Line 16: Line 16:
[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8466|TMSX-VISのgeophoneのデータシート]]
Line 21: Line 19:
{{attachment:Hgeo.png||width=500}}
Line 22: Line 21:
||geophone||serial number||
||H1||#4||
||H2||#5||
||H3||#2||
||V1||404084242||
||V2||404080976||
||V3||404084241||
 * G_e: Generator Constant [V/(m/s)]
 * η : Damping ratio
 * ω0:Resonant angular frequency of geophone [rad/s]


[[https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=8466|TMSX-VISのgeophoneのデータシート]]

||geophone||serial number||G_e||η||ω0=2pi*f0||
||H1||#4||287.91||0.285||2pi*1.01||
||H2||#5||285.68||0.284||2pi*1.01||
||H3||#2||281.92||0.290||2pi*0.97||
||V1||404084242||274.76||0.296||2pi*0.96||
||V2||404080976||286.11||0.299||2pi*1||
||V3||404084241||273.74||0.293||2pi*1||

19/05/22

What we did

  • geophoneのcalibration

今のところ(追記,19/05/24現在),フィルター通過後のgeophoneの信号が変になったり,発散する様子は見られない。 データシートを元に,geophoneのcalibrationファクターを詰めた。

参考

関口さんのD論

ETMXのIPのgeophoneでのcalibration

  • geophoneの周波数応答

Hgeo.png

  • G_e: Generator Constant [V/(m/s)]
  • η : Damping ratio
  • ω0:Resonant angular frequency of geophone [rad/s]

TMSX-VISのgeophoneのデータシート

geophone

serial number

G_e

η

ω0=2pi*f0

H1

#4

287.91

0.285

2pi*1.01

H2

#5

285.68

0.284

2pi*1.01

H3

#2

281.92

0.290

2pi*0.97

V1

404084242

274.76

0.296

2pi*0.96

V2

404080976

286.11

0.299

2pi*1

V3

404084241

273.74

0.293

2pi*1

geophoneの応答

KAGRA/Subgroups/AOS/TMSX_O3/2019-05-22 (last edited 2019-06-04 19:46:45 by kenta.tanaka)