the initial alignment script
 * Put the script usage
 
 * [[http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5027]]
 * Educate Experts
Line 29: Line 102:
 * Check if the GW channel is properly recorded in the frames  * MonitCheck Check if all the DAQ system is working properly
 * DaqCheck1 Check if the real time models are sending data to DAQ
 * DaqCheck2 Check if the GW channel is properly recorded in the frames
 * Check Detchar screen?
 * Check glitch rate (prepare a script?)
 * Check noise lines (prepare a script?)
Line 31: Line 109:
= In the case of IFO/susbsystem failure or emergency = = In the case of IFO/subsystem failure or emergency =

IFO Operation Manual

Take logbook if

  • PSL PMC looses lock
  • IMC does not lock even if PSL PMC is locking
  • IMC keeps locking and unlocking for more than an hour
  • Main interferometer does not lock even if IMC is locked
  • Main interferometer does not lock for more than 20 mins
  • Each mirror’s oplev feedback signals go more than 6000 cts.

CHECK LIST

  • Record K1:PSL-PMC_TRANS_DC_OUTPUT.
    • If it is ~6, OK.
    • If not, ask Miyakawa, Kawamura, or Enomoto to manually lock PMC.
  • Check the state of IMC_LOCK guardian.
    • If it is IMC_LOCKED, OK. Record K1:IMC-MCL_TRANS_OUTPUT.
      • If it is ~0.4, OK.
      • If not, check IMC alignment.
    • If not, check IMC alignment.
  • Record IMC oplev values below. Check if they are drifting or not.

     K1:VIS-MCE_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCE_TM_OPLEV_YAW_OUTPUT
     K1:VIS-MCI_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCI_TM_OPLEV_YAW_OUTPUT
     K1:VIS-MCO_TM_OPLEV_PIT_OUTPUT
     K1:VIS-MCO_TM_OPLEV_YAW_OUTPUT
  • Check the state of MICH_LOCK guardian.
    • If it is OPERATION, OK. Check if K1:LSC-REFL_PDA1_DC_OUTPUT is fluctuating or not.
      • If it is kind of constant, OK.
      • If not, check Michelson alignment (see JGW-T1605027 for alignment procedure).

    • If not, check Michelson alignment (see JGW-T1605027 for alignment procedure).

  • Record oplev values for Michelson mirrors. Check if they are drifting or not.

     K1:VIS-PR3_TM_OPLEV_PIT_OUTPUT
     K1:VIS-PR3_TM_OPLEV_YAW_OUTPUT
     K1:VIS-BS_TM_OPLEV_PIT_OUTPUT
     K1:VIS-BS_TM_OPLEV_YAW_OUTPUT
     K1:VIS-ETMX_TM_OPLEV_PIT_OUTPUT
     K1:VIS-ETMX_TM_OPLEV_YAW_OUTPUT
     K1:VIS-ETMY_TM_OPLEV_PIT_OUTPUT
     K1:VIS-ETMY_TM_OPLEV_YAW_OUTPUT


General

  • Type "mk" on terminal and you will open a "sitemap" screen

PSL (no PSL remote control in March)

Lock PMC

  • Open the "PSL PMC" medm screen from sitemap

PMC medm

  • Open StripTool /users/striptool/pmc.stp (link is put on the PMC screen)

  • Find PMC_TRANS Camera
  • Open the Guardian screen (link on the PMC screen)

PMC Guardian

  • 1. Request "PSL_LOCKED"
  • 2. Execute (Choose "EXEC")
  • Make sure that PMC is locked with the StripTool, PMC_TRANS camera, and the Guardian. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 6, PMC is locked. If K1:PSL-PMC_TRANS_DC_OUTPUT ~ 0, unlocked.

IMC

Lock IMC

  • Check MCI, MCO and MCE (link to VIS page. DC alignment values, Watchdog, Dackill)
  • Open IMC medm from sitemap

IMC medm

IMC Guardian

  • The usage of the IMC Guardian is almost same as one of PMC. Request "IMC_LOCKED" and execute.
  • NOTE: If PMC is unlocked, you will see a message "PSL_is_unlocked!" in USERMSG line in the IMC Guardian. When you see it, check if PMC is locked or not
  • Open StripTool /users/stiptool/IMC.stp

  • Find MCE_TRANS camera
  • Make sure IMC is locked with the StripTool, MCE_TRANS camera, and the Guardian. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.40 - 0.46 , IMC is locked. If K1:IMC-MCL_TRANS_OUTPUT ~ 0.020, unlocked.

Main IFO

Lock Michelson IFO

  • Open /users/striptool/REFLPD_RMS.stp. You can see REFL PDA1
  • Control panel for MICH length-sensing control (LSC) is sitemap => LSC => iKAGRA LSC

  • make sure MICH_LOCK Guardian showing "operation"

    MICH Guardian]]

Check VIS

  • Check PR3, BS, IX and IY (link to VIS page)

Initial Alignment

Lock Michelson IFO

  • Establish the lock procedure
  • Prepare Guardian and put the manual here
  • Prepare StripTool (list of check channels)

Check DAQ

  • MonitCheck Check if all the DAQ system is working properly

  • DaqCheck1 Check if the real time models are sending data to DAQ

  • DaqCheck2 Check if the GW channel is properly recorded in the frames

  • Check Detchar screen?
  • Check glitch rate (prepare a script?)
  • Check noise lines (prepare a script?)

In the case of IFO/subsystem failure or emergency

KAGRA/Observation/iKAGRA/Manual/MIF (last edited 2016-03-25 10:56:26 by YutaMichimura)