Line 5: | Line 5: |
* | * 09:25 Unlock michelson * 9:32 Started line injection to BS * K1:LSC-REFLAIR_A_RF17_Q_ERR_256_DQ (RF signal of PD) * K1:LSC-REFLAIR_B_LF_OUT_256_DQ (DC signal of PD) * K1:CAL-CS_LINE_LINE8_LKIN_OSC_FREQ (injection line frequency) * K1:CAL-CS_LINE_MTRX_1_8 (Amplitude of calibration line) Table of line injection to BS || Time || Freq[Hz] || amp[count] || || 9:25 || 0.0 || 0 || || 9:32 || 0.2 || 5 || || 9:33 || 0.2 || 10 || || 9:34 || 0.2 || 20 || || 9:36 || 0.2 || 100 || || 9:47 || 0.2 || 250 || || 9:53 || 0.2 || 500 || * 10:10 Finish free swining measurement * Offline analysis * /users/CAL/PHASE1/scripts/EllipseFitting.C * Pre-offline best result[[attachment:openloop_best|png]] |
Calibration work memo 2018/4/28 09:00-
Site worker : T.Yamamoto(leader), S.Haino, Y.Inoue, S. Tsuchida, A.Miyamoto, T.Yokozawa
Free swinging
- 09:25 Unlock michelson
- 9:32 Started line injection to BS
- K1:LSC-REFLAIR_A_RF17_Q_ERR_256_DQ (RF signal of PD)
- K1:LSC-REFLAIR_B_LF_OUT_256_DQ (DC signal of PD)
- K1:CAL-CS_LINE_LINE8_LKIN_OSC_FREQ (injection line frequency)
- K1:CAL-CS_LINE_MTRX_1_8 (Amplitude of calibration line)
Table of line injection to BS
Time |
Freq[Hz] |
amp[count] |
9:25 |
0.0 |
0 |
9:32 |
0.2 |
5 |
9:33 |
0.2 |
10 |
9:34 |
0.2 |
20 |
9:36 |
0.2 |
100 |
9:47 |
0.2 |
250 |
9:53 |
0.2 |
500 |
- 10:10 Finish free swining measurement
- Offline analysis
- /users/CAL/PHASE1/scripts/EllipseFitting.C
Pre-offline best resultpng