Calibration work memo 2018/4/28 09:00-
Site worker : T.Yamamoto(leader), S.Haino, Y.Inoue, S. Tsuchida, A.Miyamoto, T.Yokozawa
Free swinging
- 09:25 Unlock michelson
- 9:32 Started line injection to BS
- K1:LSC-REFLAIR_A_RF17_Q_ERR_256_DQ (RF signal of PD)
- K1:LSC-REFLAIR_B_LF_OUT_256_DQ (DC signal of PD)
- K1:CAL-CS_LINE_LINE8_LKIN_OSC_FREQ (injection line frequency)
- K1:CAL-CS_LINE_MTRX_1_8 (Amplitude of calibration line)
Table of line injection to BS
Time |
Freq[Hz] |
amp[count] |
9:25 |
0.0 |
0 |
9:32 |
0.2 |
5 |
9:33 |
0.2 |
10 |
9:34 |
0.2 |
20 |
9:36 |
0.2 |
100 |
9:47 |
0.2 |
250 |
9:53 |
0.2 |
500 |
- 10:10 Finish free swining measurement
- Offline analysis
- /users/CAL/PHASE1/scripts/EllipseFitting.C
Pre-offline best result png
2*pi*2A/lambda = 1.00e+11[cnt/m] = 3.00e+14[cnt/strain]
- 11:35 Open loop transfer function measurement
- template diaggui /users/CAL/PHASE1/templates/DTT/T_OPTF.dtt.xml
- 1st measurement : /users/CAL/PHASE1/measurements/DTT/T_OLTF/T_OLTF_180428_114110.xml
- 2nd measurement : /users/CAL/PHASE1/measurements/DTT/T_OLTF/T_OLTF_180428_115610.xml
- 3rd measurement : /users/CAL/PHASE1/measurements/DTT/T_OLTF/T_OLTF_180428_124641.xml
- 12:15 Coil driver transfer function measurement
- transfer function measurement between coilout and FC
- EXC(K1:VIS-BS_TM_COILOUTF_H4_EXC) MEADURE(K1:VIS-BS_SCMON_TM_H4)
- Be careful, channel name is including SC, but connect to FC.
- ??:?? Modified calibration model(C00 Strain)
- z-8 delay between error and control
- optical gain
- Discussed about unknown time delay (1block:1/16384=61usec)
3 known block delay : LSC->BS, DAC, LSC->CAL?
- aAA, dAA, aAI, dAI
- iKAGRA has 130usec unknown delay...
- ??:?? Start Open loop transfer function measurement
- with envelop, measurement accuracy becomes much better!
- measurements : /users/CAL/PHASE1/measurements/DTT/T_OLTF/T_OLTF_180428_******.xml