 1. Check point (1): Do OSEM values stay stable within +/- 3 um?
   The
 * Checkpoint (1): Do OSEM values stay stable within +/- 3 um?
   The values in the circle of check point (1) at the figure above represent the OSEM input value calibrated to um and decoupled to each degree of freedom.
  These values should not fluctuate more than 3 um during the damping servo is on.
 If an earthquake occurs, the value will fluctuate around 3 um. However, they should become stable within 1 min.
Line 29: Line 31:
 * Checkpoint (2): OpLev values are at the target values?
   The values in the circle of check point (2) at the figure above represent the Oplev input value calibrated to urad.
  In order to align the laser beam, these values are as same as the -1 x (target values) which suggested from MIF.

 * Checkpoint (3): Are all the servo filters on?
   If the servo filters are on, then the rectangular ramps are green. If off, they are red.
  Check if the color of the all rectangular ramps, which are in the rectangle of check point (3), are same as the figure above.

 * Checkpoint (4): Is the set-point and DC control set properly?
   Click 'OL DAMP' box above 'IM TEST' box. Check the value in the 'offset' in each path, whether these values are same as the target values.
 
 * Checkpoint (5): Do you have TM H1 input value?
   Check the values in the rectangles of check point (5) in the figure above.
  If the value at the right is almost half of the value at the left, it is OK. If the value at the right is close to 0, Go to Troubleshoot B.

 * Checkpoint (6): Is WatchDog or DACKill tripped?
   Check if you can see the red bars in check point (6) rectangle in the figure above.
  If the ramp on the 'WatchDog' box is red, the WatchDog is tripped.
 If the ramp on the 'DACKILL IOP' box is red, the DACKill is working.


==== How to troubleshoot ====
 * Troubleshoot A: Set DC control
    

Mode Cleaner Suspensions

Each MC suspension has 4 coils and 2 pico-motors (pitch, yaw).

ETM

Each iKAGRA ETM suspension has 4 coils and 4 pico-motors (pitch:1-6, yaw:2-7, z:3-8, y:4-9).

If you find a problem about coil-magnet actuator, please check 2 things.

1. Coil drivers in particular for Y-end had a loose connection with power line (-18V, -15V).

Please check the LED on coil driver. If the LED is out, try to fasten the power line.

2. Sometimes watchdog? system stop the output current from the coil driver in order to protect mechanical system.

PR3

Check the PR3 status according to the flowchart below.

text describing image

Details of the check points

text describing image text describing image

  • Checkpoint (1): Do OSEM values stay stable within +/- 3 um?
    • The values in the circle of check point (1) at the figure above represent the OSEM input value calibrated to um and decoupled to each degree of freedom.
    • These values should not fluctuate more than 3 um during the damping servo is on.
    If an earthquake occurs, the value will fluctuate around 3 um. However, they should become stable within 1 min.
  • Checkpoint (2): OpLev values are at the target values?

    • The values in the circle of check point (2) at the figure above represent the Oplev input value calibrated to urad.
    • In order to align the laser beam, these values are as same as the -1 x (target values) which suggested from MIF.
  • Checkpoint (3): Are all the servo filters on?
    • If the servo filters are on, then the rectangular ramps are green. If off, they are red.
    • Check if the color of the all rectangular ramps, which are in the rectangle of check point (3), are same as the figure above.
  • Checkpoint (4): Is the set-point and DC control set properly?
    • Click 'OL DAMP' box above 'IM TEST' box. Check the value in the 'offset' in each path, whether these values are same as the target values.
  • Checkpoint (5): Do you have TM H1 input value?
    • Check the values in the rectangles of check point (5) in the figure above.
    • If the value at the right is almost half of the value at the left, it is OK. If the value at the right is close to 0, Go to Troubleshoot B.
  • Checkpoint (6): Is WatchDog or DACKill tripped?

    • Check if you can see the red bars in check point (6) rectangle in the figure above.
    If the ramp on the 'DACKILL IOP' box is red, the DACKill is working.

How to troubleshoot

  • Troubleshoot A: Set DC control

Appendix

text describing image