Differences between revisions 4 and 6 (spanning 2 versions)
Revision 4 as of 2019-11-18 17:15:30
Size: 1025
Comment:
Revision 6 as of 2019-11-18 17:20:09
Size: 1241
Comment:
Deletions are marked like this. Additions are marked like this.
Line 18: Line 18:

 * The total opening time is 54msec. The shutter speed is higher than this if the beam is placed near the center of the shutter.
 * The shutter is vacuum compatible: 1e-8 torr.

 * The shutter driver is VED24.

Beam shutter

Shutter for the input beam

  • We tried to install a medium-speed shutter in 2017 to shut the input beam after the IMMTs so that installation work after the PRM can be done without wearing safety glasses.
  • However, the shutter did not work somehow and the anticipation failed.

Shutter for the output mode-cleaner

  • We planned to reuse the shutter to shut the beam before the OMC so that the OMC can be protected until the main interferometer is robustly operated in the dark fringe.
  • We planned to use the same shutter we purchased for the input beam in 2017.

Shutter spec

  • The shutter is a product by Uniblitz. The product name is CS65E1ZM0-ECE. The blade is made of BeCu and coated by AlMgF2. The damage threshold for 1064nm CW is 2.5W/mm2.

  • If the shutter is used after the IMMT2, the beam radius is 4mm and the max tolerable power will be 125W.
  • If the shutter is used before the OMC, the beam radius is 0.45mm and the max tolerable power will be 1.5W.
  • The total opening time is 54msec. The shutter speed is higher than this if the beam is placed near the center of the shutter.
  • The shutter is vacuum compatible: 1e-8 torr.
  • The shutter driver is VED24.

KAGRA/Subgroups/IOO/Shutter (last edited 2019-11-18 17:20:09 by KentaroSomiya)